Evaluation of Bio-Rad D-100 HbA1c analyzer against Tosoh G8 and Menarini HA-8180V
OBJECTIVES: To evaluate the Bio-Rad D-100®, an HPLC analyzer for glycated hemoglobin (HbA(1c)) determination, and to compare its performance with the Menarini HA-8180V® and Sysmex G8®. METHODS: Method comparison was performed according to Clinical and Laboratory Standards Institute (CLSI) EP9-A2 gui...
Autores principales: | , , , , , |
---|---|
Formato: | Online Artículo Texto |
Lenguaje: | English |
Publicado: |
Elsevier
2016
|
Materias: | |
Acceso en línea: | https://www.ncbi.nlm.nih.gov/pmc/articles/PMC5574509/ https://www.ncbi.nlm.nih.gov/pubmed/28856205 http://dx.doi.org/10.1016/j.plabm.2016.05.002 |
_version_ | 1783259854075330560 |
---|---|
author | Maesa, José María Fernández-Riejos, Patricia Mora, Catalina Sánchez de Toro, María Valladares, Paloma Menéndez González-Rodriguez, Concepción |
author_facet | Maesa, José María Fernández-Riejos, Patricia Mora, Catalina Sánchez de Toro, María Valladares, Paloma Menéndez González-Rodriguez, Concepción |
author_sort | Maesa, José María |
collection | PubMed |
description | OBJECTIVES: To evaluate the Bio-Rad D-100®, an HPLC analyzer for glycated hemoglobin (HbA(1c)) determination, and to compare its performance with the Menarini HA-8180V® and Sysmex G8®. METHODS: Method comparison was performed according to Clinical and Laboratory Standards Institute (CLSI) EP9-A2 guidelines. We selected 100 samples from the routine laboratory workload and analyzed them in duplicate with the three analyzers. The imprecision study was performed according to CLSI EP5-A2 guidelines for both inter-assay and intra-assay variability. Bias was assessed with external quality control material. To establish linearity, CLSI EP6-A protocol was followed. RESULTS: Method comparison (95% confidence intervals in parentheses): D-100 vs G8: Passing-Bablok regression; y=0.973(0.963–0.983)×−0.07(−0.07−0.069); r=0.9989. Bland-Altman mean difference: −0.229%HbA(1c) (−0.256: −0.202); Relative bias plot: D-100/G8 vs D100-G8 mean ratio=0.971(0.967−0.975). D-100 vs HA-8180V: Passing-Bablok regression; y=0.944(0.932–0.958)×−0.078(0.024−0.173); r=0.9989. Bland-Altman mean difference: −0.363%HbA(1c) (−0.401: −0.325); Relative bias plot D-100/HA-8180V vs D100-HA-8180V mean ratio=0.955(0.952−0.958). Inter-assay coefficient of variation (CV): 0.81%. Intra-assay CV: 1.04% (low level), and 0.78% (high level). Bias against target value=2.332%. Linearity: r(2)=0.998 in the concentration range 4.4−13.9%HbA(1c). Carry-over: 0.0024%. CONCLUSIONS: The Bio-Rad D-100 shows good correlation with G8 and HA-8180V. There is a small proportional systematic difference (2.7% and 5.6%, respectively) in both comparisons. Inter and intra-assay CVs are both lower than the lowest CV obtained in studies performed with D-100 and other instruments. |
format | Online Article Text |
id | pubmed-5574509 |
institution | National Center for Biotechnology Information |
language | English |
publishDate | 2016 |
publisher | Elsevier |
record_format | MEDLINE/PubMed |
spelling | pubmed-55745092017-08-30 Evaluation of Bio-Rad D-100 HbA1c analyzer against Tosoh G8 and Menarini HA-8180V Maesa, José María Fernández-Riejos, Patricia Mora, Catalina Sánchez de Toro, María Valladares, Paloma Menéndez González-Rodriguez, Concepción Pract Lab Med Article OBJECTIVES: To evaluate the Bio-Rad D-100®, an HPLC analyzer for glycated hemoglobin (HbA(1c)) determination, and to compare its performance with the Menarini HA-8180V® and Sysmex G8®. METHODS: Method comparison was performed according to Clinical and Laboratory Standards Institute (CLSI) EP9-A2 guidelines. We selected 100 samples from the routine laboratory workload and analyzed them in duplicate with the three analyzers. The imprecision study was performed according to CLSI EP5-A2 guidelines for both inter-assay and intra-assay variability. Bias was assessed with external quality control material. To establish linearity, CLSI EP6-A protocol was followed. RESULTS: Method comparison (95% confidence intervals in parentheses): D-100 vs G8: Passing-Bablok regression; y=0.973(0.963–0.983)×−0.07(−0.07−0.069); r=0.9989. Bland-Altman mean difference: −0.229%HbA(1c) (−0.256: −0.202); Relative bias plot: D-100/G8 vs D100-G8 mean ratio=0.971(0.967−0.975). D-100 vs HA-8180V: Passing-Bablok regression; y=0.944(0.932–0.958)×−0.078(0.024−0.173); r=0.9989. Bland-Altman mean difference: −0.363%HbA(1c) (−0.401: −0.325); Relative bias plot D-100/HA-8180V vs D100-HA-8180V mean ratio=0.955(0.952−0.958). Inter-assay coefficient of variation (CV): 0.81%. Intra-assay CV: 1.04% (low level), and 0.78% (high level). Bias against target value=2.332%. Linearity: r(2)=0.998 in the concentration range 4.4−13.9%HbA(1c). Carry-over: 0.0024%. CONCLUSIONS: The Bio-Rad D-100 shows good correlation with G8 and HA-8180V. There is a small proportional systematic difference (2.7% and 5.6%, respectively) in both comparisons. Inter and intra-assay CVs are both lower than the lowest CV obtained in studies performed with D-100 and other instruments. Elsevier 2016-05-14 /pmc/articles/PMC5574509/ /pubmed/28856205 http://dx.doi.org/10.1016/j.plabm.2016.05.002 Text en © 2016 The Authors http://creativecommons.org/licenses/by-nc-nd/4.0/ This is an open access article under the CC BY-NC-ND license (http://creativecommons.org/licenses/by-nc-nd/4.0/). |
spellingShingle | Article Maesa, José María Fernández-Riejos, Patricia Mora, Catalina Sánchez de Toro, María Valladares, Paloma Menéndez González-Rodriguez, Concepción Evaluation of Bio-Rad D-100 HbA1c analyzer against Tosoh G8 and Menarini HA-8180V |
title | Evaluation of Bio-Rad D-100 HbA1c analyzer against Tosoh G8 and Menarini HA-8180V |
title_full | Evaluation of Bio-Rad D-100 HbA1c analyzer against Tosoh G8 and Menarini HA-8180V |
title_fullStr | Evaluation of Bio-Rad D-100 HbA1c analyzer against Tosoh G8 and Menarini HA-8180V |
title_full_unstemmed | Evaluation of Bio-Rad D-100 HbA1c analyzer against Tosoh G8 and Menarini HA-8180V |
title_short | Evaluation of Bio-Rad D-100 HbA1c analyzer against Tosoh G8 and Menarini HA-8180V |
title_sort | evaluation of bio-rad d-100 hba1c analyzer against tosoh g8 and menarini ha-8180v |
topic | Article |
url | https://www.ncbi.nlm.nih.gov/pmc/articles/PMC5574509/ https://www.ncbi.nlm.nih.gov/pubmed/28856205 http://dx.doi.org/10.1016/j.plabm.2016.05.002 |
work_keys_str_mv | AT maesajosemaria evaluationofbioradd100hba1canalyzeragainsttosohg8andmenariniha8180v AT fernandezriejospatricia evaluationofbioradd100hba1canalyzeragainsttosohg8andmenariniha8180v AT moracatalinasanchez evaluationofbioradd100hba1canalyzeragainsttosohg8andmenariniha8180v AT detoromaria evaluationofbioradd100hba1canalyzeragainsttosohg8andmenariniha8180v AT valladarespalomamenendez evaluationofbioradd100hba1canalyzeragainsttosohg8andmenariniha8180v AT gonzalezrodriguezconcepcion evaluationofbioradd100hba1canalyzeragainsttosohg8andmenariniha8180v |