Cargando…

A new sensitive automated assay for procalcitonin detection: LIAISON(®) BRAHMS PCT(®) II GEN

OBJECTIVES: To assess the performance of LIAISON(®) BRAHMS PCT(®) II GEN (DiaSorin, Saluggia, Italy) in procalcitonin (PCT) determination by comparing it to the assay reference method B·R·A·H·M·S PCT KRYPTOR (Thermo Fisher Scientific Clinical Diagnostics, Hennigsdorf, Germany) and assessing its abil...

Descripción completa

Detalles Bibliográficos
Autor principal: Fortunato, Antonio
Formato: Online Artículo Texto
Lenguaje:English
Publicado: Elsevier 2016
Materias:
Acceso en línea:https://www.ncbi.nlm.nih.gov/pmc/articles/PMC5574861/
https://www.ncbi.nlm.nih.gov/pubmed/28856207
http://dx.doi.org/10.1016/j.plabm.2016.06.002
_version_ 1783259919326117888
author Fortunato, Antonio
author_facet Fortunato, Antonio
author_sort Fortunato, Antonio
collection PubMed
description OBJECTIVES: To assess the performance of LIAISON(®) BRAHMS PCT(®) II GEN (DiaSorin, Saluggia, Italy) in procalcitonin (PCT) determination by comparing it to the assay reference method B·R·A·H·M·S PCT KRYPTOR (Thermo Fisher Scientific Clinical Diagnostics, Hennigsdorf, Germany) and assessing its ability to discriminate between healthy subjects and patients with suspected infection. METHODS: Diagnostic performance was evaluated on: a) 193 selected samples covering the assay range, whose procalcitonin levels were already evaluated with the B·R·A·H·M·S PCT(®) KRYPTOR; b) prospective samples: 150 apparently healthy specimens obtained from a blood bank, 161 hospitalized patients (not with specific pathologies), 243 apparently healthy children. RESULTS: The comparison of LIAISON(®) BRAHMS PCT(®) II GEN to the reference method B·R·A·H·M·S PCT KRYPTOR yielded high correlation coefficients: slope of Deming fit equal to 1.04 (95% CI: 0.99–1.09) with an intercept equal to 0.05 (95% CI: −0.09 to 0.19) and a high concordance (98.4% (95% CI: 95.5–99.7%)) at the 0.5 ng/mL cut-off. Moreover, the results obtained using prospective samples showed: (i) no samples with PCT concentration >0.5 ng/mL (cut-off) for the apparently healthy adults (highest value=0.033 ng/mL, 95th percentile and 97.5th percentile <0.02 ng/mL); (ii) 2 samples >0.5 ng/mL for hospitalized patients (highest value=0.715 ng/mL, 95th percentile: 0.054 ng/mL; 97.5th percentile: 0.088 ng/mL); (iii) 3 samples >0.5 ng/mL for the healthy children population (highest value=0.713 ng/mL, 95th percentile: 0.155 ng/mL; 97.5th percentile: 0.275 ng/mL). CONCLUSION: The fully automated LIAISON(®) BRAHMS PCT(®) II GEN agrees well with the reference method and is suitable for early diagnosis of sepsis, severe bacterial infection and guiding antibiotic therapy.
format Online
Article
Text
id pubmed-5574861
institution National Center for Biotechnology Information
language English
publishDate 2016
publisher Elsevier
record_format MEDLINE/PubMed
spelling pubmed-55748612017-08-30 A new sensitive automated assay for procalcitonin detection: LIAISON(®) BRAHMS PCT(®) II GEN Fortunato, Antonio Pract Lab Med Article OBJECTIVES: To assess the performance of LIAISON(®) BRAHMS PCT(®) II GEN (DiaSorin, Saluggia, Italy) in procalcitonin (PCT) determination by comparing it to the assay reference method B·R·A·H·M·S PCT KRYPTOR (Thermo Fisher Scientific Clinical Diagnostics, Hennigsdorf, Germany) and assessing its ability to discriminate between healthy subjects and patients with suspected infection. METHODS: Diagnostic performance was evaluated on: a) 193 selected samples covering the assay range, whose procalcitonin levels were already evaluated with the B·R·A·H·M·S PCT(®) KRYPTOR; b) prospective samples: 150 apparently healthy specimens obtained from a blood bank, 161 hospitalized patients (not with specific pathologies), 243 apparently healthy children. RESULTS: The comparison of LIAISON(®) BRAHMS PCT(®) II GEN to the reference method B·R·A·H·M·S PCT KRYPTOR yielded high correlation coefficients: slope of Deming fit equal to 1.04 (95% CI: 0.99–1.09) with an intercept equal to 0.05 (95% CI: −0.09 to 0.19) and a high concordance (98.4% (95% CI: 95.5–99.7%)) at the 0.5 ng/mL cut-off. Moreover, the results obtained using prospective samples showed: (i) no samples with PCT concentration >0.5 ng/mL (cut-off) for the apparently healthy adults (highest value=0.033 ng/mL, 95th percentile and 97.5th percentile <0.02 ng/mL); (ii) 2 samples >0.5 ng/mL for hospitalized patients (highest value=0.715 ng/mL, 95th percentile: 0.054 ng/mL; 97.5th percentile: 0.088 ng/mL); (iii) 3 samples >0.5 ng/mL for the healthy children population (highest value=0.713 ng/mL, 95th percentile: 0.155 ng/mL; 97.5th percentile: 0.275 ng/mL). CONCLUSION: The fully automated LIAISON(®) BRAHMS PCT(®) II GEN agrees well with the reference method and is suitable for early diagnosis of sepsis, severe bacterial infection and guiding antibiotic therapy. Elsevier 2016-06-27 /pmc/articles/PMC5574861/ /pubmed/28856207 http://dx.doi.org/10.1016/j.plabm.2016.06.002 Text en © 2016 The Author http://creativecommons.org/licenses/by/4.0/ This is an open access article under the CC BY license (http://creativecommons.org/licenses/by/4.0/).
spellingShingle Article
Fortunato, Antonio
A new sensitive automated assay for procalcitonin detection: LIAISON(®) BRAHMS PCT(®) II GEN
title A new sensitive automated assay for procalcitonin detection: LIAISON(®) BRAHMS PCT(®) II GEN
title_full A new sensitive automated assay for procalcitonin detection: LIAISON(®) BRAHMS PCT(®) II GEN
title_fullStr A new sensitive automated assay for procalcitonin detection: LIAISON(®) BRAHMS PCT(®) II GEN
title_full_unstemmed A new sensitive automated assay for procalcitonin detection: LIAISON(®) BRAHMS PCT(®) II GEN
title_short A new sensitive automated assay for procalcitonin detection: LIAISON(®) BRAHMS PCT(®) II GEN
title_sort new sensitive automated assay for procalcitonin detection: liaison(®) brahms pct(®) ii gen
topic Article
url https://www.ncbi.nlm.nih.gov/pmc/articles/PMC5574861/
https://www.ncbi.nlm.nih.gov/pubmed/28856207
http://dx.doi.org/10.1016/j.plabm.2016.06.002
work_keys_str_mv AT fortunatoantonio anewsensitiveautomatedassayforprocalcitonindetectionliaisonbrahmspctiigen
AT fortunatoantonio newsensitiveautomatedassayforprocalcitonindetectionliaisonbrahmspctiigen