Precisão e confiabilidade de um teste imuno-cromatográfico rápido NS1 para diagnóstico DENV-1 no ponto de atendimento e no laboratório

BACKGROUND: Rapid immunochromatographic tests (ICT) for dengue non-structural protein 1 (NS1) have shown good performance for diagnosing acute-phase dengue in serum in laboratory settings, but rarely have been assessed in whole blood and at point of care (POC). This study compare the accuracy and in...

Descripción completa

Detalles Bibliográficos
Autores principales: Mata, Verónica Elizabeth, Passos, Sonia Regina Lambert, Hökerberg, Yara Hahr Marques, Berardinelli, Guilherme Miguéis, dos Santos, Maria Angelica Borges, Fukuoka, Levy Vilas Boas, Maciel, Anna Carolina Fontoura Seixas Rangel, dos Santos Rodrigues, Cintia Damasceno, da Silva Santos, Aline, de Vasconcellos Carvalhaes de Oliveira, Raquel
Formato: Online Artículo Texto
Lenguaje:English
Publicado: BioMed Central 2017
Materias:
Acceso en línea:https://www.ncbi.nlm.nih.gov/pmc/articles/PMC5576130/
https://www.ncbi.nlm.nih.gov/pubmed/28851293
http://dx.doi.org/10.1186/s12879-017-2679-z
Descripción
Sumario:BACKGROUND: Rapid immunochromatographic tests (ICT) for dengue non-structural protein 1 (NS1) have shown good performance for diagnosing acute-phase dengue in serum in laboratory settings, but rarely have been assessed in whole blood and at point of care (POC). This study compare the accuracy and inter- and intra-observer reliability of the NS1 Bioeasy™ ICT in whole blood at POC versus serum in the laboratory, during a DENV-1 epidemic. METHODS: Cross-sectional study involving 144 adults spontaneously demanding care in an emergency department within 4 days of onset of acute febrile illness. Accuracy of NS1 Bioeasy™ ICT was compared in whole blood and serum, both at 15 and 30 min, blinded to the reference RT-PCR or NS1 ELISA. Non-dengue patients were also tested for Zika virus with RT-PCR. Reliability of whole blood and serum readings by the same or different observers was measured by simple kappa (95% CI). RESULTS: At 15 min, sensitivity (Sn) of NS1 Bioeasy™ ICT in whole blood/POC was 76.7% (95% CI: 68.0–84.1) and specificity (Sp) was 87.0% (95% CI: 66.4–97.2). Sn in serum/laboratory was 82% (95% CI: 74.1–88.6) and Sp 100% (95% CI: 85.8–100). Positive likelihood ratio was 5.9 (95% CI: 2.0–17.0) for whole blood/POC and 19.8 (95% CI: 2.9–135.1) for serum/laboratory. Reliability of matched readings of whole blood/POC and serum/laboratory by the same observer (k = 0.83, 95% CI: 0.74–0.93) or different observers (k = 0.81, 95% CI: 0.72–0.92) was almost perfect, with higher discordant levels in the absence of dengue. Results did not differ statistically at 5%. CONCLUSIONS: NS1 Bioeasy™ ICT in DENV-1 epidemics is a potentially confirmatory test. Invalid results at 15 min should be reread at 30 min. To optimize impact of implementing ICT in the management of false-negatives it should be incorporated into an algorithm according to setting and available specimen. TRIAL REGISTRATION: UTN U1111-1145-9451.