A Novel Spinal Implant for Fusionless Scoliosis Correction: A Biomechanical Analysis of the Motion Preserving Properties of a Posterior Periapical Concave Distraction Device

STUDY DESIGN: Biomechanical study. OBJECTIVE: Recently, a posterior concave periapical distraction device for fusionless scoliosis correction was introduced. The goal of this study was to quantify the effect of the periapical distraction device on spinal range of motion (ROM) in comparison with trad...

Descripción completa

Detalles Bibliográficos
Autores principales: Holewijn, Roderick M., de Kleuver, Marinus, van der Veen, Albert J., Emanuel, Kaj S., Bisschop, Arno, Stadhouder, Agnita, van Royen, Barend J., Kingma, Idsart
Formato: Online Artículo Texto
Lenguaje:English
Publicado: SAGE Publications 2017
Materias:
Acceso en línea:https://www.ncbi.nlm.nih.gov/pmc/articles/PMC5544155/
https://www.ncbi.nlm.nih.gov/pubmed/28811983
http://dx.doi.org/10.1177/2192568217699377
_version_ 1783255230147723264
author Holewijn, Roderick M.
de Kleuver, Marinus
van der Veen, Albert J.
Emanuel, Kaj S.
Bisschop, Arno
Stadhouder, Agnita
van Royen, Barend J.
Kingma, Idsart
author_facet Holewijn, Roderick M.
de Kleuver, Marinus
van der Veen, Albert J.
Emanuel, Kaj S.
Bisschop, Arno
Stadhouder, Agnita
van Royen, Barend J.
Kingma, Idsart
author_sort Holewijn, Roderick M.
collection PubMed
description STUDY DESIGN: Biomechanical study. OBJECTIVE: Recently, a posterior concave periapical distraction device for fusionless scoliosis correction was introduced. The goal of this study was to quantify the effect of the periapical distraction device on spinal range of motion (ROM) in comparison with traditional rigid pedicle screw-rod instrumentation. METHODS: Using a spinal motion simulator, 6 human spines were loaded with 4 N m and 6 porcine spines with 2 N m to induce flexion-extension (FE), lateral bending (LB), and axial rotation (AR). ROM was measured in 3 conditions: untreated, periapical distraction device, and rigid pedicle screw-rod instrumentation. RESULTS: The periapical distraction device caused a significant (P < .05) decrease in ROM of FE (human, −40.0% and porcine, −55.9%) and LB (human, −18.2% and porcine, −17.9%) as compared to the untreated spine, while ROM of AR remained unaffected. In comparison, rigid instrumentation caused a significantly (P < .05) larger decrease in ROM of FE (human, −80.9% and porcine, −94.0%), LB (human, −75.0% and porcine, −92.2%), and AR (human, −71.3% and porcine, −86.9%). CONCLUSIONS: Although no destructive forces were applied, no device failures were observed. Spinal ROM was significantly less constrained by the periapical distraction device compared to rigid pedicle screw-rod instrumentation. Therefore, provided that scoliosis correction is achieved, a more physiological spinal motion is expected after scoliosis correction with the posterior concave periapical distraction device.
format Online
Article
Text
id pubmed-5544155
institution National Center for Biotechnology Information
language English
publishDate 2017
publisher SAGE Publications
record_format MEDLINE/PubMed
spelling pubmed-55441552017-08-15 A Novel Spinal Implant for Fusionless Scoliosis Correction: A Biomechanical Analysis of the Motion Preserving Properties of a Posterior Periapical Concave Distraction Device Holewijn, Roderick M. de Kleuver, Marinus van der Veen, Albert J. Emanuel, Kaj S. Bisschop, Arno Stadhouder, Agnita van Royen, Barend J. Kingma, Idsart Global Spine J Original Articles STUDY DESIGN: Biomechanical study. OBJECTIVE: Recently, a posterior concave periapical distraction device for fusionless scoliosis correction was introduced. The goal of this study was to quantify the effect of the periapical distraction device on spinal range of motion (ROM) in comparison with traditional rigid pedicle screw-rod instrumentation. METHODS: Using a spinal motion simulator, 6 human spines were loaded with 4 N m and 6 porcine spines with 2 N m to induce flexion-extension (FE), lateral bending (LB), and axial rotation (AR). ROM was measured in 3 conditions: untreated, periapical distraction device, and rigid pedicle screw-rod instrumentation. RESULTS: The periapical distraction device caused a significant (P < .05) decrease in ROM of FE (human, −40.0% and porcine, −55.9%) and LB (human, −18.2% and porcine, −17.9%) as compared to the untreated spine, while ROM of AR remained unaffected. In comparison, rigid instrumentation caused a significantly (P < .05) larger decrease in ROM of FE (human, −80.9% and porcine, −94.0%), LB (human, −75.0% and porcine, −92.2%), and AR (human, −71.3% and porcine, −86.9%). CONCLUSIONS: Although no destructive forces were applied, no device failures were observed. Spinal ROM was significantly less constrained by the periapical distraction device compared to rigid pedicle screw-rod instrumentation. Therefore, provided that scoliosis correction is achieved, a more physiological spinal motion is expected after scoliosis correction with the posterior concave periapical distraction device. SAGE Publications 2017-04-07 2017-08 /pmc/articles/PMC5544155/ /pubmed/28811983 http://dx.doi.org/10.1177/2192568217699377 Text en © The Author(s) 2017 http://creativecommons.org/licenses/by-nc-nd/4.0/ This article is distributed under the terms of the Creative Commons Attribution-NonCommercial-NoDerivs 4.0 License (http://www.creativecommons.org/licenses/by-nc-nd/4.0/) which permits non-commercial use, reproduction and distribution of the work as published without adaptation or alteration, without further permission provided the original work is attributed as specified on the SAGE and Open Access pages (https://us.sagepub.com/en-us/nam/open-access-at-sage).
spellingShingle Original Articles
Holewijn, Roderick M.
de Kleuver, Marinus
van der Veen, Albert J.
Emanuel, Kaj S.
Bisschop, Arno
Stadhouder, Agnita
van Royen, Barend J.
Kingma, Idsart
A Novel Spinal Implant for Fusionless Scoliosis Correction: A Biomechanical Analysis of the Motion Preserving Properties of a Posterior Periapical Concave Distraction Device
title A Novel Spinal Implant for Fusionless Scoliosis Correction: A Biomechanical Analysis of the Motion Preserving Properties of a Posterior Periapical Concave Distraction Device
title_full A Novel Spinal Implant for Fusionless Scoliosis Correction: A Biomechanical Analysis of the Motion Preserving Properties of a Posterior Periapical Concave Distraction Device
title_fullStr A Novel Spinal Implant for Fusionless Scoliosis Correction: A Biomechanical Analysis of the Motion Preserving Properties of a Posterior Periapical Concave Distraction Device
title_full_unstemmed A Novel Spinal Implant for Fusionless Scoliosis Correction: A Biomechanical Analysis of the Motion Preserving Properties of a Posterior Periapical Concave Distraction Device
title_short A Novel Spinal Implant for Fusionless Scoliosis Correction: A Biomechanical Analysis of the Motion Preserving Properties of a Posterior Periapical Concave Distraction Device
title_sort novel spinal implant for fusionless scoliosis correction: a biomechanical analysis of the motion preserving properties of a posterior periapical concave distraction device
topic Original Articles
url https://www.ncbi.nlm.nih.gov/pmc/articles/PMC5544155/
https://www.ncbi.nlm.nih.gov/pubmed/28811983
http://dx.doi.org/10.1177/2192568217699377
work_keys_str_mv AT holewijnroderickm anovelspinalimplantforfusionlessscoliosiscorrectionabiomechanicalanalysisofthemotionpreservingpropertiesofaposteriorperiapicalconcavedistractiondevice
AT dekleuvermarinus anovelspinalimplantforfusionlessscoliosiscorrectionabiomechanicalanalysisofthemotionpreservingpropertiesofaposteriorperiapicalconcavedistractiondevice
AT vanderveenalbertj anovelspinalimplantforfusionlessscoliosiscorrectionabiomechanicalanalysisofthemotionpreservingpropertiesofaposteriorperiapicalconcavedistractiondevice
AT emanuelkajs anovelspinalimplantforfusionlessscoliosiscorrectionabiomechanicalanalysisofthemotionpreservingpropertiesofaposteriorperiapicalconcavedistractiondevice
AT bisschoparno anovelspinalimplantforfusionlessscoliosiscorrectionabiomechanicalanalysisofthemotionpreservingpropertiesofaposteriorperiapicalconcavedistractiondevice
AT stadhouderagnita anovelspinalimplantforfusionlessscoliosiscorrectionabiomechanicalanalysisofthemotionpreservingpropertiesofaposteriorperiapicalconcavedistractiondevice
AT vanroyenbarendj anovelspinalimplantforfusionlessscoliosiscorrectionabiomechanicalanalysisofthemotionpreservingpropertiesofaposteriorperiapicalconcavedistractiondevice
AT kingmaidsart anovelspinalimplantforfusionlessscoliosiscorrectionabiomechanicalanalysisofthemotionpreservingpropertiesofaposteriorperiapicalconcavedistractiondevice
AT holewijnroderickm novelspinalimplantforfusionlessscoliosiscorrectionabiomechanicalanalysisofthemotionpreservingpropertiesofaposteriorperiapicalconcavedistractiondevice
AT dekleuvermarinus novelspinalimplantforfusionlessscoliosiscorrectionabiomechanicalanalysisofthemotionpreservingpropertiesofaposteriorperiapicalconcavedistractiondevice
AT vanderveenalbertj novelspinalimplantforfusionlessscoliosiscorrectionabiomechanicalanalysisofthemotionpreservingpropertiesofaposteriorperiapicalconcavedistractiondevice
AT emanuelkajs novelspinalimplantforfusionlessscoliosiscorrectionabiomechanicalanalysisofthemotionpreservingpropertiesofaposteriorperiapicalconcavedistractiondevice
AT bisschoparno novelspinalimplantforfusionlessscoliosiscorrectionabiomechanicalanalysisofthemotionpreservingpropertiesofaposteriorperiapicalconcavedistractiondevice
AT stadhouderagnita novelspinalimplantforfusionlessscoliosiscorrectionabiomechanicalanalysisofthemotionpreservingpropertiesofaposteriorperiapicalconcavedistractiondevice
AT vanroyenbarendj novelspinalimplantforfusionlessscoliosiscorrectionabiomechanicalanalysisofthemotionpreservingpropertiesofaposteriorperiapicalconcavedistractiondevice
AT kingmaidsart novelspinalimplantforfusionlessscoliosiscorrectionabiomechanicalanalysisofthemotionpreservingpropertiesofaposteriorperiapicalconcavedistractiondevice