An Improved Design of the Spiral-Coil EMAT for Enhancing the Signal Amplitude

The low energy transition efficiency of electromagnetic ultrasonic transducer (EMAT) is a common problem in practical application. For the purpose of enhancing the amplitude of the received signal, an improved double-coil bulk wave EMAT is proposed for the thickness measurement of metallic block. Th...

Descripción completa

Detalles Bibliográficos
Autores principales: Jia, Xiaojuan, Ouyang, Qi, Zhang, Xinglan
Formato: Online Artículo Texto
Lenguaje:English
Publicado: MDPI 2017
Materias:
Acceso en línea:https://www.ncbi.nlm.nih.gov/pmc/articles/PMC5470782/
https://www.ncbi.nlm.nih.gov/pubmed/28498307
http://dx.doi.org/10.3390/s17051106
Descripción
Sumario:The low energy transition efficiency of electromagnetic ultrasonic transducer (EMAT) is a common problem in practical application. For the purpose of enhancing the amplitude of the received signal, an improved double-coil bulk wave EMAT is proposed for the thickness measurement of metallic block. This new configuration of magnets consists of a solid cylindrical magnet and a ring-shaped magnet encircling the outer side of the solid cylindrical one. A double-coil was applied instead of a single spiral-coil. Numerical simulations were performed to analyze and optimize the proposed configuration of the EMAT by the 2-D axisymmetric finite element model (FEM). The experiment effectively verifies the rationality of the new configuration and the feasibility of improving the signal strength.