Breakage of a Circular Catheter Wedged in a Right Pulmonary Vein during Cryoballoon Pulmonary Vein Isolation
A 50-year-old man with paroxysmal atrial fibrillation underwent pulmonary vein isolation (PVI) using a cryoballoon. During the PVI procedure, a circular diagnostic catheter (CDC) was wedged deeply into the right PV (RPV). The lung tissues seemed to be tightening around the CDC. After various attempt...
Autores principales: | , , , |
---|---|
Formato: | Online Artículo Texto |
Lenguaje: | English |
Publicado: |
The Japanese Society of Internal Medicine
2017
|
Materias: | |
Acceso en línea: | https://www.ncbi.nlm.nih.gov/pmc/articles/PMC5478567/ https://www.ncbi.nlm.nih.gov/pubmed/28458312 http://dx.doi.org/10.2169/internalmedicine.56.7924 |
_version_ | 1783244973620068352 |
---|---|
author | Makimoto, Hisaki Kelm, Malte Shin, Dong-In Blockhaus, Christian |
author_facet | Makimoto, Hisaki Kelm, Malte Shin, Dong-In Blockhaus, Christian |
author_sort | Makimoto, Hisaki |
collection | PubMed |
description | A 50-year-old man with paroxysmal atrial fibrillation underwent pulmonary vein isolation (PVI) using a cryoballoon. During the PVI procedure, a circular diagnostic catheter (CDC) was wedged deeply into the right PV (RPV). The lung tissues seemed to be tightening around the CDC. After various attempts to resolve the situation, the shaft of the CDC was completely torn, and the circular part remained in the RPV. At a follow-up of three months, the patient showed neither atrial fibrillation events nor clinical symptoms. We herein report a broken CDC shaft during PVI with a cryoballoon, which has not yet been reported. |
format | Online Article Text |
id | pubmed-5478567 |
institution | National Center for Biotechnology Information |
language | English |
publishDate | 2017 |
publisher | The Japanese Society of Internal Medicine |
record_format | MEDLINE/PubMed |
spelling | pubmed-54785672017-06-21 Breakage of a Circular Catheter Wedged in a Right Pulmonary Vein during Cryoballoon Pulmonary Vein Isolation Makimoto, Hisaki Kelm, Malte Shin, Dong-In Blockhaus, Christian Intern Med Case Report A 50-year-old man with paroxysmal atrial fibrillation underwent pulmonary vein isolation (PVI) using a cryoballoon. During the PVI procedure, a circular diagnostic catheter (CDC) was wedged deeply into the right PV (RPV). The lung tissues seemed to be tightening around the CDC. After various attempts to resolve the situation, the shaft of the CDC was completely torn, and the circular part remained in the RPV. At a follow-up of three months, the patient showed neither atrial fibrillation events nor clinical symptoms. We herein report a broken CDC shaft during PVI with a cryoballoon, which has not yet been reported. The Japanese Society of Internal Medicine 2017-05-01 /pmc/articles/PMC5478567/ /pubmed/28458312 http://dx.doi.org/10.2169/internalmedicine.56.7924 Text en Copyright © 2017 by The Japanese Society of Internal Medicine https://creativecommons.org/licenses/by-nc-nd/4.0/ The Internal Medicine is an Open Access article distributed under the Creative Commons Attribution-NonCommercial-NoDerivatives 4.0 International License. To view the details of this license, please visit (https://creativecommons.org/licenses/by-nc-nd/4.0/). |
spellingShingle | Case Report Makimoto, Hisaki Kelm, Malte Shin, Dong-In Blockhaus, Christian Breakage of a Circular Catheter Wedged in a Right Pulmonary Vein during Cryoballoon Pulmonary Vein Isolation |
title | Breakage of a Circular Catheter Wedged in a Right Pulmonary Vein during Cryoballoon Pulmonary Vein Isolation |
title_full | Breakage of a Circular Catheter Wedged in a Right Pulmonary Vein during Cryoballoon Pulmonary Vein Isolation |
title_fullStr | Breakage of a Circular Catheter Wedged in a Right Pulmonary Vein during Cryoballoon Pulmonary Vein Isolation |
title_full_unstemmed | Breakage of a Circular Catheter Wedged in a Right Pulmonary Vein during Cryoballoon Pulmonary Vein Isolation |
title_short | Breakage of a Circular Catheter Wedged in a Right Pulmonary Vein during Cryoballoon Pulmonary Vein Isolation |
title_sort | breakage of a circular catheter wedged in a right pulmonary vein during cryoballoon pulmonary vein isolation |
topic | Case Report |
url | https://www.ncbi.nlm.nih.gov/pmc/articles/PMC5478567/ https://www.ncbi.nlm.nih.gov/pubmed/28458312 http://dx.doi.org/10.2169/internalmedicine.56.7924 |
work_keys_str_mv | AT makimotohisaki breakageofacircularcatheterwedgedinarightpulmonaryveinduringcryoballoonpulmonaryveinisolation AT kelmmalte breakageofacircularcatheterwedgedinarightpulmonaryveinduringcryoballoonpulmonaryveinisolation AT shindongin breakageofacircularcatheterwedgedinarightpulmonaryveinduringcryoballoonpulmonaryveinisolation AT blockhauschristian breakageofacircularcatheterwedgedinarightpulmonaryveinduringcryoballoonpulmonaryveinisolation |