Normal foot loading parameters and repeatability of the Footscan® platform system
BACKGROUND: The Footscan® platform system is one of the most commonly used clinical tools for the measurement of the foot pressure. The present study was designed to assess the repeatability of the system and identify the range of loading parameters observed in the normal foot. METHODS: Measurements...
Autores principales: | , , , , , , |
---|---|
Formato: | Online Artículo Texto |
Lenguaje: | English |
Publicado: |
BioMed Central
2017
|
Materias: | |
Acceso en línea: | https://www.ncbi.nlm.nih.gov/pmc/articles/PMC5512844/ https://www.ncbi.nlm.nih.gov/pubmed/28725271 http://dx.doi.org/10.1186/s13047-017-0209-2 |
_version_ | 1783250541259784192 |
---|---|
author | Xu, Chao Wen, Xin-Xin Huang, Lu-Yu Shang, Lei Cheng, Xi-Xia Yan, Ya-Bo Lei, Wei |
author_facet | Xu, Chao Wen, Xin-Xin Huang, Lu-Yu Shang, Lei Cheng, Xi-Xia Yan, Ya-Bo Lei, Wei |
author_sort | Xu, Chao |
collection | PubMed |
description | BACKGROUND: The Footscan® platform system is one of the most commonly used clinical tools for the measurement of the foot pressure. The present study was designed to assess the repeatability of the system and identify the range of loading parameters observed in the normal foot. METHODS: Measurements were collected from 32 healthy participants, 15 females and 17 males, twice at an interval of 1 week. Peak pressure (PP), contact time (CT), contact area (CA), pressure-time integral (PTI), and maximum force (MaF) were recorded; these parameters were investigated in 10 areas of the foot: medial heel, lateral heel, midfoot, first to fifth metatarsals, hallux, and toes 2–5. The intra-session repeatability was evaluated by calculating the intraclass correlation coefficients (ICCs) and coefficients of variation (CVs) across the three repeated trials within the same session. The inter-session repeatability was assessed using the average of the three trials in each session to determine the ICCs and CVs. RESULTS: The ICCs showed moderate to good repeatability for every variable of interest, and the CVs were all <28%. The highest zones of PP were found under the second and third metatarsals, followed by the medial heel. The CT was 68.5–82.8% of the total stance time under the metatarsal heads. CA was highest under the midfoot, PTI was highest under the second metatarsal, and MaF was highest under the medial heel. CONCLUSIONS: Footscan® platform system was found to be repeatable. Thus, it can be used as a valuable tool in the assessment of plantar pressure distribution, and the normal values of the foot loading parameters identified in this study can be employed to provide a reference range for the gait analysis performed by the Footscan® system. |
format | Online Article Text |
id | pubmed-5512844 |
institution | National Center for Biotechnology Information |
language | English |
publishDate | 2017 |
publisher | BioMed Central |
record_format | MEDLINE/PubMed |
spelling | pubmed-55128442017-07-19 Normal foot loading parameters and repeatability of the Footscan® platform system Xu, Chao Wen, Xin-Xin Huang, Lu-Yu Shang, Lei Cheng, Xi-Xia Yan, Ya-Bo Lei, Wei J Foot Ankle Res Research BACKGROUND: The Footscan® platform system is one of the most commonly used clinical tools for the measurement of the foot pressure. The present study was designed to assess the repeatability of the system and identify the range of loading parameters observed in the normal foot. METHODS: Measurements were collected from 32 healthy participants, 15 females and 17 males, twice at an interval of 1 week. Peak pressure (PP), contact time (CT), contact area (CA), pressure-time integral (PTI), and maximum force (MaF) were recorded; these parameters were investigated in 10 areas of the foot: medial heel, lateral heel, midfoot, first to fifth metatarsals, hallux, and toes 2–5. The intra-session repeatability was evaluated by calculating the intraclass correlation coefficients (ICCs) and coefficients of variation (CVs) across the three repeated trials within the same session. The inter-session repeatability was assessed using the average of the three trials in each session to determine the ICCs and CVs. RESULTS: The ICCs showed moderate to good repeatability for every variable of interest, and the CVs were all <28%. The highest zones of PP were found under the second and third metatarsals, followed by the medial heel. The CT was 68.5–82.8% of the total stance time under the metatarsal heads. CA was highest under the midfoot, PTI was highest under the second metatarsal, and MaF was highest under the medial heel. CONCLUSIONS: Footscan® platform system was found to be repeatable. Thus, it can be used as a valuable tool in the assessment of plantar pressure distribution, and the normal values of the foot loading parameters identified in this study can be employed to provide a reference range for the gait analysis performed by the Footscan® system. BioMed Central 2017-07-17 /pmc/articles/PMC5512844/ /pubmed/28725271 http://dx.doi.org/10.1186/s13047-017-0209-2 Text en © The Author(s). 2017 Open AccessThis article is distributed under the terms of the Creative Commons Attribution 4.0 International License (http://creativecommons.org/licenses/by/4.0/), which permits unrestricted use, distribution, and reproduction in any medium, provided you give appropriate credit to the original author(s) and the source, provide a link to the Creative Commons license, and indicate if changes were made. The Creative Commons Public Domain Dedication waiver (http://creativecommons.org/publicdomain/zero/1.0/) applies to the data made available in this article, unless otherwise stated. |
spellingShingle | Research Xu, Chao Wen, Xin-Xin Huang, Lu-Yu Shang, Lei Cheng, Xi-Xia Yan, Ya-Bo Lei, Wei Normal foot loading parameters and repeatability of the Footscan® platform system |
title | Normal foot loading parameters and repeatability of the Footscan® platform system |
title_full | Normal foot loading parameters and repeatability of the Footscan® platform system |
title_fullStr | Normal foot loading parameters and repeatability of the Footscan® platform system |
title_full_unstemmed | Normal foot loading parameters and repeatability of the Footscan® platform system |
title_short | Normal foot loading parameters and repeatability of the Footscan® platform system |
title_sort | normal foot loading parameters and repeatability of the footscan® platform system |
topic | Research |
url | https://www.ncbi.nlm.nih.gov/pmc/articles/PMC5512844/ https://www.ncbi.nlm.nih.gov/pubmed/28725271 http://dx.doi.org/10.1186/s13047-017-0209-2 |
work_keys_str_mv | AT xuchao normalfootloadingparametersandrepeatabilityofthefootscanplatformsystem AT wenxinxin normalfootloadingparametersandrepeatabilityofthefootscanplatformsystem AT huangluyu normalfootloadingparametersandrepeatabilityofthefootscanplatformsystem AT shanglei normalfootloadingparametersandrepeatabilityofthefootscanplatformsystem AT chengxixia normalfootloadingparametersandrepeatabilityofthefootscanplatformsystem AT yanyabo normalfootloadingparametersandrepeatabilityofthefootscanplatformsystem AT leiwei normalfootloadingparametersandrepeatabilityofthefootscanplatformsystem |