Cargando…

Comparative efficacy and safety of contact force-sensing catheter and second-generation cryoballoon ablation for atrial fibrillation: a meta-analysis

This meta-analysis compared the efficacy and safety of the contact force (CF)-sensing catheter and second-generation cryoballoon (CB) ablation for treating atrial fibrillation (AF). Six controlled clinical trials comparing ablation for AF using a CF-sensing catheter or second-generation CB were iden...

Descripción completa

Detalles Bibliográficos
Autores principales: Zhou, X., Lv, W., Zhang, W., Ye, Y., Li, Y., Zhou, Q., Zhang, J., Xing, Q., Lu, Y., Zhang, L., Wang, H., Qin, W., Tang, B.
Formato: Online Artículo Texto
Lenguaje:English
Publicado: Associação Brasileira de Divulgação Científica 2017
Materias:
Acceso en línea:https://www.ncbi.nlm.nih.gov/pmc/articles/PMC5572854/
https://www.ncbi.nlm.nih.gov/pubmed/28793059
http://dx.doi.org/10.1590/1414-431X20176409
Descripción
Sumario:This meta-analysis compared the efficacy and safety of the contact force (CF)-sensing catheter and second-generation cryoballoon (CB) ablation for treating atrial fibrillation (AF). Six controlled clinical trials comparing ablation for AF using a CF-sensing catheter or second-generation CB were identified from PubMed, EMBASE, Cochrane Library, Wanfang Data, and China National Knowledge Infrastructure. The procedure duration was significantly lower in the CB group compared with that in the CF group [mean difference (MD)=29.4; 95%CI=17.84–40.96; P=0.01], whereas there was no difference between the groups for fluoroscopy duration (MD=0.59; 95%CI=–4.48–5.66; P=0.82). Moreover, there was no difference in the incidence of non-lethal complications (embolic event, tamponade, femoral/subclavian hematoma, arteriovenous fistula, pulmonary vein stenosis, phrenic nerve palsy, and esophageal injury) between the CB and the CF groups (8.38 vs 5.35%; RR=0.66; 95%CI=0.37–1.17; P=0.15). Transient phrenic nerve palsy occurred in 17 of 326 patients (5.2%) of the CB group vs none in the CF group (RR=0.12; 95%CI=0.03–0.43; P=0.001). A comparable proportion of patients in CF and CB groups suffered from AF recurrence during the 12-month follow-up after a single ablation procedure [risk ratio (RR)=1.03; 95%CI=0.78–1.35; P=0.84]. AF ablation using CF-sensing catheters and second-generation CB showed comparable fluoroscopy duration and efficacy (during a 12-month follow-up), with shorter procedure duration and different complications in the CB group.