Reliability and Repeatability of Cone Density Measurements in Patients With Stargardt Disease and RPGR-Associated Retinopathy

PURPOSE: To assess reliability and repeatability of cone density measurements by using confocal and (nonconfocal) split-detector adaptive optics scanning light ophthalmoscopy (AOSLO) imaging. It will be determined whether cone density values are significantly different between modalities in Stargard...

Descripción completa

Detalles Bibliográficos
Autores principales: Tanna, Preena, Kasilian, Melissa, Strauss, Rupert, Tee, James, Kalitzeos, Angelos, Tarima, Sergey, Visotcky, Alexis, Dubra, Alfredo, Carroll, Joseph, Michaelides, Michel
Formato: Online Artículo Texto
Lenguaje:English
Publicado: The Association for Research in Vision and Ophthalmology 2017
Materias:
Acceso en línea:https://www.ncbi.nlm.nih.gov/pmc/articles/PMC5525557/
https://www.ncbi.nlm.nih.gov/pubmed/28738413
http://dx.doi.org/10.1167/iovs.17-21904
_version_ 1783252658532909056
author Tanna, Preena
Kasilian, Melissa
Strauss, Rupert
Tee, James
Kalitzeos, Angelos
Tarima, Sergey
Visotcky, Alexis
Dubra, Alfredo
Carroll, Joseph
Michaelides, Michel
author_facet Tanna, Preena
Kasilian, Melissa
Strauss, Rupert
Tee, James
Kalitzeos, Angelos
Tarima, Sergey
Visotcky, Alexis
Dubra, Alfredo
Carroll, Joseph
Michaelides, Michel
author_sort Tanna, Preena
collection PubMed
description PURPOSE: To assess reliability and repeatability of cone density measurements by using confocal and (nonconfocal) split-detector adaptive optics scanning light ophthalmoscopy (AOSLO) imaging. It will be determined whether cone density values are significantly different between modalities in Stargardt disease (STGD) and retinitis pigmentosa GTPase regulator (RPGR)–associated retinopathy. METHODS: Twelve patients with STGD (aged 9–52 years) and eight with RPGR-associated retinopathy (aged 11–31 years) were imaged using both confocal and split-detector AOSLO simultaneously. Four graders manually identified cone locations in each image that were used to calculate local densities. Each imaging modality was evaluated independently. The data set consisted of 1584 assessments of 99 STGD images (each image in two modalities and four graders who graded each image twice) and 928 RPGR assessments of 58 images (each image in two modalities and four graders who graded each image twice). RESULTS: For STGD assessments the reliability for confocal and split-detector AOSLO was 67.9% and 95.9%, respectively, and the repeatability was 71.2% and 97.3%, respectively. The differences in the measured cone density values between modalities were statistically significant for one grader. For RPGR assessments the reliability for confocal and split-detector AOSLO was 22.1% and 88.5%, respectively, and repeatability was 63.2% and 94.5%, respectively. The differences in cone density between modalities were statistically significant for all graders. CONCLUSIONS: Split-detector AOSLO greatly improved the reliability and repeatability of cone density measurements in both disorders and will be valuable for natural history studies and clinical trials using AOSLO. However, it appears that these indices may be disease dependent, implying the need for similar investigations in other conditions.
format Online
Article
Text
id pubmed-5525557
institution National Center for Biotechnology Information
language English
publishDate 2017
publisher The Association for Research in Vision and Ophthalmology
record_format MEDLINE/PubMed
spelling pubmed-55255572017-07-26 Reliability and Repeatability of Cone Density Measurements in Patients With Stargardt Disease and RPGR-Associated Retinopathy Tanna, Preena Kasilian, Melissa Strauss, Rupert Tee, James Kalitzeos, Angelos Tarima, Sergey Visotcky, Alexis Dubra, Alfredo Carroll, Joseph Michaelides, Michel Invest Ophthalmol Vis Sci Retina PURPOSE: To assess reliability and repeatability of cone density measurements by using confocal and (nonconfocal) split-detector adaptive optics scanning light ophthalmoscopy (AOSLO) imaging. It will be determined whether cone density values are significantly different between modalities in Stargardt disease (STGD) and retinitis pigmentosa GTPase regulator (RPGR)–associated retinopathy. METHODS: Twelve patients with STGD (aged 9–52 years) and eight with RPGR-associated retinopathy (aged 11–31 years) were imaged using both confocal and split-detector AOSLO simultaneously. Four graders manually identified cone locations in each image that were used to calculate local densities. Each imaging modality was evaluated independently. The data set consisted of 1584 assessments of 99 STGD images (each image in two modalities and four graders who graded each image twice) and 928 RPGR assessments of 58 images (each image in two modalities and four graders who graded each image twice). RESULTS: For STGD assessments the reliability for confocal and split-detector AOSLO was 67.9% and 95.9%, respectively, and the repeatability was 71.2% and 97.3%, respectively. The differences in the measured cone density values between modalities were statistically significant for one grader. For RPGR assessments the reliability for confocal and split-detector AOSLO was 22.1% and 88.5%, respectively, and repeatability was 63.2% and 94.5%, respectively. The differences in cone density between modalities were statistically significant for all graders. CONCLUSIONS: Split-detector AOSLO greatly improved the reliability and repeatability of cone density measurements in both disorders and will be valuable for natural history studies and clinical trials using AOSLO. However, it appears that these indices may be disease dependent, implying the need for similar investigations in other conditions. The Association for Research in Vision and Ophthalmology 2017-07 /pmc/articles/PMC5525557/ /pubmed/28738413 http://dx.doi.org/10.1167/iovs.17-21904 Text en Copyright 2017 The Authors http://creativecommons.org/licenses/by-nc-nd/4.0/ This work is licensed under a Creative Commons Attribution-NonCommercial-NoDerivatives 4.0 International License.
spellingShingle Retina
Tanna, Preena
Kasilian, Melissa
Strauss, Rupert
Tee, James
Kalitzeos, Angelos
Tarima, Sergey
Visotcky, Alexis
Dubra, Alfredo
Carroll, Joseph
Michaelides, Michel
Reliability and Repeatability of Cone Density Measurements in Patients With Stargardt Disease and RPGR-Associated Retinopathy
title Reliability and Repeatability of Cone Density Measurements in Patients With Stargardt Disease and RPGR-Associated Retinopathy
title_full Reliability and Repeatability of Cone Density Measurements in Patients With Stargardt Disease and RPGR-Associated Retinopathy
title_fullStr Reliability and Repeatability of Cone Density Measurements in Patients With Stargardt Disease and RPGR-Associated Retinopathy
title_full_unstemmed Reliability and Repeatability of Cone Density Measurements in Patients With Stargardt Disease and RPGR-Associated Retinopathy
title_short Reliability and Repeatability of Cone Density Measurements in Patients With Stargardt Disease and RPGR-Associated Retinopathy
title_sort reliability and repeatability of cone density measurements in patients with stargardt disease and rpgr-associated retinopathy
topic Retina
url https://www.ncbi.nlm.nih.gov/pmc/articles/PMC5525557/
https://www.ncbi.nlm.nih.gov/pubmed/28738413
http://dx.doi.org/10.1167/iovs.17-21904
work_keys_str_mv AT tannapreena reliabilityandrepeatabilityofconedensitymeasurementsinpatientswithstargardtdiseaseandrpgrassociatedretinopathy
AT kasilianmelissa reliabilityandrepeatabilityofconedensitymeasurementsinpatientswithstargardtdiseaseandrpgrassociatedretinopathy
AT straussrupert reliabilityandrepeatabilityofconedensitymeasurementsinpatientswithstargardtdiseaseandrpgrassociatedretinopathy
AT teejames reliabilityandrepeatabilityofconedensitymeasurementsinpatientswithstargardtdiseaseandrpgrassociatedretinopathy
AT kalitzeosangelos reliabilityandrepeatabilityofconedensitymeasurementsinpatientswithstargardtdiseaseandrpgrassociatedretinopathy
AT tarimasergey reliabilityandrepeatabilityofconedensitymeasurementsinpatientswithstargardtdiseaseandrpgrassociatedretinopathy
AT visotckyalexis reliabilityandrepeatabilityofconedensitymeasurementsinpatientswithstargardtdiseaseandrpgrassociatedretinopathy
AT dubraalfredo reliabilityandrepeatabilityofconedensitymeasurementsinpatientswithstargardtdiseaseandrpgrassociatedretinopathy
AT carrolljoseph reliabilityandrepeatabilityofconedensitymeasurementsinpatientswithstargardtdiseaseandrpgrassociatedretinopathy
AT michaelidesmichel reliabilityandrepeatabilityofconedensitymeasurementsinpatientswithstargardtdiseaseandrpgrassociatedretinopathy