Cargando…

The Roche Total Mycophenolic Acid® assay: An application protocol for the ABX Pentra 400 analyzer and comparison with LC–MS in children with idiopathic nephrotic syndrome

BACKGROUND: For TDM of mycophenolate acid (MPA), the Roche Total Mycophenolic Acid® assay based on the inhibition of recombinant inosine monophosphate dehydrogenase (IMPDH) has been shown to be a simple and reliable alternative to chromatographic methods. We have adapted this assay on the ABX Pentra...

Descripción completa

Detalles Bibliográficos
Autores principales: Parant, François, Ranchin, Bruno, Gagnieu, Marie-Claude
Formato: Online Artículo Texto
Lenguaje:English
Publicado: Elsevier 2017
Materias:
Acceso en línea:https://www.ncbi.nlm.nih.gov/pmc/articles/PMC5575364/
https://www.ncbi.nlm.nih.gov/pubmed/28856214
http://dx.doi.org/10.1016/j.plabm.2016.12.002
Descripción
Sumario:BACKGROUND: For TDM of mycophenolate acid (MPA), the Roche Total Mycophenolic Acid® assay based on the inhibition of recombinant inosine monophosphate dehydrogenase (IMPDH) has been shown to be a simple and reliable alternative to chromatographic methods. We have adapted this assay on the ABX Pentra 400 analyzer (HORIBA). OBJECTIVE: To investigate the analytical performances of the Roche Total Mycophenolic Acid® assay on the ABX Pentra 400 and to compare it to an LC-MS method using samples from children with nephrotic syndrome treated with mycophenolate mofetil (MMF). MATERIAL AND METHODS: Configuration of the open-channel on the ABX Pentra 400 was based on the Roche MPA assay package insert. Precision was determined as described in the CLSI protocol EP5-A2. Comparison with the LC-MS method was performed using 356 plasma samples from 42 children with nephrotic syndrome (8 h pharmacokinetic profiles). RESULTS: The enzymatic assay demonstrated high precision. The %CV for Within Run Imprecision ranged from 5.5% at 1.2 mg/L to 1.5% at 14.1 mg/L and Total Imprecision ranged from 9.3% to 2.5%. The method comparison with plasma samples from children yielded overall a good correlation and a good agreement between both methods. The Passing Bablok regression analysis showed the following results: [Roche MPA assay]=1.058 [MPA LC-MS] −0.06; rho=0.996. CONCLUSION: The Roche Total Mycophenolic Acid® assay is adaptable to the ABX Pentra 400 analyzer, and demonstrates accurate and precise measurement of MPA in plasma obtained from children with nephrotic syndrome.