Dosimetric evaluation of MobiusFX in the RapidArc delivery quality assurance comparing with 3DVH
The dosimetric characteristics of MobiusFX, which uses the treatment machine log file to calculate the dose inside the patient body, were analyzed for use in the RapidArc delivery quality assurance (DQA) process. The accuracy and usefulness of MobiusFX in clinical cases was evaluated by comparing th...
Autores principales: | , , , , , |
---|---|
Formato: | Online Artículo Texto |
Lenguaje: | English |
Publicado: |
Public Library of Science
2017
|
Materias: | |
Acceso en línea: | https://www.ncbi.nlm.nih.gov/pmc/articles/PMC5568283/ https://www.ncbi.nlm.nih.gov/pubmed/28832605 http://dx.doi.org/10.1371/journal.pone.0183165 |
_version_ | 1783258828898304000 |
---|---|
author | Song, Ju-Young Jeong, Jae-Uk Yoon, Mee Sun Ahn, Sung-Ja Chung, Woong-Ki Nam, Taek-Keun |
author_facet | Song, Ju-Young Jeong, Jae-Uk Yoon, Mee Sun Ahn, Sung-Ja Chung, Woong-Ki Nam, Taek-Keun |
author_sort | Song, Ju-Young |
collection | PubMed |
description | The dosimetric characteristics of MobiusFX, which uses the treatment machine log file to calculate the dose inside the patient body, were analyzed for use in the RapidArc delivery quality assurance (DQA) process. The accuracy and usefulness of MobiusFX in clinical cases was evaluated by comparing the dose calculated by MobiusFX with that calculated by the conventional measurement dose based program, 3DVH. The results of gamma evaluation with three different criteria (3%–3 mm, 4%–3 mm, 5%–3 mm) were analyzed, and the dose changes were calculated while simulating variable position errors (6 mm, 3 mm) and dosimetric output increases (6%, 3%). Although the doses calculated by each tool were not identical due to differences in the calculation algorithms, the doses calculated by MobiusFX were generally similar to those calculated by 3DVH. Based on these results, MobiusFX exhibited the required accuracy for clinical application. However, it could not determine the dosimetric output variation. It should therefore be considered a supplementary DQA tool that can verify the error in the daily treatment process, but not an ideal DQA tool that can replace conventional measurement based DQA methods. |
format | Online Article Text |
id | pubmed-5568283 |
institution | National Center for Biotechnology Information |
language | English |
publishDate | 2017 |
publisher | Public Library of Science |
record_format | MEDLINE/PubMed |
spelling | pubmed-55682832017-09-09 Dosimetric evaluation of MobiusFX in the RapidArc delivery quality assurance comparing with 3DVH Song, Ju-Young Jeong, Jae-Uk Yoon, Mee Sun Ahn, Sung-Ja Chung, Woong-Ki Nam, Taek-Keun PLoS One Research Article The dosimetric characteristics of MobiusFX, which uses the treatment machine log file to calculate the dose inside the patient body, were analyzed for use in the RapidArc delivery quality assurance (DQA) process. The accuracy and usefulness of MobiusFX in clinical cases was evaluated by comparing the dose calculated by MobiusFX with that calculated by the conventional measurement dose based program, 3DVH. The results of gamma evaluation with three different criteria (3%–3 mm, 4%–3 mm, 5%–3 mm) were analyzed, and the dose changes were calculated while simulating variable position errors (6 mm, 3 mm) and dosimetric output increases (6%, 3%). Although the doses calculated by each tool were not identical due to differences in the calculation algorithms, the doses calculated by MobiusFX were generally similar to those calculated by 3DVH. Based on these results, MobiusFX exhibited the required accuracy for clinical application. However, it could not determine the dosimetric output variation. It should therefore be considered a supplementary DQA tool that can verify the error in the daily treatment process, but not an ideal DQA tool that can replace conventional measurement based DQA methods. Public Library of Science 2017-08-23 /pmc/articles/PMC5568283/ /pubmed/28832605 http://dx.doi.org/10.1371/journal.pone.0183165 Text en © 2017 Song et al http://creativecommons.org/licenses/by/4.0/ This is an open access article distributed under the terms of the Creative Commons Attribution License (http://creativecommons.org/licenses/by/4.0/) , which permits unrestricted use, distribution, and reproduction in any medium, provided the original author and source are credited. |
spellingShingle | Research Article Song, Ju-Young Jeong, Jae-Uk Yoon, Mee Sun Ahn, Sung-Ja Chung, Woong-Ki Nam, Taek-Keun Dosimetric evaluation of MobiusFX in the RapidArc delivery quality assurance comparing with 3DVH |
title | Dosimetric evaluation of MobiusFX in the RapidArc delivery quality assurance comparing with 3DVH |
title_full | Dosimetric evaluation of MobiusFX in the RapidArc delivery quality assurance comparing with 3DVH |
title_fullStr | Dosimetric evaluation of MobiusFX in the RapidArc delivery quality assurance comparing with 3DVH |
title_full_unstemmed | Dosimetric evaluation of MobiusFX in the RapidArc delivery quality assurance comparing with 3DVH |
title_short | Dosimetric evaluation of MobiusFX in the RapidArc delivery quality assurance comparing with 3DVH |
title_sort | dosimetric evaluation of mobiusfx in the rapidarc delivery quality assurance comparing with 3dvh |
topic | Research Article |
url | https://www.ncbi.nlm.nih.gov/pmc/articles/PMC5568283/ https://www.ncbi.nlm.nih.gov/pubmed/28832605 http://dx.doi.org/10.1371/journal.pone.0183165 |
work_keys_str_mv | AT songjuyoung dosimetricevaluationofmobiusfxintherapidarcdeliveryqualityassurancecomparingwith3dvh AT jeongjaeuk dosimetricevaluationofmobiusfxintherapidarcdeliveryqualityassurancecomparingwith3dvh AT yoonmeesun dosimetricevaluationofmobiusfxintherapidarcdeliveryqualityassurancecomparingwith3dvh AT ahnsungja dosimetricevaluationofmobiusfxintherapidarcdeliveryqualityassurancecomparingwith3dvh AT chungwoongki dosimetricevaluationofmobiusfxintherapidarcdeliveryqualityassurancecomparingwith3dvh AT namtaekkeun dosimetricevaluationofmobiusfxintherapidarcdeliveryqualityassurancecomparingwith3dvh |