Cargando…
Positioning Performance of Power and Manual Drivers in Posterior Spinal Fusion Procedures
This work presents an analysis and comparison of the efficacy of two methods for pedicle screw placement during posterior spinal fusion surgery. A total of 100 screws (64 manual and 36 power driven), all placed utilizing a surgical navigation system, were analyzed and compared. Final screw placement...
Autores principales: | , , , , |
---|---|
Formato: | Online Artículo Texto |
Lenguaje: | English |
Publicado: |
Hindawi
2017
|
Materias: | |
Acceso en línea: | https://www.ncbi.nlm.nih.gov/pmc/articles/PMC5551465/ https://www.ncbi.nlm.nih.gov/pubmed/28819344 http://dx.doi.org/10.1155/2017/7262841 |
_version_ | 1783256311301931008 |
---|---|
author | Prendergast, J. Micah Perry, Alexander C. Patel, Vikas V. Lindley, Emily M. Rentschler, Mark E. |
author_facet | Prendergast, J. Micah Perry, Alexander C. Patel, Vikas V. Lindley, Emily M. Rentschler, Mark E. |
author_sort | Prendergast, J. Micah |
collection | PubMed |
description | This work presents an analysis and comparison of the efficacy of two methods for pedicle screw placement during posterior spinal fusion surgery. A total of 100 screws (64 manual and 36 power driven), all placed utilizing a surgical navigation system, were analyzed and compared. Final screw placement was compared to initial surgical plans using the navigation system, and the final screw locations were analyzed on the basis of angular deviation from these planned trajectories as well as screw translation within a critical reference plane. The power driver was found to insignificantly decrease the resulting angular deviation of these pedicle screws with a mean deviation of 3.35 degrees compared to 3.44 degrees with the manual driver (p = 0.853). Conversely, the power driver was found to increase the translational distance in the critical region, with mean deviations of 2.45 mm for the power driver compared to 1.54 mm with the manual driver. The increase in translational deviation was significant (p = 0.002) indicating that there may be some loss in performance from the adoption of the power driver. |
format | Online Article Text |
id | pubmed-5551465 |
institution | National Center for Biotechnology Information |
language | English |
publishDate | 2017 |
publisher | Hindawi |
record_format | MEDLINE/PubMed |
spelling | pubmed-55514652017-08-17 Positioning Performance of Power and Manual Drivers in Posterior Spinal Fusion Procedures Prendergast, J. Micah Perry, Alexander C. Patel, Vikas V. Lindley, Emily M. Rentschler, Mark E. Appl Bionics Biomech Research Article This work presents an analysis and comparison of the efficacy of two methods for pedicle screw placement during posterior spinal fusion surgery. A total of 100 screws (64 manual and 36 power driven), all placed utilizing a surgical navigation system, were analyzed and compared. Final screw placement was compared to initial surgical plans using the navigation system, and the final screw locations were analyzed on the basis of angular deviation from these planned trajectories as well as screw translation within a critical reference plane. The power driver was found to insignificantly decrease the resulting angular deviation of these pedicle screws with a mean deviation of 3.35 degrees compared to 3.44 degrees with the manual driver (p = 0.853). Conversely, the power driver was found to increase the translational distance in the critical region, with mean deviations of 2.45 mm for the power driver compared to 1.54 mm with the manual driver. The increase in translational deviation was significant (p = 0.002) indicating that there may be some loss in performance from the adoption of the power driver. Hindawi 2017 2017-07-27 /pmc/articles/PMC5551465/ /pubmed/28819344 http://dx.doi.org/10.1155/2017/7262841 Text en Copyright © 2017 J. Micah Prendergast et al. http://creativecommons.org/licenses/by/4.0/ This is an open access article distributed under the Creative Commons Attribution License, which permits unrestricted use, distribution, and reproduction in any medium, provided the original work is properly cited. |
spellingShingle | Research Article Prendergast, J. Micah Perry, Alexander C. Patel, Vikas V. Lindley, Emily M. Rentschler, Mark E. Positioning Performance of Power and Manual Drivers in Posterior Spinal Fusion Procedures |
title | Positioning Performance of Power and Manual Drivers in Posterior Spinal Fusion Procedures |
title_full | Positioning Performance of Power and Manual Drivers in Posterior Spinal Fusion Procedures |
title_fullStr | Positioning Performance of Power and Manual Drivers in Posterior Spinal Fusion Procedures |
title_full_unstemmed | Positioning Performance of Power and Manual Drivers in Posterior Spinal Fusion Procedures |
title_short | Positioning Performance of Power and Manual Drivers in Posterior Spinal Fusion Procedures |
title_sort | positioning performance of power and manual drivers in posterior spinal fusion procedures |
topic | Research Article |
url | https://www.ncbi.nlm.nih.gov/pmc/articles/PMC5551465/ https://www.ncbi.nlm.nih.gov/pubmed/28819344 http://dx.doi.org/10.1155/2017/7262841 |
work_keys_str_mv | AT prendergastjmicah positioningperformanceofpowerandmanualdriversinposteriorspinalfusionprocedures AT perryalexanderc positioningperformanceofpowerandmanualdriversinposteriorspinalfusionprocedures AT patelvikasv positioningperformanceofpowerandmanualdriversinposteriorspinalfusionprocedures AT lindleyemilym positioningperformanceofpowerandmanualdriversinposteriorspinalfusionprocedures AT rentschlermarke positioningperformanceofpowerandmanualdriversinposteriorspinalfusionprocedures |