The Influence of Acquisition Mode on the Dosimetric Performance of an Amorphous Silicon Electronic Portal Imaging Device
AIMS: This study investigates the impact of cine acquisition mode on the dosimetric characteristics of a Varian aS500 amorphous silicon electronic portal imaging device (a-Si EPID). MATERIALS AND METHODS: The performance of an a-Si EPID operated in cine mode was assessed and compared to its performa...
Autores principales: | , , , , |
---|---|
Formato: | Online Artículo Texto |
Lenguaje: | English |
Publicado: |
Medknow Publications & Media Pvt Ltd
2017
|
Materias: | |
Acceso en línea: | https://www.ncbi.nlm.nih.gov/pmc/articles/PMC5496276/ https://www.ncbi.nlm.nih.gov/pubmed/28706355 http://dx.doi.org/10.4103/jmp.JMP_98_16 |
_version_ | 1783247942656720896 |
---|---|
author | Bawazeer, Omemh Herath, Sisira Sarasanandarajah, Siva Kron, Tomas Deb, Pradip |
author_facet | Bawazeer, Omemh Herath, Sisira Sarasanandarajah, Siva Kron, Tomas Deb, Pradip |
author_sort | Bawazeer, Omemh |
collection | PubMed |
description | AIMS: This study investigates the impact of cine acquisition mode on the dosimetric characteristics of a Varian aS500 amorphous silicon electronic portal imaging device (a-Si EPID). MATERIALS AND METHODS: The performance of an a-Si EPID operated in cine mode was assessed and compared to its performance when operated in an integrated mode and dose measurements using an ionization chamber. This study was conducted at different photon energies and the EPID performance was assessed as function of the delivered dose, dose rate, multileaf collimator speed, field size, phantom thickness, and intensity-modulated radiation therapy fields. RESULTS: The worst nonlinearity was observed at low monitor unit (MU) settings < 100 MU with the highest dose per frame. The nonlinearity of response at a low MU setting was attributed due to the loss of four cine images during each delivery. The EPID response with changing dose rate for 10 MU delivered had similar results to its performance in an integrated mode and ionization chamber. Despite the nonlinearity of response with low MU delivered, EPID performance operated in cine and integrated acquisition modes had comparable responses within 2%. CONCLUSIONS: For EPID dosimetry application using cine mode, this study recommends the calibration of the EPID images to be undertaken at a large MU. There were no additional corrections that were required when the EPID operated in cine acquisition mode as compared to calibration in integrated mode. |
format | Online Article Text |
id | pubmed-5496276 |
institution | National Center for Biotechnology Information |
language | English |
publishDate | 2017 |
publisher | Medknow Publications & Media Pvt Ltd |
record_format | MEDLINE/PubMed |
spelling | pubmed-54962762017-07-13 The Influence of Acquisition Mode on the Dosimetric Performance of an Amorphous Silicon Electronic Portal Imaging Device Bawazeer, Omemh Herath, Sisira Sarasanandarajah, Siva Kron, Tomas Deb, Pradip J Med Phys Technical Note AIMS: This study investigates the impact of cine acquisition mode on the dosimetric characteristics of a Varian aS500 amorphous silicon electronic portal imaging device (a-Si EPID). MATERIALS AND METHODS: The performance of an a-Si EPID operated in cine mode was assessed and compared to its performance when operated in an integrated mode and dose measurements using an ionization chamber. This study was conducted at different photon energies and the EPID performance was assessed as function of the delivered dose, dose rate, multileaf collimator speed, field size, phantom thickness, and intensity-modulated radiation therapy fields. RESULTS: The worst nonlinearity was observed at low monitor unit (MU) settings < 100 MU with the highest dose per frame. The nonlinearity of response at a low MU setting was attributed due to the loss of four cine images during each delivery. The EPID response with changing dose rate for 10 MU delivered had similar results to its performance in an integrated mode and ionization chamber. Despite the nonlinearity of response with low MU delivered, EPID performance operated in cine and integrated acquisition modes had comparable responses within 2%. CONCLUSIONS: For EPID dosimetry application using cine mode, this study recommends the calibration of the EPID images to be undertaken at a large MU. There were no additional corrections that were required when the EPID operated in cine acquisition mode as compared to calibration in integrated mode. Medknow Publications & Media Pvt Ltd 2017 /pmc/articles/PMC5496276/ /pubmed/28706355 http://dx.doi.org/10.4103/jmp.JMP_98_16 Text en Copyright: © 2017 Journal of Medical Physics http://creativecommons.org/licenses/by-nc-sa/3.0 This is an open access article distributed under the terms of the Creative Commons Attribution-NonCommercial-ShareAlike 3.0 License, which allows others to remix, tweak, and build upon the work non-commercially, as long as the author is credited and the new creations are licensed under the identical terms. |
spellingShingle | Technical Note Bawazeer, Omemh Herath, Sisira Sarasanandarajah, Siva Kron, Tomas Deb, Pradip The Influence of Acquisition Mode on the Dosimetric Performance of an Amorphous Silicon Electronic Portal Imaging Device |
title | The Influence of Acquisition Mode on the Dosimetric Performance of an Amorphous Silicon Electronic Portal Imaging Device |
title_full | The Influence of Acquisition Mode on the Dosimetric Performance of an Amorphous Silicon Electronic Portal Imaging Device |
title_fullStr | The Influence of Acquisition Mode on the Dosimetric Performance of an Amorphous Silicon Electronic Portal Imaging Device |
title_full_unstemmed | The Influence of Acquisition Mode on the Dosimetric Performance of an Amorphous Silicon Electronic Portal Imaging Device |
title_short | The Influence of Acquisition Mode on the Dosimetric Performance of an Amorphous Silicon Electronic Portal Imaging Device |
title_sort | influence of acquisition mode on the dosimetric performance of an amorphous silicon electronic portal imaging device |
topic | Technical Note |
url | https://www.ncbi.nlm.nih.gov/pmc/articles/PMC5496276/ https://www.ncbi.nlm.nih.gov/pubmed/28706355 http://dx.doi.org/10.4103/jmp.JMP_98_16 |
work_keys_str_mv | AT bawazeeromemh theinfluenceofacquisitionmodeonthedosimetricperformanceofanamorphoussiliconelectronicportalimagingdevice AT herathsisira theinfluenceofacquisitionmodeonthedosimetricperformanceofanamorphoussiliconelectronicportalimagingdevice AT sarasanandarajahsiva theinfluenceofacquisitionmodeonthedosimetricperformanceofanamorphoussiliconelectronicportalimagingdevice AT krontomas theinfluenceofacquisitionmodeonthedosimetricperformanceofanamorphoussiliconelectronicportalimagingdevice AT debpradip theinfluenceofacquisitionmodeonthedosimetricperformanceofanamorphoussiliconelectronicportalimagingdevice AT bawazeeromemh influenceofacquisitionmodeonthedosimetricperformanceofanamorphoussiliconelectronicportalimagingdevice AT herathsisira influenceofacquisitionmodeonthedosimetricperformanceofanamorphoussiliconelectronicportalimagingdevice AT sarasanandarajahsiva influenceofacquisitionmodeonthedosimetricperformanceofanamorphoussiliconelectronicportalimagingdevice AT krontomas influenceofacquisitionmodeonthedosimetricperformanceofanamorphoussiliconelectronicportalimagingdevice AT debpradip influenceofacquisitionmodeonthedosimetricperformanceofanamorphoussiliconelectronicportalimagingdevice |