Impact of cuff positioning on blood pressure measurement accuracy: may a specially designed cuff make a difference?
During blood pressure (BP) measurement, the recommended positioning of the cuff bladder center is directly above the brachial artery. We investigated the relevance of incorrect cuff positioning during (1) auscultatory measurement with an appropriate or improperly small cuff and (2) oscillometric mea...
Autores principales: | , , , , , , , , , |
---|---|
Formato: | Online Artículo Texto |
Lenguaje: | English |
Publicado: |
Nature Publishing Group
2017
|
Materias: | |
Acceso en línea: | https://www.ncbi.nlm.nih.gov/pmc/articles/PMC5506235/ https://www.ncbi.nlm.nih.gov/pubmed/28077860 http://dx.doi.org/10.1038/hr.2016.184 |
_version_ | 1783249540212588544 |
---|---|
author | Bilo, Grzegorz Sala, Oscar Perego, Carlotta Faini, Andrea Gao, Lan Głuszewska, Anna Ochoa, Juan Eugenio Pellegrini, Dario Lonati, Laura Maria Parati, Gianfranco |
author_facet | Bilo, Grzegorz Sala, Oscar Perego, Carlotta Faini, Andrea Gao, Lan Głuszewska, Anna Ochoa, Juan Eugenio Pellegrini, Dario Lonati, Laura Maria Parati, Gianfranco |
author_sort | Bilo, Grzegorz |
collection | PubMed |
description | During blood pressure (BP) measurement, the recommended positioning of the cuff bladder center is directly above the brachial artery. We investigated the relevance of incorrect cuff positioning during (1) auscultatory measurement with an appropriate or improperly small cuff and (2) oscillometric measurement with a wide-range cuff designed to guarantee accurate measurements regardless of position. In subjects with wide BP and arm circumference ranges, (1) auscultatory BP was repeatedly measured with a properly positioned cuff (reference) and, simultaneously, with an identical cuff placed on the other arm in either a correct or an incorrect position (test). The measurements were performed with a properly sized (N=57) or an improperly small cuff (N=33). (2) Auscultatory measurements obtained with a properly positioned and sized cuff were compared with oscillometric measurements obtained with a specially designed wide-range cuff (Omron IntelliWrap) placed on the contralateral arm either in a correct or an incorrect position. Auscultatory BP measures were unaffected by incorrect positioning of a properly sized cuff, whereas with undercuffing, BP was overestimated with the cuff displaced by 90° laterally (systolic/diastolic BP differences: 4.9±4.6/4.0±4.6 mm Hg, P<0.01) or by 180° (3.9±5.4/4.2±5.1 mm Hg, P<0.01) in relation to the correct position. Incorrect placement of the oscillometric cuff had no significant effect on the accuracy of the measurements (difference with correct position <1.5 mm Hg). Incorrect cuff positioning introduces a systematic overestimation of auscultatory BP when the cuff is too small in relation to arm circumference but not when it is correctly sized. No systematic error was observed with oscillometric measurements obtained with a specially designed wide-range cuff. |
format | Online Article Text |
id | pubmed-5506235 |
institution | National Center for Biotechnology Information |
language | English |
publishDate | 2017 |
publisher | Nature Publishing Group |
record_format | MEDLINE/PubMed |
spelling | pubmed-55062352017-07-17 Impact of cuff positioning on blood pressure measurement accuracy: may a specially designed cuff make a difference? Bilo, Grzegorz Sala, Oscar Perego, Carlotta Faini, Andrea Gao, Lan Głuszewska, Anna Ochoa, Juan Eugenio Pellegrini, Dario Lonati, Laura Maria Parati, Gianfranco Hypertens Res Original Article During blood pressure (BP) measurement, the recommended positioning of the cuff bladder center is directly above the brachial artery. We investigated the relevance of incorrect cuff positioning during (1) auscultatory measurement with an appropriate or improperly small cuff and (2) oscillometric measurement with a wide-range cuff designed to guarantee accurate measurements regardless of position. In subjects with wide BP and arm circumference ranges, (1) auscultatory BP was repeatedly measured with a properly positioned cuff (reference) and, simultaneously, with an identical cuff placed on the other arm in either a correct or an incorrect position (test). The measurements were performed with a properly sized (N=57) or an improperly small cuff (N=33). (2) Auscultatory measurements obtained with a properly positioned and sized cuff were compared with oscillometric measurements obtained with a specially designed wide-range cuff (Omron IntelliWrap) placed on the contralateral arm either in a correct or an incorrect position. Auscultatory BP measures were unaffected by incorrect positioning of a properly sized cuff, whereas with undercuffing, BP was overestimated with the cuff displaced by 90° laterally (systolic/diastolic BP differences: 4.9±4.6/4.0±4.6 mm Hg, P<0.01) or by 180° (3.9±5.4/4.2±5.1 mm Hg, P<0.01) in relation to the correct position. Incorrect placement of the oscillometric cuff had no significant effect on the accuracy of the measurements (difference with correct position <1.5 mm Hg). Incorrect cuff positioning introduces a systematic overestimation of auscultatory BP when the cuff is too small in relation to arm circumference but not when it is correctly sized. No systematic error was observed with oscillometric measurements obtained with a specially designed wide-range cuff. Nature Publishing Group 2017-06 2017-01-12 /pmc/articles/PMC5506235/ /pubmed/28077860 http://dx.doi.org/10.1038/hr.2016.184 Text en Copyright © 2017 The Author(s) http://creativecommons.org/licenses/by-nc-sa/4.0/ This work is licensed under a Creative Commons Attribution-NonCommercial-ShareAlike 4.0 International License. The images or other third party material in this article are included in the article’s Creative Commons license, unless indicated otherwise in the credit line; if the material is not included under the Creative Commons license, users will need to obtain permission from the license holder to reproduce the material. To view a copy of this license, visit http://creativecommons.org/licenses/by-nc-sa/4.0/ |
spellingShingle | Original Article Bilo, Grzegorz Sala, Oscar Perego, Carlotta Faini, Andrea Gao, Lan Głuszewska, Anna Ochoa, Juan Eugenio Pellegrini, Dario Lonati, Laura Maria Parati, Gianfranco Impact of cuff positioning on blood pressure measurement accuracy: may a specially designed cuff make a difference? |
title | Impact of cuff positioning on blood pressure measurement accuracy: may a specially designed cuff make a difference? |
title_full | Impact of cuff positioning on blood pressure measurement accuracy: may a specially designed cuff make a difference? |
title_fullStr | Impact of cuff positioning on blood pressure measurement accuracy: may a specially designed cuff make a difference? |
title_full_unstemmed | Impact of cuff positioning on blood pressure measurement accuracy: may a specially designed cuff make a difference? |
title_short | Impact of cuff positioning on blood pressure measurement accuracy: may a specially designed cuff make a difference? |
title_sort | impact of cuff positioning on blood pressure measurement accuracy: may a specially designed cuff make a difference? |
topic | Original Article |
url | https://www.ncbi.nlm.nih.gov/pmc/articles/PMC5506235/ https://www.ncbi.nlm.nih.gov/pubmed/28077860 http://dx.doi.org/10.1038/hr.2016.184 |
work_keys_str_mv | AT bilogrzegorz impactofcuffpositioningonbloodpressuremeasurementaccuracymayaspeciallydesignedcuffmakeadifference AT salaoscar impactofcuffpositioningonbloodpressuremeasurementaccuracymayaspeciallydesignedcuffmakeadifference AT peregocarlotta impactofcuffpositioningonbloodpressuremeasurementaccuracymayaspeciallydesignedcuffmakeadifference AT fainiandrea impactofcuffpositioningonbloodpressuremeasurementaccuracymayaspeciallydesignedcuffmakeadifference AT gaolan impactofcuffpositioningonbloodpressuremeasurementaccuracymayaspeciallydesignedcuffmakeadifference AT głuszewskaanna impactofcuffpositioningonbloodpressuremeasurementaccuracymayaspeciallydesignedcuffmakeadifference AT ochoajuaneugenio impactofcuffpositioningonbloodpressuremeasurementaccuracymayaspeciallydesignedcuffmakeadifference AT pellegrinidario impactofcuffpositioningonbloodpressuremeasurementaccuracymayaspeciallydesignedcuffmakeadifference AT lonatilauramaria impactofcuffpositioningonbloodpressuremeasurementaccuracymayaspeciallydesignedcuffmakeadifference AT paratigianfranco impactofcuffpositioningonbloodpressuremeasurementaccuracymayaspeciallydesignedcuffmakeadifference |