Cargando…
Chemical and Structural Characterization of Several Mid-Term Explanted Breast Prostheses
The recent scandal of poly implant prostheses (PIP), which were found in some cases to be made of non-medical grade silicone (as reported by the European Scientific Committee on Emerging and Newly Identified Health Risks), had a great social impact. Thousands of patients asked for implant removal wi...
Autores principales: | , , , , , |
---|---|
Formato: | Online Artículo Texto |
Lenguaje: | English |
Publicado: |
MDPI
2016
|
Materias: | |
Acceso en línea: | https://www.ncbi.nlm.nih.gov/pmc/articles/PMC5510735/ https://www.ncbi.nlm.nih.gov/pubmed/28773799 http://dx.doi.org/10.3390/ma9080678 |
Sumario: | The recent scandal of poly implant prostheses (PIP), which were found in some cases to be made of non-medical grade silicone (as reported by the European Scientific Committee on Emerging and Newly Identified Health Risks), had a great social impact. Thousands of patients asked for implant removal with significant costs for public health care systems. We analysed, by a multidisciplinary approach, sixteen different breast implants after explantation by using several analytical and structural techniques, such as Fourier Transform infrared spectroscopy (FT-IR), mass spectrometry equipped by ion coupled plasma (ICP-MS), gas-chromatography (GC-MS), and tensile testing. Traces of organic (fatty acid) and inorganic (Fe, Cr, Pt, Na, and other metals) substances were found in all samples, and, even if these values are under danger threshold levels, our study results highlight the possibility of bioaccumulation and tissue contamination, implying the need for continuous medical surveillance and monitoring of material aging. |
---|