Comparison of central corneal thickness with four noncontact devices: An agreement analysis of swept-source technology

PURPOSE: The purpose of this study was to compare the central corneal thickness (CCT) measurements of four noncontact devices in healthy eyes. MATERIALS AND METHODS: In a sample of 45 healthy controls, CCT was measured using an optical biometer (IOLMaster 700) based on swept-source optical coherence...

Descripción completa

Detalles Bibliográficos
Autores principales: Özyol, Erhan, Özyol, Pelin
Formato: Online Artículo Texto
Lenguaje:English
Publicado: Medknow Publications & Media Pvt Ltd 2017
Materias:
Acceso en línea:https://www.ncbi.nlm.nih.gov/pmc/articles/PMC5508455/
https://www.ncbi.nlm.nih.gov/pubmed/28643709
http://dx.doi.org/10.4103/ijo.IJO_618_16
_version_ 1783249884998008832
author Özyol, Erhan
Özyol, Pelin
author_facet Özyol, Erhan
Özyol, Pelin
author_sort Özyol, Erhan
collection PubMed
description PURPOSE: The purpose of this study was to compare the central corneal thickness (CCT) measurements of four noncontact devices in healthy eyes. MATERIALS AND METHODS: In a sample of 45 healthy controls, CCT was measured using an optical biometer (IOLMaster 700) based on swept-source optical coherence tomography (SS-OCT), high-resolution rotating Scheimpflug camera system (Pentacam HR), spectral-domain OCT (SD-OCT) device with an anterior segment module (Spectralis), and noncontact pachymetry (NCP) device (Topcon TRK-2P). Agreement among the devices was analyzed using mean differences (i.e., bias) and Bland–Altman analysis with 95% limits of agreement (LoA). RESULTS: Mean CCT measurements were 537.5 ± 47.5 μm for SS-OCT optical biometer, 532.3 ± 43.8 μm for Scheimpflug system, 521.3 ± 44.7 μm for SD-OCT device, and 518.0 ± 43.1 μm for NCP (P < 0.001). The SD-OCT device and NCP showed the closest agreement, with a bias of 2.6 μm (95% LoA, −3.6–8.8 μm), whereas the SS-OCT optical biometer and NCP showed the least agreement, with a bias of 18.7 μm (95% LoA, −2.1–39.5 μm). Bias was 16.1 μm (95% LoA, −3.1–35.3 μm) for SS-OCT optical biometer and SD-OCT, 5.1 μm (95% LoA, −6.8–17.0 μm) for SS-OCT optical biometer and Scheimpflug system, 10.9 μm (95% LoA, −15.1–36.9 μm) for SD-OCT device and Scheimpflug system, and 13.6 μm (95% LoA, −5–32.2 μm) for Scheimpflug system and NCP. CONCLUSIONS: SS-OCT optical biometer overestimates CCT measurements compared to Scheimpflug system, SD-OCT device, and NCP. Given mean differences and range variations in CCT measurements between devices, SS-OCT optical biometer and Scheimpflug system are interchangeable as are SD-OCT and NCP.
format Online
Article
Text
id pubmed-5508455
institution National Center for Biotechnology Information
language English
publishDate 2017
publisher Medknow Publications & Media Pvt Ltd
record_format MEDLINE/PubMed
spelling pubmed-55084552017-07-17 Comparison of central corneal thickness with four noncontact devices: An agreement analysis of swept-source technology Özyol, Erhan Özyol, Pelin Indian J Ophthalmol Original Article PURPOSE: The purpose of this study was to compare the central corneal thickness (CCT) measurements of four noncontact devices in healthy eyes. MATERIALS AND METHODS: In a sample of 45 healthy controls, CCT was measured using an optical biometer (IOLMaster 700) based on swept-source optical coherence tomography (SS-OCT), high-resolution rotating Scheimpflug camera system (Pentacam HR), spectral-domain OCT (SD-OCT) device with an anterior segment module (Spectralis), and noncontact pachymetry (NCP) device (Topcon TRK-2P). Agreement among the devices was analyzed using mean differences (i.e., bias) and Bland–Altman analysis with 95% limits of agreement (LoA). RESULTS: Mean CCT measurements were 537.5 ± 47.5 μm for SS-OCT optical biometer, 532.3 ± 43.8 μm for Scheimpflug system, 521.3 ± 44.7 μm for SD-OCT device, and 518.0 ± 43.1 μm for NCP (P < 0.001). The SD-OCT device and NCP showed the closest agreement, with a bias of 2.6 μm (95% LoA, −3.6–8.8 μm), whereas the SS-OCT optical biometer and NCP showed the least agreement, with a bias of 18.7 μm (95% LoA, −2.1–39.5 μm). Bias was 16.1 μm (95% LoA, −3.1–35.3 μm) for SS-OCT optical biometer and SD-OCT, 5.1 μm (95% LoA, −6.8–17.0 μm) for SS-OCT optical biometer and Scheimpflug system, 10.9 μm (95% LoA, −15.1–36.9 μm) for SD-OCT device and Scheimpflug system, and 13.6 μm (95% LoA, −5–32.2 μm) for Scheimpflug system and NCP. CONCLUSIONS: SS-OCT optical biometer overestimates CCT measurements compared to Scheimpflug system, SD-OCT device, and NCP. Given mean differences and range variations in CCT measurements between devices, SS-OCT optical biometer and Scheimpflug system are interchangeable as are SD-OCT and NCP. Medknow Publications & Media Pvt Ltd 2017-06 /pmc/articles/PMC5508455/ /pubmed/28643709 http://dx.doi.org/10.4103/ijo.IJO_618_16 Text en Copyright: © 2017 Indian Journal of Ophthalmology http://creativecommons.org/licenses/by-nc-sa/3.0 This is an open access article distributed under the terms of the Creative Commons Attribution-NonCommercial-ShareAlike 3.0 License, which allows others to remix, tweak, and build upon the work non-commercially, as long as the author is credited and the new creations are licensed under the identical terms.
spellingShingle Original Article
Özyol, Erhan
Özyol, Pelin
Comparison of central corneal thickness with four noncontact devices: An agreement analysis of swept-source technology
title Comparison of central corneal thickness with four noncontact devices: An agreement analysis of swept-source technology
title_full Comparison of central corneal thickness with four noncontact devices: An agreement analysis of swept-source technology
title_fullStr Comparison of central corneal thickness with four noncontact devices: An agreement analysis of swept-source technology
title_full_unstemmed Comparison of central corneal thickness with four noncontact devices: An agreement analysis of swept-source technology
title_short Comparison of central corneal thickness with four noncontact devices: An agreement analysis of swept-source technology
title_sort comparison of central corneal thickness with four noncontact devices: an agreement analysis of swept-source technology
topic Original Article
url https://www.ncbi.nlm.nih.gov/pmc/articles/PMC5508455/
https://www.ncbi.nlm.nih.gov/pubmed/28643709
http://dx.doi.org/10.4103/ijo.IJO_618_16
work_keys_str_mv AT ozyolerhan comparisonofcentralcornealthicknesswithfournoncontactdevicesanagreementanalysisofsweptsourcetechnology
AT ozyolpelin comparisonofcentralcornealthicknesswithfournoncontactdevicesanagreementanalysisofsweptsourcetechnology