Evaluation of a low-cost, low-power syringe pump to deliver magnesium sulfate intravenously to pre-eclamptic women in a Malawian referral hospital

BACKGROUND: Magnesium sulfate is an affordable and effective treatment for pre-eclampsia and eclampsia. In settings where infusion pumps are not available to regulate the flow rate of intravenous delivery, healthcare providers must administer magnesium sulfate (MgSO(4)) via time-consuming and painfu...

Descripción completa

Detalles Bibliográficos
Autores principales: Skerrett, Erica, Kommwa, Edward, Maynard, Kelley, Juarez, Alexa, Mataya, Ronald, Richards-Kortum, Rebecca, Oden, Z. Maria
Formato: Online Artículo Texto
Lenguaje:English
Publicado: BioMed Central 2017
Materias:
Acceso en línea:https://www.ncbi.nlm.nih.gov/pmc/articles/PMC5474875/
https://www.ncbi.nlm.nih.gov/pubmed/28629437
http://dx.doi.org/10.1186/s12884-017-1382-9
_version_ 1783244519001554944
author Skerrett, Erica
Kommwa, Edward
Maynard, Kelley
Juarez, Alexa
Mataya, Ronald
Richards-Kortum, Rebecca
Oden, Z. Maria
author_facet Skerrett, Erica
Kommwa, Edward
Maynard, Kelley
Juarez, Alexa
Mataya, Ronald
Richards-Kortum, Rebecca
Oden, Z. Maria
author_sort Skerrett, Erica
collection PubMed
description BACKGROUND: Magnesium sulfate is an affordable and effective treatment for pre-eclampsia and eclampsia. In settings where infusion pumps are not available to regulate the flow rate of intravenous delivery, healthcare providers must administer magnesium sulfate (MgSO(4)) via time-consuming and painful, large-volume intramuscular injections. As an alternative to costly commercially available syringe pumps, we developed AutoSyp, an accurate, low-cost, and low-powered syringe pump designed to meet the needs and constraints these low-resource settings. This paper describes results of a pilot study to evaluate the feasibility of using AutoSyp to administer MgSO(4) intravenously to women suffering from pre-eclampsia at a referral hospital in Blantyre, Malawi. METHODS: AutoSyp was programmed to deliver MgSO(4) following the Zuspan regimen to pregnant and post-partum women suffering from pre-eclampsia at Queen Elizabeth Central Hospital in Blatnyre, Malawi. Given the selection of either loading or maintenance dose on AutoSyp’s user interface, the flow rate was automatically programmed to dispense 60 mL/h or 5 mL/h of 20% MgSO(4) solution, respectively. During each treatment, the dispensed volume was automatically calculated by the device based on the plunger position and stored on a computer for accuracy analysis of the mean flow rate and total volume delivered. The clinical results for both the loading and maintenance dose administrations were compared to the device’s accuracy during tests performed in the laboratory setting. RESULTS: Twenty-two women were enrolled in this study. In both the clinical and laboratory settings, the mean flow rate errors for the loading and maintenance dose infusions were under 2%. During 466 h of testing, the device sounded 129 occlusion alarms across 14 subjects. Of these, 71 alarms were false positives. CONCLUSION: Results of this study support the use of AutoSyp as a less painful and accurate means of MgSO(4) administration in clinical environments that lack infusion systems. There were a large number of false alarms in the current system which will be addressed in future designs. AutoSyp maintains the comfort of intravenous MgSO(4) administration, but unlike commercially available syringe pumps, it is capable of operating with a variety of syringe brands and sizes and requires no additional consumables. AutoSyp’s appropriate design will benefit its implementation and sustained use in low-resource settings. TRIAL REGISTRATION: Trial registered prospectively on November 18, 2014 with ClinicalTrials.gov (NCT02296931)
format Online
Article
Text
id pubmed-5474875
institution National Center for Biotechnology Information
language English
publishDate 2017
publisher BioMed Central
record_format MEDLINE/PubMed
spelling pubmed-54748752017-06-21 Evaluation of a low-cost, low-power syringe pump to deliver magnesium sulfate intravenously to pre-eclamptic women in a Malawian referral hospital Skerrett, Erica Kommwa, Edward Maynard, Kelley Juarez, Alexa Mataya, Ronald Richards-Kortum, Rebecca Oden, Z. Maria BMC Pregnancy Childbirth Research Article BACKGROUND: Magnesium sulfate is an affordable and effective treatment for pre-eclampsia and eclampsia. In settings where infusion pumps are not available to regulate the flow rate of intravenous delivery, healthcare providers must administer magnesium sulfate (MgSO(4)) via time-consuming and painful, large-volume intramuscular injections. As an alternative to costly commercially available syringe pumps, we developed AutoSyp, an accurate, low-cost, and low-powered syringe pump designed to meet the needs and constraints these low-resource settings. This paper describes results of a pilot study to evaluate the feasibility of using AutoSyp to administer MgSO(4) intravenously to women suffering from pre-eclampsia at a referral hospital in Blantyre, Malawi. METHODS: AutoSyp was programmed to deliver MgSO(4) following the Zuspan regimen to pregnant and post-partum women suffering from pre-eclampsia at Queen Elizabeth Central Hospital in Blatnyre, Malawi. Given the selection of either loading or maintenance dose on AutoSyp’s user interface, the flow rate was automatically programmed to dispense 60 mL/h or 5 mL/h of 20% MgSO(4) solution, respectively. During each treatment, the dispensed volume was automatically calculated by the device based on the plunger position and stored on a computer for accuracy analysis of the mean flow rate and total volume delivered. The clinical results for both the loading and maintenance dose administrations were compared to the device’s accuracy during tests performed in the laboratory setting. RESULTS: Twenty-two women were enrolled in this study. In both the clinical and laboratory settings, the mean flow rate errors for the loading and maintenance dose infusions were under 2%. During 466 h of testing, the device sounded 129 occlusion alarms across 14 subjects. Of these, 71 alarms were false positives. CONCLUSION: Results of this study support the use of AutoSyp as a less painful and accurate means of MgSO(4) administration in clinical environments that lack infusion systems. There were a large number of false alarms in the current system which will be addressed in future designs. AutoSyp maintains the comfort of intravenous MgSO(4) administration, but unlike commercially available syringe pumps, it is capable of operating with a variety of syringe brands and sizes and requires no additional consumables. AutoSyp’s appropriate design will benefit its implementation and sustained use in low-resource settings. TRIAL REGISTRATION: Trial registered prospectively on November 18, 2014 with ClinicalTrials.gov (NCT02296931) BioMed Central 2017-06-19 /pmc/articles/PMC5474875/ /pubmed/28629437 http://dx.doi.org/10.1186/s12884-017-1382-9 Text en © The Author(s). 2017 Open AccessThis article is distributed under the terms of the Creative Commons Attribution 4.0 International License (http://creativecommons.org/licenses/by/4.0/), which permits unrestricted use, distribution, and reproduction in any medium, provided you give appropriate credit to the original author(s) and the source, provide a link to the Creative Commons license, and indicate if changes were made. The Creative Commons Public Domain Dedication waiver (http://creativecommons.org/publicdomain/zero/1.0/) applies to the data made available in this article, unless otherwise stated.
spellingShingle Research Article
Skerrett, Erica
Kommwa, Edward
Maynard, Kelley
Juarez, Alexa
Mataya, Ronald
Richards-Kortum, Rebecca
Oden, Z. Maria
Evaluation of a low-cost, low-power syringe pump to deliver magnesium sulfate intravenously to pre-eclamptic women in a Malawian referral hospital
title Evaluation of a low-cost, low-power syringe pump to deliver magnesium sulfate intravenously to pre-eclamptic women in a Malawian referral hospital
title_full Evaluation of a low-cost, low-power syringe pump to deliver magnesium sulfate intravenously to pre-eclamptic women in a Malawian referral hospital
title_fullStr Evaluation of a low-cost, low-power syringe pump to deliver magnesium sulfate intravenously to pre-eclamptic women in a Malawian referral hospital
title_full_unstemmed Evaluation of a low-cost, low-power syringe pump to deliver magnesium sulfate intravenously to pre-eclamptic women in a Malawian referral hospital
title_short Evaluation of a low-cost, low-power syringe pump to deliver magnesium sulfate intravenously to pre-eclamptic women in a Malawian referral hospital
title_sort evaluation of a low-cost, low-power syringe pump to deliver magnesium sulfate intravenously to pre-eclamptic women in a malawian referral hospital
topic Research Article
url https://www.ncbi.nlm.nih.gov/pmc/articles/PMC5474875/
https://www.ncbi.nlm.nih.gov/pubmed/28629437
http://dx.doi.org/10.1186/s12884-017-1382-9
work_keys_str_mv AT skerretterica evaluationofalowcostlowpowersyringepumptodelivermagnesiumsulfateintravenouslytopreeclampticwomeninamalawianreferralhospital
AT kommwaedward evaluationofalowcostlowpowersyringepumptodelivermagnesiumsulfateintravenouslytopreeclampticwomeninamalawianreferralhospital
AT maynardkelley evaluationofalowcostlowpowersyringepumptodelivermagnesiumsulfateintravenouslytopreeclampticwomeninamalawianreferralhospital
AT juarezalexa evaluationofalowcostlowpowersyringepumptodelivermagnesiumsulfateintravenouslytopreeclampticwomeninamalawianreferralhospital
AT matayaronald evaluationofalowcostlowpowersyringepumptodelivermagnesiumsulfateintravenouslytopreeclampticwomeninamalawianreferralhospital
AT richardskortumrebecca evaluationofalowcostlowpowersyringepumptodelivermagnesiumsulfateintravenouslytopreeclampticwomeninamalawianreferralhospital
AT odenzmaria evaluationofalowcostlowpowersyringepumptodelivermagnesiumsulfateintravenouslytopreeclampticwomeninamalawianreferralhospital