Assessment of magnetic field interactions and radiofrequency‐radiation‐induced heating of metallic spinal implants in 7 T field

The safety of metallic spinal implants in magnetic resonance imaging (MRI) performed using ultrahigh fields has not been established. Hence, we examined whether the displacement forces caused by a static magnetic field and the heating induced by radiofrequency radiation are substantial for spinal im...

Descripción completa

Detalles Bibliográficos
Autores principales: Tsukimura, Itsuko, Murakami, Hideki, Sasaki, Makoto, Endo, Hirooki, Yamabe, Daisuke, Oikawa, Ryosuke, Doita, Minoru
Formato: Online Artículo Texto
Lenguaje:English
Publicado: John Wiley and Sons Inc. 2017
Materias:
Acceso en línea:https://www.ncbi.nlm.nih.gov/pmc/articles/PMC5573950/
https://www.ncbi.nlm.nih.gov/pubmed/27769107
http://dx.doi.org/10.1002/jor.23464
_version_ 1783259745536180224
author Tsukimura, Itsuko
Murakami, Hideki
Sasaki, Makoto
Endo, Hirooki
Yamabe, Daisuke
Oikawa, Ryosuke
Doita, Minoru
author_facet Tsukimura, Itsuko
Murakami, Hideki
Sasaki, Makoto
Endo, Hirooki
Yamabe, Daisuke
Oikawa, Ryosuke
Doita, Minoru
author_sort Tsukimura, Itsuko
collection PubMed
description The safety of metallic spinal implants in magnetic resonance imaging (MRI) performed using ultrahigh fields has not been established. Hence, we examined whether the displacement forces caused by a static magnetic field and the heating induced by radiofrequency radiation are substantial for spinal implants in a 7 T field. We investigated spinal rods of various lengths and materials, a screw, and a cross‐linking bridge in accordance with the American Society for Testing and Materials guidelines. The displacement forces of the metallic implants in static 7 T and 3 T static magnetic fields were measured and compared. The temperature changes of the implants during 15‐min‐long fast spin‐echo and balanced gradient‐echo image acquisition sequences were measured in the 7 T field. The deflection angles of the metallic spinal materials in the 7 T field were 5.0–21.0° [median: 6.7°], significantly larger than those in the 3 T field (1.0–6.3° [2.2°]). Among the metallic rods, the cobalt–chrome rods had significantly larger deflection angles (17.8–21.0° [19.8°]) than the pure titanium and titanium alloy rods (5.0–7.7° [6.2°]). The temperature changes of the implants, including the cross‐linked rods, were 0.7–1.0°C [0.8°C] and 0.6–1.0°C [0.7°C] during the fast spin‐echo and balanced gradient‐echo sequences, respectively; these changes were slightly larger than those of the controls (0.4–1.1°C [0.5°C] and 0.3–0.9°C [0.6°C], respectively). All of the metallic spinal implants exhibited small displacement forces and minimal heating, indicating that MRI examinations using 7 T fields may be performed safely on patients with these implants. © 2016 The Authors. Journal of Orthopaedic Research Published by Wiley Periodicals, Inc. on behalf of Orthopaedic Research Society. J Orthop Res 35:1831–1837, 2017.
format Online
Article
Text
id pubmed-5573950
institution National Center for Biotechnology Information
language English
publishDate 2017
publisher John Wiley and Sons Inc.
record_format MEDLINE/PubMed
spelling pubmed-55739502017-09-15 Assessment of magnetic field interactions and radiofrequency‐radiation‐induced heating of metallic spinal implants in 7 T field Tsukimura, Itsuko Murakami, Hideki Sasaki, Makoto Endo, Hirooki Yamabe, Daisuke Oikawa, Ryosuke Doita, Minoru J Orthop Res Research Articles The safety of metallic spinal implants in magnetic resonance imaging (MRI) performed using ultrahigh fields has not been established. Hence, we examined whether the displacement forces caused by a static magnetic field and the heating induced by radiofrequency radiation are substantial for spinal implants in a 7 T field. We investigated spinal rods of various lengths and materials, a screw, and a cross‐linking bridge in accordance with the American Society for Testing and Materials guidelines. The displacement forces of the metallic implants in static 7 T and 3 T static magnetic fields were measured and compared. The temperature changes of the implants during 15‐min‐long fast spin‐echo and balanced gradient‐echo image acquisition sequences were measured in the 7 T field. The deflection angles of the metallic spinal materials in the 7 T field were 5.0–21.0° [median: 6.7°], significantly larger than those in the 3 T field (1.0–6.3° [2.2°]). Among the metallic rods, the cobalt–chrome rods had significantly larger deflection angles (17.8–21.0° [19.8°]) than the pure titanium and titanium alloy rods (5.0–7.7° [6.2°]). The temperature changes of the implants, including the cross‐linked rods, were 0.7–1.0°C [0.8°C] and 0.6–1.0°C [0.7°C] during the fast spin‐echo and balanced gradient‐echo sequences, respectively; these changes were slightly larger than those of the controls (0.4–1.1°C [0.5°C] and 0.3–0.9°C [0.6°C], respectively). All of the metallic spinal implants exhibited small displacement forces and minimal heating, indicating that MRI examinations using 7 T fields may be performed safely on patients with these implants. © 2016 The Authors. Journal of Orthopaedic Research Published by Wiley Periodicals, Inc. on behalf of Orthopaedic Research Society. J Orthop Res 35:1831–1837, 2017. John Wiley and Sons Inc. 2017-03-08 2017-08 /pmc/articles/PMC5573950/ /pubmed/27769107 http://dx.doi.org/10.1002/jor.23464 Text en © 2016 The Authors. Journal of Orthopaedic Research Published by Wiley Periodicals, Inc. on behalf of Orthopaedic Research Society. This is an open access article under the terms of the Creative Commons Attribution‐NonCommercial (http://creativecommons.org/licenses/by-nc/4.0/) License, which permits use, distribution and reproduction in any medium, provided the original work is properly cited and is not used for commercial purposes.
spellingShingle Research Articles
Tsukimura, Itsuko
Murakami, Hideki
Sasaki, Makoto
Endo, Hirooki
Yamabe, Daisuke
Oikawa, Ryosuke
Doita, Minoru
Assessment of magnetic field interactions and radiofrequency‐radiation‐induced heating of metallic spinal implants in 7 T field
title Assessment of magnetic field interactions and radiofrequency‐radiation‐induced heating of metallic spinal implants in 7 T field
title_full Assessment of magnetic field interactions and radiofrequency‐radiation‐induced heating of metallic spinal implants in 7 T field
title_fullStr Assessment of magnetic field interactions and radiofrequency‐radiation‐induced heating of metallic spinal implants in 7 T field
title_full_unstemmed Assessment of magnetic field interactions and radiofrequency‐radiation‐induced heating of metallic spinal implants in 7 T field
title_short Assessment of magnetic field interactions and radiofrequency‐radiation‐induced heating of metallic spinal implants in 7 T field
title_sort assessment of magnetic field interactions and radiofrequency‐radiation‐induced heating of metallic spinal implants in 7 t field
topic Research Articles
url https://www.ncbi.nlm.nih.gov/pmc/articles/PMC5573950/
https://www.ncbi.nlm.nih.gov/pubmed/27769107
http://dx.doi.org/10.1002/jor.23464
work_keys_str_mv AT tsukimuraitsuko assessmentofmagneticfieldinteractionsandradiofrequencyradiationinducedheatingofmetallicspinalimplantsin7tfield
AT murakamihideki assessmentofmagneticfieldinteractionsandradiofrequencyradiationinducedheatingofmetallicspinalimplantsin7tfield
AT sasakimakoto assessmentofmagneticfieldinteractionsandradiofrequencyradiationinducedheatingofmetallicspinalimplantsin7tfield
AT endohirooki assessmentofmagneticfieldinteractionsandradiofrequencyradiationinducedheatingofmetallicspinalimplantsin7tfield
AT yamabedaisuke assessmentofmagneticfieldinteractionsandradiofrequencyradiationinducedheatingofmetallicspinalimplantsin7tfield
AT oikawaryosuke assessmentofmagneticfieldinteractionsandradiofrequencyradiationinducedheatingofmetallicspinalimplantsin7tfield
AT doitaminoru assessmentofmagneticfieldinteractionsandradiofrequencyradiationinducedheatingofmetallicspinalimplantsin7tfield