Evaluation of a new electronic preoperative reference marker for toric intraocular lens implantation by two different methods of analysis: Adobe Photoshop versus iTrace
PURPOSE: The aim of this study is to compare two different methods of analysis of preoperative reference marking for toric intraocular lens (IOL) after marking with an electronic marker. SETTING/VENUE: Cataract and IOL Implantation Service, Shroff Eye Centre, New Delhi, India. PATIENTS AND METHODS:...
Autores principales: | , , , , |
---|---|
Formato: | Online Artículo Texto |
Lenguaje: | English |
Publicado: |
Medknow Publications & Media Pvt Ltd
2017
|
Materias: | |
Acceso en línea: | https://www.ncbi.nlm.nih.gov/pmc/articles/PMC5516471/ https://www.ncbi.nlm.nih.gov/pubmed/28757694 http://dx.doi.org/10.4103/ojo.OJO_163_2013 |
_version_ | 1783251168672088064 |
---|---|
author | Farooqui, Javed Hussain Sharma, Mansi Koul, Archana Dutta, Ranjan Shroff, Noshir Minoo |
author_facet | Farooqui, Javed Hussain Sharma, Mansi Koul, Archana Dutta, Ranjan Shroff, Noshir Minoo |
author_sort | Farooqui, Javed Hussain |
collection | PubMed |
description | PURPOSE: The aim of this study is to compare two different methods of analysis of preoperative reference marking for toric intraocular lens (IOL) after marking with an electronic marker. SETTING/VENUE: Cataract and IOL Implantation Service, Shroff Eye Centre, New Delhi, India. PATIENTS AND METHODS: Fifty-two eyes of thirty patients planned for toric IOL implantation were included in the study. All patients had preoperative marking performed with an electronic preoperative two-step toric IOL reference marker (ASICO AE-2929). Reference marks were placed at 3-and 9-o'clock positions. Marks were analyzed with two systems. First, slit-lamp photographs taken and analyzed using Adobe Photoshop (version 7.0). Second, Tracey iTrace Visual Function Analyzer (version 5.1.1) was used for capturing corneal topograph examination and position of marks noted. Amount of alignment error was calculated. RESULTS: Mean absolute rotation error was 2.38 ± 1.78° by Photoshop and 2.87 ± 2.03° by iTrace which was not statistically significant (P = 0.215). Nearly 72.7% of eyes by Photoshop and 61.4% by iTrace had rotation error ≤3° (P = 0.359); and 90.9% of eyes by Photoshop and 81.8% by iTrace had rotation error ≤5° (P = 0.344). No significant difference in absolute amount of rotation between eyes when analyzed by either method. CONCLUSIONS: Difference in reference mark positions when analyzed by two systems suggests the presence of varying cyclotorsion at different points of time. Both analysis methods showed an approximately 3° of alignment error, which could contribute to 10% loss of astigmatic correction of toric IOL. This can be further compounded by intra-operative marking errors and final placement of IOL in the bag. |
format | Online Article Text |
id | pubmed-5516471 |
institution | National Center for Biotechnology Information |
language | English |
publishDate | 2017 |
publisher | Medknow Publications & Media Pvt Ltd |
record_format | MEDLINE/PubMed |
spelling | pubmed-55164712017-07-28 Evaluation of a new electronic preoperative reference marker for toric intraocular lens implantation by two different methods of analysis: Adobe Photoshop versus iTrace Farooqui, Javed Hussain Sharma, Mansi Koul, Archana Dutta, Ranjan Shroff, Noshir Minoo Oman J Ophthalmol Original Article PURPOSE: The aim of this study is to compare two different methods of analysis of preoperative reference marking for toric intraocular lens (IOL) after marking with an electronic marker. SETTING/VENUE: Cataract and IOL Implantation Service, Shroff Eye Centre, New Delhi, India. PATIENTS AND METHODS: Fifty-two eyes of thirty patients planned for toric IOL implantation were included in the study. All patients had preoperative marking performed with an electronic preoperative two-step toric IOL reference marker (ASICO AE-2929). Reference marks were placed at 3-and 9-o'clock positions. Marks were analyzed with two systems. First, slit-lamp photographs taken and analyzed using Adobe Photoshop (version 7.0). Second, Tracey iTrace Visual Function Analyzer (version 5.1.1) was used for capturing corneal topograph examination and position of marks noted. Amount of alignment error was calculated. RESULTS: Mean absolute rotation error was 2.38 ± 1.78° by Photoshop and 2.87 ± 2.03° by iTrace which was not statistically significant (P = 0.215). Nearly 72.7% of eyes by Photoshop and 61.4% by iTrace had rotation error ≤3° (P = 0.359); and 90.9% of eyes by Photoshop and 81.8% by iTrace had rotation error ≤5° (P = 0.344). No significant difference in absolute amount of rotation between eyes when analyzed by either method. CONCLUSIONS: Difference in reference mark positions when analyzed by two systems suggests the presence of varying cyclotorsion at different points of time. Both analysis methods showed an approximately 3° of alignment error, which could contribute to 10% loss of astigmatic correction of toric IOL. This can be further compounded by intra-operative marking errors and final placement of IOL in the bag. Medknow Publications & Media Pvt Ltd 2017 /pmc/articles/PMC5516471/ /pubmed/28757694 http://dx.doi.org/10.4103/ojo.OJO_163_2013 Text en Copyright: © 2017 Oman Ophthalmic Society http://creativecommons.org/licenses/by-nc-sa/3.0 This is an open access article distributed under the terms of the Creative Commons Attribution-NonCommercial-ShareAlike 3.0 License, which allows others to remix, tweak, and build upon the work non-commercially, as long as the author is credited and the new creations are licensed under the identical terms. |
spellingShingle | Original Article Farooqui, Javed Hussain Sharma, Mansi Koul, Archana Dutta, Ranjan Shroff, Noshir Minoo Evaluation of a new electronic preoperative reference marker for toric intraocular lens implantation by two different methods of analysis: Adobe Photoshop versus iTrace |
title | Evaluation of a new electronic preoperative reference marker for toric intraocular lens implantation by two different methods of analysis: Adobe Photoshop versus iTrace |
title_full | Evaluation of a new electronic preoperative reference marker for toric intraocular lens implantation by two different methods of analysis: Adobe Photoshop versus iTrace |
title_fullStr | Evaluation of a new electronic preoperative reference marker for toric intraocular lens implantation by two different methods of analysis: Adobe Photoshop versus iTrace |
title_full_unstemmed | Evaluation of a new electronic preoperative reference marker for toric intraocular lens implantation by two different methods of analysis: Adobe Photoshop versus iTrace |
title_short | Evaluation of a new electronic preoperative reference marker for toric intraocular lens implantation by two different methods of analysis: Adobe Photoshop versus iTrace |
title_sort | evaluation of a new electronic preoperative reference marker for toric intraocular lens implantation by two different methods of analysis: adobe photoshop versus itrace |
topic | Original Article |
url | https://www.ncbi.nlm.nih.gov/pmc/articles/PMC5516471/ https://www.ncbi.nlm.nih.gov/pubmed/28757694 http://dx.doi.org/10.4103/ojo.OJO_163_2013 |
work_keys_str_mv | AT farooquijavedhussain evaluationofanewelectronicpreoperativereferencemarkerfortoricintraocularlensimplantationbytwodifferentmethodsofanalysisadobephotoshopversusitrace AT sharmamansi evaluationofanewelectronicpreoperativereferencemarkerfortoricintraocularlensimplantationbytwodifferentmethodsofanalysisadobephotoshopversusitrace AT koularchana evaluationofanewelectronicpreoperativereferencemarkerfortoricintraocularlensimplantationbytwodifferentmethodsofanalysisadobephotoshopversusitrace AT duttaranjan evaluationofanewelectronicpreoperativereferencemarkerfortoricintraocularlensimplantationbytwodifferentmethodsofanalysisadobephotoshopversusitrace AT shroffnoshirminoo evaluationofanewelectronicpreoperativereferencemarkerfortoricintraocularlensimplantationbytwodifferentmethodsofanalysisadobephotoshopversusitrace |