Comparison of central corneal thickness with four noncontact devices: An agreement analysis of swept-source technology
PURPOSE: The purpose of this study was to compare the central corneal thickness (CCT) measurements of four noncontact devices in healthy eyes. MATERIALS AND METHODS: In a sample of 45 healthy controls, CCT was measured using an optical biometer (IOLMaster 700) based on swept-source optical coherence...
Autores principales: | , |
---|---|
Formato: | Online Artículo Texto |
Lenguaje: | English |
Publicado: |
Medknow Publications & Media Pvt Ltd
2017
|
Materias: | |
Acceso en línea: | https://www.ncbi.nlm.nih.gov/pmc/articles/PMC5508455/ https://www.ncbi.nlm.nih.gov/pubmed/28643709 http://dx.doi.org/10.4103/ijo.IJO_618_16 |
Sumario: | PURPOSE: The purpose of this study was to compare the central corneal thickness (CCT) measurements of four noncontact devices in healthy eyes. MATERIALS AND METHODS: In a sample of 45 healthy controls, CCT was measured using an optical biometer (IOLMaster 700) based on swept-source optical coherence tomography (SS-OCT), high-resolution rotating Scheimpflug camera system (Pentacam HR), spectral-domain OCT (SD-OCT) device with an anterior segment module (Spectralis), and noncontact pachymetry (NCP) device (Topcon TRK-2P). Agreement among the devices was analyzed using mean differences (i.e., bias) and Bland–Altman analysis with 95% limits of agreement (LoA). RESULTS: Mean CCT measurements were 537.5 ± 47.5 μm for SS-OCT optical biometer, 532.3 ± 43.8 μm for Scheimpflug system, 521.3 ± 44.7 μm for SD-OCT device, and 518.0 ± 43.1 μm for NCP (P < 0.001). The SD-OCT device and NCP showed the closest agreement, with a bias of 2.6 μm (95% LoA, −3.6–8.8 μm), whereas the SS-OCT optical biometer and NCP showed the least agreement, with a bias of 18.7 μm (95% LoA, −2.1–39.5 μm). Bias was 16.1 μm (95% LoA, −3.1–35.3 μm) for SS-OCT optical biometer and SD-OCT, 5.1 μm (95% LoA, −6.8–17.0 μm) for SS-OCT optical biometer and Scheimpflug system, 10.9 μm (95% LoA, −15.1–36.9 μm) for SD-OCT device and Scheimpflug system, and 13.6 μm (95% LoA, −5–32.2 μm) for Scheimpflug system and NCP. CONCLUSIONS: SS-OCT optical biometer overestimates CCT measurements compared to Scheimpflug system, SD-OCT device, and NCP. Given mean differences and range variations in CCT measurements between devices, SS-OCT optical biometer and Scheimpflug system are interchangeable as are SD-OCT and NCP. |
---|