The Influence of Acquisition Mode on the Dosimetric Performance of an Amorphous Silicon Electronic Portal Imaging Device

AIMS: This study investigates the impact of cine acquisition mode on the dosimetric characteristics of a Varian aS500 amorphous silicon electronic portal imaging device (a-Si EPID). MATERIALS AND METHODS: The performance of an a-Si EPID operated in cine mode was assessed and compared to its performa...

Descripción completa

Detalles Bibliográficos
Autores principales: Bawazeer, Omemh, Herath, Sisira, Sarasanandarajah, Siva, Kron, Tomas, Deb, Pradip
Formato: Online Artículo Texto
Lenguaje:English
Publicado: Medknow Publications & Media Pvt Ltd 2017
Materias:
Acceso en línea:https://www.ncbi.nlm.nih.gov/pmc/articles/PMC5496276/
https://www.ncbi.nlm.nih.gov/pubmed/28706355
http://dx.doi.org/10.4103/jmp.JMP_98_16
_version_ 1783247942656720896
author Bawazeer, Omemh
Herath, Sisira
Sarasanandarajah, Siva
Kron, Tomas
Deb, Pradip
author_facet Bawazeer, Omemh
Herath, Sisira
Sarasanandarajah, Siva
Kron, Tomas
Deb, Pradip
author_sort Bawazeer, Omemh
collection PubMed
description AIMS: This study investigates the impact of cine acquisition mode on the dosimetric characteristics of a Varian aS500 amorphous silicon electronic portal imaging device (a-Si EPID). MATERIALS AND METHODS: The performance of an a-Si EPID operated in cine mode was assessed and compared to its performance when operated in an integrated mode and dose measurements using an ionization chamber. This study was conducted at different photon energies and the EPID performance was assessed as function of the delivered dose, dose rate, multileaf collimator speed, field size, phantom thickness, and intensity-modulated radiation therapy fields. RESULTS: The worst nonlinearity was observed at low monitor unit (MU) settings < 100 MU with the highest dose per frame. The nonlinearity of response at a low MU setting was attributed due to the loss of four cine images during each delivery. The EPID response with changing dose rate for 10 MU delivered had similar results to its performance in an integrated mode and ionization chamber. Despite the nonlinearity of response with low MU delivered, EPID performance operated in cine and integrated acquisition modes had comparable responses within 2%. CONCLUSIONS: For EPID dosimetry application using cine mode, this study recommends the calibration of the EPID images to be undertaken at a large MU. There were no additional corrections that were required when the EPID operated in cine acquisition mode as compared to calibration in integrated mode.
format Online
Article
Text
id pubmed-5496276
institution National Center for Biotechnology Information
language English
publishDate 2017
publisher Medknow Publications & Media Pvt Ltd
record_format MEDLINE/PubMed
spelling pubmed-54962762017-07-13 The Influence of Acquisition Mode on the Dosimetric Performance of an Amorphous Silicon Electronic Portal Imaging Device Bawazeer, Omemh Herath, Sisira Sarasanandarajah, Siva Kron, Tomas Deb, Pradip J Med Phys Technical Note AIMS: This study investigates the impact of cine acquisition mode on the dosimetric characteristics of a Varian aS500 amorphous silicon electronic portal imaging device (a-Si EPID). MATERIALS AND METHODS: The performance of an a-Si EPID operated in cine mode was assessed and compared to its performance when operated in an integrated mode and dose measurements using an ionization chamber. This study was conducted at different photon energies and the EPID performance was assessed as function of the delivered dose, dose rate, multileaf collimator speed, field size, phantom thickness, and intensity-modulated radiation therapy fields. RESULTS: The worst nonlinearity was observed at low monitor unit (MU) settings < 100 MU with the highest dose per frame. The nonlinearity of response at a low MU setting was attributed due to the loss of four cine images during each delivery. The EPID response with changing dose rate for 10 MU delivered had similar results to its performance in an integrated mode and ionization chamber. Despite the nonlinearity of response with low MU delivered, EPID performance operated in cine and integrated acquisition modes had comparable responses within 2%. CONCLUSIONS: For EPID dosimetry application using cine mode, this study recommends the calibration of the EPID images to be undertaken at a large MU. There were no additional corrections that were required when the EPID operated in cine acquisition mode as compared to calibration in integrated mode. Medknow Publications & Media Pvt Ltd 2017 /pmc/articles/PMC5496276/ /pubmed/28706355 http://dx.doi.org/10.4103/jmp.JMP_98_16 Text en Copyright: © 2017 Journal of Medical Physics http://creativecommons.org/licenses/by-nc-sa/3.0 This is an open access article distributed under the terms of the Creative Commons Attribution-NonCommercial-ShareAlike 3.0 License, which allows others to remix, tweak, and build upon the work non-commercially, as long as the author is credited and the new creations are licensed under the identical terms.
spellingShingle Technical Note
Bawazeer, Omemh
Herath, Sisira
Sarasanandarajah, Siva
Kron, Tomas
Deb, Pradip
The Influence of Acquisition Mode on the Dosimetric Performance of an Amorphous Silicon Electronic Portal Imaging Device
title The Influence of Acquisition Mode on the Dosimetric Performance of an Amorphous Silicon Electronic Portal Imaging Device
title_full The Influence of Acquisition Mode on the Dosimetric Performance of an Amorphous Silicon Electronic Portal Imaging Device
title_fullStr The Influence of Acquisition Mode on the Dosimetric Performance of an Amorphous Silicon Electronic Portal Imaging Device
title_full_unstemmed The Influence of Acquisition Mode on the Dosimetric Performance of an Amorphous Silicon Electronic Portal Imaging Device
title_short The Influence of Acquisition Mode on the Dosimetric Performance of an Amorphous Silicon Electronic Portal Imaging Device
title_sort influence of acquisition mode on the dosimetric performance of an amorphous silicon electronic portal imaging device
topic Technical Note
url https://www.ncbi.nlm.nih.gov/pmc/articles/PMC5496276/
https://www.ncbi.nlm.nih.gov/pubmed/28706355
http://dx.doi.org/10.4103/jmp.JMP_98_16
work_keys_str_mv AT bawazeeromemh theinfluenceofacquisitionmodeonthedosimetricperformanceofanamorphoussiliconelectronicportalimagingdevice
AT herathsisira theinfluenceofacquisitionmodeonthedosimetricperformanceofanamorphoussiliconelectronicportalimagingdevice
AT sarasanandarajahsiva theinfluenceofacquisitionmodeonthedosimetricperformanceofanamorphoussiliconelectronicportalimagingdevice
AT krontomas theinfluenceofacquisitionmodeonthedosimetricperformanceofanamorphoussiliconelectronicportalimagingdevice
AT debpradip theinfluenceofacquisitionmodeonthedosimetricperformanceofanamorphoussiliconelectronicportalimagingdevice
AT bawazeeromemh influenceofacquisitionmodeonthedosimetricperformanceofanamorphoussiliconelectronicportalimagingdevice
AT herathsisira influenceofacquisitionmodeonthedosimetricperformanceofanamorphoussiliconelectronicportalimagingdevice
AT sarasanandarajahsiva influenceofacquisitionmodeonthedosimetricperformanceofanamorphoussiliconelectronicportalimagingdevice
AT krontomas influenceofacquisitionmodeonthedosimetricperformanceofanamorphoussiliconelectronicportalimagingdevice
AT debpradip influenceofacquisitionmodeonthedosimetricperformanceofanamorphoussiliconelectronicportalimagingdevice