Lactate point-of-care testing for acidosis: Cross-comparison of two devices with routine laboratory results

OBJECTIVES: Lactate is a major parameter in medical decision making. During labor, it is an indicator for fetal acidosis and immediate intervention. In the Emergency Department (ED), rapid analysis of lactate/blood gas is crucial for optimal patient care. Our objectives were to cross-compare-for the...

Descripción completa

Detalles Bibliográficos
Autores principales: van Horssen, Remco, Schuurman, Teska N., de Groot, Monique J.M., Jakobs, Bernadette S.
Formato: Online Artículo Texto
Lenguaje:English
Publicado: Elsevier 2015
Materias:
Acceso en línea:https://www.ncbi.nlm.nih.gov/pmc/articles/PMC5574518/
https://www.ncbi.nlm.nih.gov/pubmed/28856192
http://dx.doi.org/10.1016/j.plabm.2015.12.005
Descripción
Sumario:OBJECTIVES: Lactate is a major parameter in medical decision making. During labor, it is an indicator for fetal acidosis and immediate intervention. In the Emergency Department (ED), rapid analysis of lactate/blood gas is crucial for optimal patient care. Our objectives were to cross-compare-for the first time-two point-of-care testing (POCT) lactate devices with routine laboratory results using novel tight precision targets and evaluate different lactate cut-off concentrations to predict metabolic acidosis. DESIGN AND METHODS: Blood samples from the delivery room (n=66) and from the ED (n=85) were analyzed on two POCT devices, the StatStrip-Lactate (Nova Biomedical) and the iSTAT-1 (CG4+ cassettes, Abbott), and compared to the routine laboratory analyzer (ABL-735, Radiometer). Lactate concentrations were cross-compared between these analyzers. RESULTS: The StatStrip correlated well with the ABL-735 (R=0.9737) and with the iSTAT-1 (R=0.9774) for lactate in umbilical cord blood. Lactate concentrations in ED samples measured on the iSTAT-1 and ABL-735 showed a correlation coefficient of R=0.9953. Analytical imprecision was excellent for lactate and pH, while for pO(2) and pCO(2) the coefficient of variation was relatively high using the iSTAT-1. CONCLUSION: Both POCT devices showed adequate analytical performance to measure lactate. The StatStrip can indicate metabolic acidosis in 1 μl blood and will be implemented at the delivery room.