Cargando…

A comparative assessment of the accuracies of Raypex 5, Raypex 6, iPex and iPex II electronic apex locators: An in vitro study

PURPOSE: The aims of this study were to examine the accuracy of iPex II and to compare it with those of Raypex 5, Raypex 6 and iPex electronic apex locators (EALs). MATERIALS AND METHODS: Thirty fresh human mandibular premolar teeth were used in this study. Crown segments were cut and root canals we...

Descripción completa

Detalles Bibliográficos
Autores principales: Gurel, Mugem Asli, Helvacioglu Kivanc, Bagdagul, Ekici, Adil
Formato: Online Artículo Texto
Lenguaje:English
Publicado: Istanbul University Faculty of Dentisty 2017
Materias:
Acceso en línea:https://www.ncbi.nlm.nih.gov/pmc/articles/PMC5573492/
https://www.ncbi.nlm.nih.gov/pubmed/28955583
http://dx.doi.org/10.17096/jiufd.61309
Descripción
Sumario:PURPOSE: The aims of this study were to examine the accuracy of iPex II and to compare it with those of Raypex 5, Raypex 6 and iPex electronic apex locators (EALs). MATERIALS AND METHODS: Thirty fresh human mandibular premolar teeth were used in this study. Crown segments were cut and root canals were coronally flared. A #10 K-file was inserted until its tip can be seen within apical foramen to determine actual working length (AWL). Teeth were embedded in alginate and each multi-frequency EALs were randomly tested to determine the electronic working length (EWL). Differences between AWLs and EWLs were statistically compared. RESULTS: No significant differences were found between four EALs. EWL measurements by Raypex 5 were accurate in 64.29%, Raypex 6 in 53.58%, iPex in 64.29% and iPex II in 50% of the specimens, within the range of ±0.5 mm from the AWL. CONCLUSION: Within the limitations of this in vitro experiment, our findings indicate that the accuracy of working length measurements calculated with iPex II was similar to those of other multi-frequency EALs used in this study.