Accuracy of bone mineral density quantification using dual-layer spectral detector CT: a phantom study
OBJECTIVES: To investigate the accuracy of bone mineral density (BMD) quantification using dual-layer spectral detector CT (SDCT) at various scan protocols. METHODS: Two validated anthropomorphic phantoms containing inserts of 50–200 mg/cm(3) calcium hydroxyapatite (HA) were scanned using a 64-slice...
Autores principales: | , , , , , , , , |
---|---|
Formato: | Online Artículo Texto |
Lenguaje: | English |
Publicado: |
Springer Berlin Heidelberg
2017
|
Materias: | |
Acceso en línea: | https://www.ncbi.nlm.nih.gov/pmc/articles/PMC5579207/ https://www.ncbi.nlm.nih.gov/pubmed/28374079 http://dx.doi.org/10.1007/s00330-017-4801-4 |
_version_ | 1783260664874139648 |
---|---|
author | van Hamersvelt, Robbert W. Schilham, Arnold M. R. Engelke, Klaus den Harder, Annemarie M. de Keizer, Bart Verhaar, Harald J. Leiner, Tim de Jong, Pim A. Willemink, Martin J. |
author_facet | van Hamersvelt, Robbert W. Schilham, Arnold M. R. Engelke, Klaus den Harder, Annemarie M. de Keizer, Bart Verhaar, Harald J. Leiner, Tim de Jong, Pim A. Willemink, Martin J. |
author_sort | van Hamersvelt, Robbert W. |
collection | PubMed |
description | OBJECTIVES: To investigate the accuracy of bone mineral density (BMD) quantification using dual-layer spectral detector CT (SDCT) at various scan protocols. METHODS: Two validated anthropomorphic phantoms containing inserts of 50–200 mg/cm(3) calcium hydroxyapatite (HA) were scanned using a 64-slice SDCT scanner at various acquisition protocols (120 and 140 kVp, and 50, 100 and 200 mAs). Regions of interest (ROIs) were placed in each insert and mean attenuation profiles at monochromatic energy levels (90–200 keV) were constructed. These profiles were fitted to attenuation profiles of pure HA and water to calculate HA concentrations. For comparison, one phantom was scanned using dual energy X-ray absorptiometry (DXA). RESULTS: At both 120 and 140 kVp, excellent correlations (R = 0.97, P < 0.001) were found between true and measured HA concentrations. Mean error for all measurements at 120 kVp was -5.6 ± 5.7 mg/cm(3) (-3.6 ± 3.2%) and at 140 kVp -2.4 ± 3.7 mg/cm(3) (-0.8 ± 2.8%). Mean measurement errors were smaller than 6% for all acquisition protocols. Strong linear correlations (R(2) ≥ 0.970, P < 0.001) with DXA were found. CONCLUSIONS: SDCT allows for accurate BMD quantification and potentially opens up the possibility for osteoporosis evaluation and opportunistic screening in patients undergoing SDCT for other clinical indications. However, patient studies are needed to extend and translate our findings. KEY POINTS: • Dual-layer spectral detector CT allows for accurate bone mineral density quantification. • BMD measurements on SDCT are strongly linearly correlated to DXA. • SDCT, acquired for several indications, may allow for evaluation of osteoporosis. • This potentially opens up the possibility for opportunistic osteoporosis screening. |
format | Online Article Text |
id | pubmed-5579207 |
institution | National Center for Biotechnology Information |
language | English |
publishDate | 2017 |
publisher | Springer Berlin Heidelberg |
record_format | MEDLINE/PubMed |
spelling | pubmed-55792072017-09-18 Accuracy of bone mineral density quantification using dual-layer spectral detector CT: a phantom study van Hamersvelt, Robbert W. Schilham, Arnold M. R. Engelke, Klaus den Harder, Annemarie M. de Keizer, Bart Verhaar, Harald J. Leiner, Tim de Jong, Pim A. Willemink, Martin J. Eur Radiol Musculoskeletal OBJECTIVES: To investigate the accuracy of bone mineral density (BMD) quantification using dual-layer spectral detector CT (SDCT) at various scan protocols. METHODS: Two validated anthropomorphic phantoms containing inserts of 50–200 mg/cm(3) calcium hydroxyapatite (HA) were scanned using a 64-slice SDCT scanner at various acquisition protocols (120 and 140 kVp, and 50, 100 and 200 mAs). Regions of interest (ROIs) were placed in each insert and mean attenuation profiles at monochromatic energy levels (90–200 keV) were constructed. These profiles were fitted to attenuation profiles of pure HA and water to calculate HA concentrations. For comparison, one phantom was scanned using dual energy X-ray absorptiometry (DXA). RESULTS: At both 120 and 140 kVp, excellent correlations (R = 0.97, P < 0.001) were found between true and measured HA concentrations. Mean error for all measurements at 120 kVp was -5.6 ± 5.7 mg/cm(3) (-3.6 ± 3.2%) and at 140 kVp -2.4 ± 3.7 mg/cm(3) (-0.8 ± 2.8%). Mean measurement errors were smaller than 6% for all acquisition protocols. Strong linear correlations (R(2) ≥ 0.970, P < 0.001) with DXA were found. CONCLUSIONS: SDCT allows for accurate BMD quantification and potentially opens up the possibility for osteoporosis evaluation and opportunistic screening in patients undergoing SDCT for other clinical indications. However, patient studies are needed to extend and translate our findings. KEY POINTS: • Dual-layer spectral detector CT allows for accurate bone mineral density quantification. • BMD measurements on SDCT are strongly linearly correlated to DXA. • SDCT, acquired for several indications, may allow for evaluation of osteoporosis. • This potentially opens up the possibility for opportunistic osteoporosis screening. Springer Berlin Heidelberg 2017-04-03 2017 /pmc/articles/PMC5579207/ /pubmed/28374079 http://dx.doi.org/10.1007/s00330-017-4801-4 Text en © The Author(s) 2017 Open Access This article is distributed under the terms of the Creative Commons Attribution 4.0 International License (http://creativecommons.org/licenses/by/4.0/), which permits unrestricted use, distribution, and reproduction in any medium, provided you give appropriate credit to the original author(s) and the source, provide a link to the Creative Commons license, and indicate if changes were made. |
spellingShingle | Musculoskeletal van Hamersvelt, Robbert W. Schilham, Arnold M. R. Engelke, Klaus den Harder, Annemarie M. de Keizer, Bart Verhaar, Harald J. Leiner, Tim de Jong, Pim A. Willemink, Martin J. Accuracy of bone mineral density quantification using dual-layer spectral detector CT: a phantom study |
title | Accuracy of bone mineral density quantification using dual-layer spectral detector CT: a phantom study |
title_full | Accuracy of bone mineral density quantification using dual-layer spectral detector CT: a phantom study |
title_fullStr | Accuracy of bone mineral density quantification using dual-layer spectral detector CT: a phantom study |
title_full_unstemmed | Accuracy of bone mineral density quantification using dual-layer spectral detector CT: a phantom study |
title_short | Accuracy of bone mineral density quantification using dual-layer spectral detector CT: a phantom study |
title_sort | accuracy of bone mineral density quantification using dual-layer spectral detector ct: a phantom study |
topic | Musculoskeletal |
url | https://www.ncbi.nlm.nih.gov/pmc/articles/PMC5579207/ https://www.ncbi.nlm.nih.gov/pubmed/28374079 http://dx.doi.org/10.1007/s00330-017-4801-4 |
work_keys_str_mv | AT vanhamersveltrobbertw accuracyofbonemineraldensityquantificationusingduallayerspectraldetectorctaphantomstudy AT schilhamarnoldmr accuracyofbonemineraldensityquantificationusingduallayerspectraldetectorctaphantomstudy AT engelkeklaus accuracyofbonemineraldensityquantificationusingduallayerspectraldetectorctaphantomstudy AT denharderannemariem accuracyofbonemineraldensityquantificationusingduallayerspectraldetectorctaphantomstudy AT dekeizerbart accuracyofbonemineraldensityquantificationusingduallayerspectraldetectorctaphantomstudy AT verhaarharaldj accuracyofbonemineraldensityquantificationusingduallayerspectraldetectorctaphantomstudy AT leinertim accuracyofbonemineraldensityquantificationusingduallayerspectraldetectorctaphantomstudy AT dejongpima accuracyofbonemineraldensityquantificationusingduallayerspectraldetectorctaphantomstudy AT willeminkmartinj accuracyofbonemineraldensityquantificationusingduallayerspectraldetectorctaphantomstudy |