The stability of quantitative blood count parameters using the ADVIA 2120i hematology analyzer

OBJECTIVES: To evaluate the stability of complete blood count (CBC) parameters by measuring at multiple time points up to 72 h after venepuncture. METHODS: 36 samples submitted for routine haematology were measured at 0, 8, 24, 48, 72 h. 18 samples were kept at room temperature (23–25 °C) and the ot...

Descripción completa

Detalles Bibliográficos
Autores principales: Pintér, Erzsébet, László, Kinga, Schüszler, Ildikó, Konderák, Judit
Formato: Online Artículo Texto
Lenguaje:English
Publicado: Elsevier 2015
Materias:
Acceso en línea:https://www.ncbi.nlm.nih.gov/pmc/articles/PMC5574499/
https://www.ncbi.nlm.nih.gov/pubmed/28856188
http://dx.doi.org/10.1016/j.plabm.2015.12.001
_version_ 1783259851611176960
author Pintér, Erzsébet
László, Kinga
Schüszler, Ildikó
Konderák, Judit
author_facet Pintér, Erzsébet
László, Kinga
Schüszler, Ildikó
Konderák, Judit
author_sort Pintér, Erzsébet
collection PubMed
description OBJECTIVES: To evaluate the stability of complete blood count (CBC) parameters by measuring at multiple time points up to 72 h after venepuncture. METHODS: 36 samples submitted for routine haematology were measured at 0, 8, 24, 48, 72 h. 18 samples were kept at room temperature (23–25 °C) and the other 18 at 4 °C. The stability of the CBC parameters was determined by comparing the results to the 0 h sample. White Blood Cell Count (WBC), Hemoglobin (Hb) concentration, Red Blood Cell (RBC) count, Hematocrit (Hct), Mean Corpuscular Volume (MCV), Mean Corpuscular Hemoglobin (MCH), Mean Corpuscular Hemoglobin Concentration (MCHC), Cellular Hemoglobin Concentration Mean (CHCM), Mean Platelet Volume (MPV), RBC distribution width (RDW), Platelet (PLT) and Reticulocyte counts were studied. RESULTS: Most parameters were stable for 24 h at 4 °C, except for MCV, MCHC, CHCM and MPV. MCV and MPV increased after 8 h (p<0.0001), whereas MCHC and CHCM decreased significantly after 8 h. Significant changes were found for MCHC in samples kept at 4 °C for 48 h (p=0.002), and for CHCM kept for 72 h (p<0.001). Reticulocyte count stability was maintained for 24 h at 4 °C (p=0.3047). In samples kept at room temperature changes occurred after 8 h in RBC, Hct, MCV, MCH, MCHC, CHCM, MPV and PLT. CONCLUSION: CBC measurements are reliable for 8 h when samples are stored at room temperature. The only parameter stable for 72 h at room temperature was Hb. Blood samples kept at 4 °C for up to 24 h are suitable for haematological analysis.
format Online
Article
Text
id pubmed-5574499
institution National Center for Biotechnology Information
language English
publishDate 2015
publisher Elsevier
record_format MEDLINE/PubMed
spelling pubmed-55744992017-08-30 The stability of quantitative blood count parameters using the ADVIA 2120i hematology analyzer Pintér, Erzsébet László, Kinga Schüszler, Ildikó Konderák, Judit Pract Lab Med Research Article OBJECTIVES: To evaluate the stability of complete blood count (CBC) parameters by measuring at multiple time points up to 72 h after venepuncture. METHODS: 36 samples submitted for routine haematology were measured at 0, 8, 24, 48, 72 h. 18 samples were kept at room temperature (23–25 °C) and the other 18 at 4 °C. The stability of the CBC parameters was determined by comparing the results to the 0 h sample. White Blood Cell Count (WBC), Hemoglobin (Hb) concentration, Red Blood Cell (RBC) count, Hematocrit (Hct), Mean Corpuscular Volume (MCV), Mean Corpuscular Hemoglobin (MCH), Mean Corpuscular Hemoglobin Concentration (MCHC), Cellular Hemoglobin Concentration Mean (CHCM), Mean Platelet Volume (MPV), RBC distribution width (RDW), Platelet (PLT) and Reticulocyte counts were studied. RESULTS: Most parameters were stable for 24 h at 4 °C, except for MCV, MCHC, CHCM and MPV. MCV and MPV increased after 8 h (p<0.0001), whereas MCHC and CHCM decreased significantly after 8 h. Significant changes were found for MCHC in samples kept at 4 °C for 48 h (p=0.002), and for CHCM kept for 72 h (p<0.001). Reticulocyte count stability was maintained for 24 h at 4 °C (p=0.3047). In samples kept at room temperature changes occurred after 8 h in RBC, Hct, MCV, MCH, MCHC, CHCM, MPV and PLT. CONCLUSION: CBC measurements are reliable for 8 h when samples are stored at room temperature. The only parameter stable for 72 h at room temperature was Hb. Blood samples kept at 4 °C for up to 24 h are suitable for haematological analysis. Elsevier 2015-12-07 /pmc/articles/PMC5574499/ /pubmed/28856188 http://dx.doi.org/10.1016/j.plabm.2015.12.001 Text en © 2015 The Authors http://creativecommons.org/licenses/by-nc-nd/4.0/ This is an open access article under the CC BY-NC-ND license (http://creativecommons.org/licenses/by-nc-nd/4.0/).
spellingShingle Research Article
Pintér, Erzsébet
László, Kinga
Schüszler, Ildikó
Konderák, Judit
The stability of quantitative blood count parameters using the ADVIA 2120i hematology analyzer
title The stability of quantitative blood count parameters using the ADVIA 2120i hematology analyzer
title_full The stability of quantitative blood count parameters using the ADVIA 2120i hematology analyzer
title_fullStr The stability of quantitative blood count parameters using the ADVIA 2120i hematology analyzer
title_full_unstemmed The stability of quantitative blood count parameters using the ADVIA 2120i hematology analyzer
title_short The stability of quantitative blood count parameters using the ADVIA 2120i hematology analyzer
title_sort stability of quantitative blood count parameters using the advia 2120i hematology analyzer
topic Research Article
url https://www.ncbi.nlm.nih.gov/pmc/articles/PMC5574499/
https://www.ncbi.nlm.nih.gov/pubmed/28856188
http://dx.doi.org/10.1016/j.plabm.2015.12.001
work_keys_str_mv AT pintererzsebet thestabilityofquantitativebloodcountparametersusingtheadvia2120ihematologyanalyzer
AT laszlokinga thestabilityofquantitativebloodcountparametersusingtheadvia2120ihematologyanalyzer
AT schuszlerildiko thestabilityofquantitativebloodcountparametersusingtheadvia2120ihematologyanalyzer
AT konderakjudit thestabilityofquantitativebloodcountparametersusingtheadvia2120ihematologyanalyzer
AT pintererzsebet stabilityofquantitativebloodcountparametersusingtheadvia2120ihematologyanalyzer
AT laszlokinga stabilityofquantitativebloodcountparametersusingtheadvia2120ihematologyanalyzer
AT schuszlerildiko stabilityofquantitativebloodcountparametersusingtheadvia2120ihematologyanalyzer
AT konderakjudit stabilityofquantitativebloodcountparametersusingtheadvia2120ihematologyanalyzer