Ejection fraction in myocardial perfusion imaging assessed with a dynamic phantom: comparison between IQ-SPECT and LEHR
BACKGROUND: Developments in single photon emission tomography instrumentation and reconstruction methods present a potential for decreasing acquisition times. One of such recent options for myocardial perfusion imaging (MPI) is IQ-SPECT. This study was motivated by the inconsistency in the reported...
Autores principales: | , , , |
---|---|
Formato: | Online Artículo Texto |
Lenguaje: | English |
Publicado: |
Springer International Publishing
2017
|
Materias: | |
Acceso en línea: | https://www.ncbi.nlm.nih.gov/pmc/articles/PMC5517386/ https://www.ncbi.nlm.nih.gov/pubmed/28726076 http://dx.doi.org/10.1186/s40658-017-0187-2 |
_version_ | 1783251271953678336 |
---|---|
author | Hippeläinen, Eero Mäkelä, Teemu Kaasalainen, Touko Kaleva, Erna |
author_facet | Hippeläinen, Eero Mäkelä, Teemu Kaasalainen, Touko Kaleva, Erna |
author_sort | Hippeläinen, Eero |
collection | PubMed |
description | BACKGROUND: Developments in single photon emission tomography instrumentation and reconstruction methods present a potential for decreasing acquisition times. One of such recent options for myocardial perfusion imaging (MPI) is IQ-SPECT. This study was motivated by the inconsistency in the reported ejection fraction (EF) and left ventricular (LV) volume results between IQ-SPECT and more conventional low-energy high-resolution (LEHR) collimation protocols. IQ-SPECT and LEHR quantitative results were compared while the equivalent number of iterations (EI) was varied. The end-diastolic (EDV) and end-systolic volumes (ESV) and the derived EF values were investigated. A dynamic heart phantom was used to produce repeatable ESVs, EDVs and EFs. Phantom performance was verified by comparing the set EF values to those measured from a gated multi-slice X-ray computed tomography (CT) scan (EF(True)). The phantom with an EF setting of 45, 55, 65 and 70% was imaged with both IQ-SPECT and LEHR protocols. The data were reconstructed with different EI, and two commonly used clinical myocardium delineation software were used to evaluate the LV volumes. RESULTS: The CT verification showed that the phantom EF settings were repeatable and accurate with the EF(True) being within 1% point from the manufacture’s nominal value. Depending on EI both MPI protocols can be made to produce correct EF estimates, but IQ-SPECT protocol produced on average 41 and 42% smaller EDV and ESV when compared to the phantom’s volumes, while LEHR protocol underestimated volumes by 24 and 21%, respectively. The volume results were largely similar between the delineation methods used. CONCLUSIONS: The reconstruction parameters can greatly affect the volume estimates obtained from perfusion studies. IQ-SPECT produces systematically smaller LV volumes than the conventional LEHR MPI protocol. The volume estimates are also software dependent. |
format | Online Article Text |
id | pubmed-5517386 |
institution | National Center for Biotechnology Information |
language | English |
publishDate | 2017 |
publisher | Springer International Publishing |
record_format | MEDLINE/PubMed |
spelling | pubmed-55173862017-08-02 Ejection fraction in myocardial perfusion imaging assessed with a dynamic phantom: comparison between IQ-SPECT and LEHR Hippeläinen, Eero Mäkelä, Teemu Kaasalainen, Touko Kaleva, Erna EJNMMI Phys Original Research BACKGROUND: Developments in single photon emission tomography instrumentation and reconstruction methods present a potential for decreasing acquisition times. One of such recent options for myocardial perfusion imaging (MPI) is IQ-SPECT. This study was motivated by the inconsistency in the reported ejection fraction (EF) and left ventricular (LV) volume results between IQ-SPECT and more conventional low-energy high-resolution (LEHR) collimation protocols. IQ-SPECT and LEHR quantitative results were compared while the equivalent number of iterations (EI) was varied. The end-diastolic (EDV) and end-systolic volumes (ESV) and the derived EF values were investigated. A dynamic heart phantom was used to produce repeatable ESVs, EDVs and EFs. Phantom performance was verified by comparing the set EF values to those measured from a gated multi-slice X-ray computed tomography (CT) scan (EF(True)). The phantom with an EF setting of 45, 55, 65 and 70% was imaged with both IQ-SPECT and LEHR protocols. The data were reconstructed with different EI, and two commonly used clinical myocardium delineation software were used to evaluate the LV volumes. RESULTS: The CT verification showed that the phantom EF settings were repeatable and accurate with the EF(True) being within 1% point from the manufacture’s nominal value. Depending on EI both MPI protocols can be made to produce correct EF estimates, but IQ-SPECT protocol produced on average 41 and 42% smaller EDV and ESV when compared to the phantom’s volumes, while LEHR protocol underestimated volumes by 24 and 21%, respectively. The volume results were largely similar between the delineation methods used. CONCLUSIONS: The reconstruction parameters can greatly affect the volume estimates obtained from perfusion studies. IQ-SPECT produces systematically smaller LV volumes than the conventional LEHR MPI protocol. The volume estimates are also software dependent. Springer International Publishing 2017-07-19 /pmc/articles/PMC5517386/ /pubmed/28726076 http://dx.doi.org/10.1186/s40658-017-0187-2 Text en © The Author(s). 2017 Open AccessThis article is distributed under the terms of the Creative Commons Attribution 4.0 International License (http://creativecommons.org/licenses/by/4.0/), which permits unrestricted use, distribution, and reproduction in any medium, provided you give appropriate credit to the original author(s) and the source, provide a link to the Creative Commons license, and indicate if changes were made. |
spellingShingle | Original Research Hippeläinen, Eero Mäkelä, Teemu Kaasalainen, Touko Kaleva, Erna Ejection fraction in myocardial perfusion imaging assessed with a dynamic phantom: comparison between IQ-SPECT and LEHR |
title | Ejection fraction in myocardial perfusion imaging assessed with a dynamic phantom: comparison between IQ-SPECT and LEHR |
title_full | Ejection fraction in myocardial perfusion imaging assessed with a dynamic phantom: comparison between IQ-SPECT and LEHR |
title_fullStr | Ejection fraction in myocardial perfusion imaging assessed with a dynamic phantom: comparison between IQ-SPECT and LEHR |
title_full_unstemmed | Ejection fraction in myocardial perfusion imaging assessed with a dynamic phantom: comparison between IQ-SPECT and LEHR |
title_short | Ejection fraction in myocardial perfusion imaging assessed with a dynamic phantom: comparison between IQ-SPECT and LEHR |
title_sort | ejection fraction in myocardial perfusion imaging assessed with a dynamic phantom: comparison between iq-spect and lehr |
topic | Original Research |
url | https://www.ncbi.nlm.nih.gov/pmc/articles/PMC5517386/ https://www.ncbi.nlm.nih.gov/pubmed/28726076 http://dx.doi.org/10.1186/s40658-017-0187-2 |
work_keys_str_mv | AT hippelaineneero ejectionfractioninmyocardialperfusionimagingassessedwithadynamicphantomcomparisonbetweeniqspectandlehr AT makelateemu ejectionfractioninmyocardialperfusionimagingassessedwithadynamicphantomcomparisonbetweeniqspectandlehr AT kaasalainentouko ejectionfractioninmyocardialperfusionimagingassessedwithadynamicphantomcomparisonbetweeniqspectandlehr AT kalevaerna ejectionfractioninmyocardialperfusionimagingassessedwithadynamicphantomcomparisonbetweeniqspectandlehr |