Application of laboratory and portable attenuated total reflectance infrared spectroscopic approaches for rapid quantification of alpaca serum immunoglobulin G

The objective of this study was to develop and compare the performance of laboratory grade and portable attenuated total reflectance infrared (ATR-IR) spectroscopic approaches in combination with partial least squares regression (PLSR) for the rapid quantification of alpaca serum IgG concentration,...

Descripción completa

Detalles Bibliográficos
Autores principales: Elsohaby, Ibrahim, Burns, Jennifer B., Riley, Christopher B., Shaw, R. Anthony, McClure, J. Trenton
Formato: Online Artículo Texto
Lenguaje:English
Publicado: Public Library of Science 2017
Materias:
Acceso en línea:https://www.ncbi.nlm.nih.gov/pmc/articles/PMC5484476/
https://www.ncbi.nlm.nih.gov/pubmed/28651006
http://dx.doi.org/10.1371/journal.pone.0179644
_version_ 1783245887990923264
author Elsohaby, Ibrahim
Burns, Jennifer B.
Riley, Christopher B.
Shaw, R. Anthony
McClure, J. Trenton
author_facet Elsohaby, Ibrahim
Burns, Jennifer B.
Riley, Christopher B.
Shaw, R. Anthony
McClure, J. Trenton
author_sort Elsohaby, Ibrahim
collection PubMed
description The objective of this study was to develop and compare the performance of laboratory grade and portable attenuated total reflectance infrared (ATR-IR) spectroscopic approaches in combination with partial least squares regression (PLSR) for the rapid quantification of alpaca serum IgG concentration, and the identification of low IgG (<1000 mg/dL), which is consistent with the diagnosis of failure of transfer of passive immunity (FTPI) in neonates. Serum samples (n = 175) collected from privately owned, healthy alpacas were tested by the reference method of radial immunodiffusion (RID) assay, and laboratory grade and portable ATR-IR spectrometers. Various pre-processing strategies were applied to the ATR-IR spectra that were linked to corresponding RID-IgG concentrations, and then randomly split into two sets: calibration (training) and test sets. PLSR was applied to the calibration set and calibration models were developed, and the test set was used to assess the accuracy of the analytical method. For the test set, the Pearson correlation coefficients between the IgG measured by RID and predicted by both laboratory grade and portable ATR-IR spectrometers was 0.91. The average differences between reference serum IgG concentrations and the two IR-based methods were 120.5 mg/dL and 71 mg/dL for the laboratory and portable ATR-IR-based assays, respectively. Adopting an IgG concentration <1000 mg/dL as the cut-point for FTPI cases, the sensitivity, specificity, and accuracy for identifying serum samples below this cut point by laboratory ATR-IR assay were 86, 100 and 98%, respectively (within the entire data set). Corresponding values for the portable ATR-IR assay were 95, 99 and 99%, respectively. These results suggest that the two different ATR-IR assays performed similarly for rapid qualitative evaluation of alpaca serum IgG and for diagnosis of IgG <1000 mg/dL, the portable ATR-IR spectrometer performed slightly better, and provides more flexibility for potential application in the field.
format Online
Article
Text
id pubmed-5484476
institution National Center for Biotechnology Information
language English
publishDate 2017
publisher Public Library of Science
record_format MEDLINE/PubMed
spelling pubmed-54844762017-07-11 Application of laboratory and portable attenuated total reflectance infrared spectroscopic approaches for rapid quantification of alpaca serum immunoglobulin G Elsohaby, Ibrahim Burns, Jennifer B. Riley, Christopher B. Shaw, R. Anthony McClure, J. Trenton PLoS One Research Article The objective of this study was to develop and compare the performance of laboratory grade and portable attenuated total reflectance infrared (ATR-IR) spectroscopic approaches in combination with partial least squares regression (PLSR) for the rapid quantification of alpaca serum IgG concentration, and the identification of low IgG (<1000 mg/dL), which is consistent with the diagnosis of failure of transfer of passive immunity (FTPI) in neonates. Serum samples (n = 175) collected from privately owned, healthy alpacas were tested by the reference method of radial immunodiffusion (RID) assay, and laboratory grade and portable ATR-IR spectrometers. Various pre-processing strategies were applied to the ATR-IR spectra that were linked to corresponding RID-IgG concentrations, and then randomly split into two sets: calibration (training) and test sets. PLSR was applied to the calibration set and calibration models were developed, and the test set was used to assess the accuracy of the analytical method. For the test set, the Pearson correlation coefficients between the IgG measured by RID and predicted by both laboratory grade and portable ATR-IR spectrometers was 0.91. The average differences between reference serum IgG concentrations and the two IR-based methods were 120.5 mg/dL and 71 mg/dL for the laboratory and portable ATR-IR-based assays, respectively. Adopting an IgG concentration <1000 mg/dL as the cut-point for FTPI cases, the sensitivity, specificity, and accuracy for identifying serum samples below this cut point by laboratory ATR-IR assay were 86, 100 and 98%, respectively (within the entire data set). Corresponding values for the portable ATR-IR assay were 95, 99 and 99%, respectively. These results suggest that the two different ATR-IR assays performed similarly for rapid qualitative evaluation of alpaca serum IgG and for diagnosis of IgG <1000 mg/dL, the portable ATR-IR spectrometer performed slightly better, and provides more flexibility for potential application in the field. Public Library of Science 2017-06-26 /pmc/articles/PMC5484476/ /pubmed/28651006 http://dx.doi.org/10.1371/journal.pone.0179644 Text en © 2017 Elsohaby et al http://creativecommons.org/licenses/by/4.0/ This is an open access article distributed under the terms of the Creative Commons Attribution License (http://creativecommons.org/licenses/by/4.0/) , which permits unrestricted use, distribution, and reproduction in any medium, provided the original author and source are credited.
spellingShingle Research Article
Elsohaby, Ibrahim
Burns, Jennifer B.
Riley, Christopher B.
Shaw, R. Anthony
McClure, J. Trenton
Application of laboratory and portable attenuated total reflectance infrared spectroscopic approaches for rapid quantification of alpaca serum immunoglobulin G
title Application of laboratory and portable attenuated total reflectance infrared spectroscopic approaches for rapid quantification of alpaca serum immunoglobulin G
title_full Application of laboratory and portable attenuated total reflectance infrared spectroscopic approaches for rapid quantification of alpaca serum immunoglobulin G
title_fullStr Application of laboratory and portable attenuated total reflectance infrared spectroscopic approaches for rapid quantification of alpaca serum immunoglobulin G
title_full_unstemmed Application of laboratory and portable attenuated total reflectance infrared spectroscopic approaches for rapid quantification of alpaca serum immunoglobulin G
title_short Application of laboratory and portable attenuated total reflectance infrared spectroscopic approaches for rapid quantification of alpaca serum immunoglobulin G
title_sort application of laboratory and portable attenuated total reflectance infrared spectroscopic approaches for rapid quantification of alpaca serum immunoglobulin g
topic Research Article
url https://www.ncbi.nlm.nih.gov/pmc/articles/PMC5484476/
https://www.ncbi.nlm.nih.gov/pubmed/28651006
http://dx.doi.org/10.1371/journal.pone.0179644
work_keys_str_mv AT elsohabyibrahim applicationoflaboratoryandportableattenuatedtotalreflectanceinfraredspectroscopicapproachesforrapidquantificationofalpacaserumimmunoglobuling
AT burnsjenniferb applicationoflaboratoryandportableattenuatedtotalreflectanceinfraredspectroscopicapproachesforrapidquantificationofalpacaserumimmunoglobuling
AT rileychristopherb applicationoflaboratoryandportableattenuatedtotalreflectanceinfraredspectroscopicapproachesforrapidquantificationofalpacaserumimmunoglobuling
AT shawranthony applicationoflaboratoryandportableattenuatedtotalreflectanceinfraredspectroscopicapproachesforrapidquantificationofalpacaserumimmunoglobuling
AT mcclurejtrenton applicationoflaboratoryandportableattenuatedtotalreflectanceinfraredspectroscopicapproachesforrapidquantificationofalpacaserumimmunoglobuling