Cargando…

Evaluation of the performance of Abbott m2000 and Roche COBAS Ampliprep/COBAS Taqman assays for HIV-1 viral load determination using dried blood spots and dried plasma spots in Kenya

BACKGROUND: Routine HIV viral load testing is not widely accessible in most resource-limited settings, including Kenya. To increase access to viral load testing, alternative sample types like dried blood spots (DBS), which overcome the logistic barriers associated with plasma separation and cold cha...

Descripción completa

Detalles Bibliográficos
Autores principales: Zeh, Clement, Ndiege, Kenneth, Inzaule, Seth, Achieng, Rebecca, Williamson, John, Chih-Wei Chang, Joy, Ellenberger, Dennis, Nkengasong, John
Formato: Online Artículo Texto
Lenguaje:English
Publicado: Public Library of Science 2017
Materias:
Acceso en línea:https://www.ncbi.nlm.nih.gov/pmc/articles/PMC5473550/
https://www.ncbi.nlm.nih.gov/pubmed/28622370
http://dx.doi.org/10.1371/journal.pone.0179316
_version_ 1783244306448908288
author Zeh, Clement
Ndiege, Kenneth
Inzaule, Seth
Achieng, Rebecca
Williamson, John
Chih-Wei Chang, Joy
Ellenberger, Dennis
Nkengasong, John
author_facet Zeh, Clement
Ndiege, Kenneth
Inzaule, Seth
Achieng, Rebecca
Williamson, John
Chih-Wei Chang, Joy
Ellenberger, Dennis
Nkengasong, John
author_sort Zeh, Clement
collection PubMed
description BACKGROUND: Routine HIV viral load testing is not widely accessible in most resource-limited settings, including Kenya. To increase access to viral load testing, alternative sample types like dried blood spots (DBS), which overcome the logistic barriers associated with plasma separation and cold chain shipment need to be considered and evaluated. The current study evaluated matched dried blood spots (DBS) and dried plasma spots (DPS) against plasma using the Abbott M 2000 (Abbott) and Roche Cobas Ampliprep/Cobas TaqMan (CAP/CTM) quantitative viral load assays in western Kenya. METHODS: Matched plasma DBS and DPS were obtained from 200 HIV-1 infected antiretroviral treatment (ART)-experienced patients attending patient support centers in Western Kenya. Standard quantitative assay performance parameters with accompanying 95% confidence intervals (CI) were assessed at the assays lower detection limit (400cps/ml for CAP/CTM and 550cps/ml for Abbott) using SAS version 9.2. Receiver operating curves (ROC) were further used to assess viral-load thresholds with best assay performance (reference assay CAP/CTM plasma). RESULTS: Using the Abbott test, the sensitivity and specificity, respectively, for DPS were (97.3%, [95%CI: 93.2–99.2] and 98.1% [95%CI: 89.7–100]) and those for DBS (93.9% [95%CI: 88.8–97.2] and 88.0% [95%CI: 82.2–92.4]). The correlation and agreement using paired plasma and DPS/DBS were strong, with r(2) = 90.5 and r(c) = 68.1. The Bland-Altman relative percent change was 95.3 for DPS, (95%CI: 90.4–97.7) and 73.6 (95%CI: 51.6–86.5) for DBS. Using the CAP/CTM assay, the sensitivity for DBS was significantly higher compared to DPS (100.0% [95% CI: 97.6–100.0] vs. 94.7% [95%CI: 89.8–97.7]), while the specificity for DBS was lower: 4%, [95% CI: 0.4–13.7] compared to DPS: 94.0%, [95% CI: 83.5–98.7]. When compared under different clinical relevant thresholds, the accuracy for the Abbott assay was 95% at the 1000cps/ml cut-off with a sensitivity and specificity of 96.6% [95% CI 91.8–98.7] and 90.4% [95% CI 78.2–96.4] respectively. The optimum threshold was at 3000 cps/ml with an accuracy of 95.5%, sensitivity and specificity of 94.6% [95%CI 89.3–97.5] and 98.1% [95%CI 88.4–99.9]) respectively. The best threshold for CAP/CTM was at 4000 copies /mL, with 92.5% accuracy (sensitivity of 96.0% [95%CI 91.0–98.3] and specificity of 82.7% [95%CI 69.2–91.3]). CONCLUSIONS: There was similar performance between matched DBS, DPS and plasma using the Abbott test, and good correlation for matched DPS and plasma using the CAPCTM test. The findings suggest that DBS and DPS may be reliably used as alternative specimens to plasma to measure HIV-1 VL using Abbott, and DPS may be reliably used with CAP/CTM in resource-limited settings.
format Online
Article
Text
id pubmed-5473550
institution National Center for Biotechnology Information
language English
publishDate 2017
publisher Public Library of Science
record_format MEDLINE/PubMed
spelling pubmed-54735502017-06-22 Evaluation of the performance of Abbott m2000 and Roche COBAS Ampliprep/COBAS Taqman assays for HIV-1 viral load determination using dried blood spots and dried plasma spots in Kenya Zeh, Clement Ndiege, Kenneth Inzaule, Seth Achieng, Rebecca Williamson, John Chih-Wei Chang, Joy Ellenberger, Dennis Nkengasong, John PLoS One Research Article BACKGROUND: Routine HIV viral load testing is not widely accessible in most resource-limited settings, including Kenya. To increase access to viral load testing, alternative sample types like dried blood spots (DBS), which overcome the logistic barriers associated with plasma separation and cold chain shipment need to be considered and evaluated. The current study evaluated matched dried blood spots (DBS) and dried plasma spots (DPS) against plasma using the Abbott M 2000 (Abbott) and Roche Cobas Ampliprep/Cobas TaqMan (CAP/CTM) quantitative viral load assays in western Kenya. METHODS: Matched plasma DBS and DPS were obtained from 200 HIV-1 infected antiretroviral treatment (ART)-experienced patients attending patient support centers in Western Kenya. Standard quantitative assay performance parameters with accompanying 95% confidence intervals (CI) were assessed at the assays lower detection limit (400cps/ml for CAP/CTM and 550cps/ml for Abbott) using SAS version 9.2. Receiver operating curves (ROC) were further used to assess viral-load thresholds with best assay performance (reference assay CAP/CTM plasma). RESULTS: Using the Abbott test, the sensitivity and specificity, respectively, for DPS were (97.3%, [95%CI: 93.2–99.2] and 98.1% [95%CI: 89.7–100]) and those for DBS (93.9% [95%CI: 88.8–97.2] and 88.0% [95%CI: 82.2–92.4]). The correlation and agreement using paired plasma and DPS/DBS were strong, with r(2) = 90.5 and r(c) = 68.1. The Bland-Altman relative percent change was 95.3 for DPS, (95%CI: 90.4–97.7) and 73.6 (95%CI: 51.6–86.5) for DBS. Using the CAP/CTM assay, the sensitivity for DBS was significantly higher compared to DPS (100.0% [95% CI: 97.6–100.0] vs. 94.7% [95%CI: 89.8–97.7]), while the specificity for DBS was lower: 4%, [95% CI: 0.4–13.7] compared to DPS: 94.0%, [95% CI: 83.5–98.7]. When compared under different clinical relevant thresholds, the accuracy for the Abbott assay was 95% at the 1000cps/ml cut-off with a sensitivity and specificity of 96.6% [95% CI 91.8–98.7] and 90.4% [95% CI 78.2–96.4] respectively. The optimum threshold was at 3000 cps/ml with an accuracy of 95.5%, sensitivity and specificity of 94.6% [95%CI 89.3–97.5] and 98.1% [95%CI 88.4–99.9]) respectively. The best threshold for CAP/CTM was at 4000 copies /mL, with 92.5% accuracy (sensitivity of 96.0% [95%CI 91.0–98.3] and specificity of 82.7% [95%CI 69.2–91.3]). CONCLUSIONS: There was similar performance between matched DBS, DPS and plasma using the Abbott test, and good correlation for matched DPS and plasma using the CAPCTM test. The findings suggest that DBS and DPS may be reliably used as alternative specimens to plasma to measure HIV-1 VL using Abbott, and DPS may be reliably used with CAP/CTM in resource-limited settings. Public Library of Science 2017-06-16 /pmc/articles/PMC5473550/ /pubmed/28622370 http://dx.doi.org/10.1371/journal.pone.0179316 Text en https://creativecommons.org/publicdomain/zero/1.0/ This is an open access article, free of all copyright, and may be freely reproduced, distributed, transmitted, modified, built upon, or otherwise used by anyone for any lawful purpose. The work is made available under the Creative Commons CC0 (https://creativecommons.org/publicdomain/zero/1.0/) public domain dedication.
spellingShingle Research Article
Zeh, Clement
Ndiege, Kenneth
Inzaule, Seth
Achieng, Rebecca
Williamson, John
Chih-Wei Chang, Joy
Ellenberger, Dennis
Nkengasong, John
Evaluation of the performance of Abbott m2000 and Roche COBAS Ampliprep/COBAS Taqman assays for HIV-1 viral load determination using dried blood spots and dried plasma spots in Kenya
title Evaluation of the performance of Abbott m2000 and Roche COBAS Ampliprep/COBAS Taqman assays for HIV-1 viral load determination using dried blood spots and dried plasma spots in Kenya
title_full Evaluation of the performance of Abbott m2000 and Roche COBAS Ampliprep/COBAS Taqman assays for HIV-1 viral load determination using dried blood spots and dried plasma spots in Kenya
title_fullStr Evaluation of the performance of Abbott m2000 and Roche COBAS Ampliprep/COBAS Taqman assays for HIV-1 viral load determination using dried blood spots and dried plasma spots in Kenya
title_full_unstemmed Evaluation of the performance of Abbott m2000 and Roche COBAS Ampliprep/COBAS Taqman assays for HIV-1 viral load determination using dried blood spots and dried plasma spots in Kenya
title_short Evaluation of the performance of Abbott m2000 and Roche COBAS Ampliprep/COBAS Taqman assays for HIV-1 viral load determination using dried blood spots and dried plasma spots in Kenya
title_sort evaluation of the performance of abbott m2000 and roche cobas ampliprep/cobas taqman assays for hiv-1 viral load determination using dried blood spots and dried plasma spots in kenya
topic Research Article
url https://www.ncbi.nlm.nih.gov/pmc/articles/PMC5473550/
https://www.ncbi.nlm.nih.gov/pubmed/28622370
http://dx.doi.org/10.1371/journal.pone.0179316
work_keys_str_mv AT zehclement evaluationoftheperformanceofabbottm2000androchecobasampliprepcobastaqmanassaysforhiv1viralloaddeterminationusingdriedbloodspotsanddriedplasmaspotsinkenya
AT ndiegekenneth evaluationoftheperformanceofabbottm2000androchecobasampliprepcobastaqmanassaysforhiv1viralloaddeterminationusingdriedbloodspotsanddriedplasmaspotsinkenya
AT inzauleseth evaluationoftheperformanceofabbottm2000androchecobasampliprepcobastaqmanassaysforhiv1viralloaddeterminationusingdriedbloodspotsanddriedplasmaspotsinkenya
AT achiengrebecca evaluationoftheperformanceofabbottm2000androchecobasampliprepcobastaqmanassaysforhiv1viralloaddeterminationusingdriedbloodspotsanddriedplasmaspotsinkenya
AT williamsonjohn evaluationoftheperformanceofabbottm2000androchecobasampliprepcobastaqmanassaysforhiv1viralloaddeterminationusingdriedbloodspotsanddriedplasmaspotsinkenya
AT chihweichangjoy evaluationoftheperformanceofabbottm2000androchecobasampliprepcobastaqmanassaysforhiv1viralloaddeterminationusingdriedbloodspotsanddriedplasmaspotsinkenya
AT ellenbergerdennis evaluationoftheperformanceofabbottm2000androchecobasampliprepcobastaqmanassaysforhiv1viralloaddeterminationusingdriedbloodspotsanddriedplasmaspotsinkenya
AT nkengasongjohn evaluationoftheperformanceofabbottm2000androchecobasampliprepcobastaqmanassaysforhiv1viralloaddeterminationusingdriedbloodspotsanddriedplasmaspotsinkenya