Accuracy of bone mineral density quantification using dual-layer spectral detector CT: a phantom study

OBJECTIVES: To investigate the accuracy of bone mineral density (BMD) quantification using dual-layer spectral detector CT (SDCT) at various scan protocols. METHODS: Two validated anthropomorphic phantoms containing inserts of 50–200 mg/cm(3) calcium hydroxyapatite (HA) were scanned using a 64-slice...

Descripción completa

Detalles Bibliográficos
Autores principales: van Hamersvelt, Robbert W., Schilham, Arnold M. R., Engelke, Klaus, den Harder, Annemarie M., de Keizer, Bart, Verhaar, Harald J., Leiner, Tim, de Jong, Pim A., Willemink, Martin J.
Formato: Online Artículo Texto
Lenguaje:English
Publicado: Springer Berlin Heidelberg 2017
Materias:
Acceso en línea:https://www.ncbi.nlm.nih.gov/pmc/articles/PMC5579207/
https://www.ncbi.nlm.nih.gov/pubmed/28374079
http://dx.doi.org/10.1007/s00330-017-4801-4
_version_ 1783260664874139648
author van Hamersvelt, Robbert W.
Schilham, Arnold M. R.
Engelke, Klaus
den Harder, Annemarie M.
de Keizer, Bart
Verhaar, Harald J.
Leiner, Tim
de Jong, Pim A.
Willemink, Martin J.
author_facet van Hamersvelt, Robbert W.
Schilham, Arnold M. R.
Engelke, Klaus
den Harder, Annemarie M.
de Keizer, Bart
Verhaar, Harald J.
Leiner, Tim
de Jong, Pim A.
Willemink, Martin J.
author_sort van Hamersvelt, Robbert W.
collection PubMed
description OBJECTIVES: To investigate the accuracy of bone mineral density (BMD) quantification using dual-layer spectral detector CT (SDCT) at various scan protocols. METHODS: Two validated anthropomorphic phantoms containing inserts of 50–200 mg/cm(3) calcium hydroxyapatite (HA) were scanned using a 64-slice SDCT scanner at various acquisition protocols (120 and 140 kVp, and 50, 100 and 200 mAs). Regions of interest (ROIs) were placed in each insert and mean attenuation profiles at monochromatic energy levels (90–200 keV) were constructed. These profiles were fitted to attenuation profiles of pure HA and water to calculate HA concentrations. For comparison, one phantom was scanned using dual energy X-ray absorptiometry (DXA). RESULTS: At both 120 and 140 kVp, excellent correlations (R = 0.97, P < 0.001) were found between true and measured HA concentrations. Mean error for all measurements at 120 kVp was -5.6 ± 5.7 mg/cm(3) (-3.6 ± 3.2%) and at 140 kVp -2.4 ± 3.7 mg/cm(3) (-0.8 ± 2.8%). Mean measurement errors were smaller than 6% for all acquisition protocols. Strong linear correlations (R(2) ≥ 0.970, P < 0.001) with DXA were found. CONCLUSIONS: SDCT allows for accurate BMD quantification and potentially opens up the possibility for osteoporosis evaluation and opportunistic screening in patients undergoing SDCT for other clinical indications. However, patient studies are needed to extend and translate our findings. KEY POINTS: • Dual-layer spectral detector CT allows for accurate bone mineral density quantification. • BMD measurements on SDCT are strongly linearly correlated to DXA. • SDCT, acquired for several indications, may allow for evaluation of osteoporosis. • This potentially opens up the possibility for opportunistic osteoporosis screening.
format Online
Article
Text
id pubmed-5579207
institution National Center for Biotechnology Information
language English
publishDate 2017
publisher Springer Berlin Heidelberg
record_format MEDLINE/PubMed
spelling pubmed-55792072017-09-18 Accuracy of bone mineral density quantification using dual-layer spectral detector CT: a phantom study van Hamersvelt, Robbert W. Schilham, Arnold M. R. Engelke, Klaus den Harder, Annemarie M. de Keizer, Bart Verhaar, Harald J. Leiner, Tim de Jong, Pim A. Willemink, Martin J. Eur Radiol Musculoskeletal OBJECTIVES: To investigate the accuracy of bone mineral density (BMD) quantification using dual-layer spectral detector CT (SDCT) at various scan protocols. METHODS: Two validated anthropomorphic phantoms containing inserts of 50–200 mg/cm(3) calcium hydroxyapatite (HA) were scanned using a 64-slice SDCT scanner at various acquisition protocols (120 and 140 kVp, and 50, 100 and 200 mAs). Regions of interest (ROIs) were placed in each insert and mean attenuation profiles at monochromatic energy levels (90–200 keV) were constructed. These profiles were fitted to attenuation profiles of pure HA and water to calculate HA concentrations. For comparison, one phantom was scanned using dual energy X-ray absorptiometry (DXA). RESULTS: At both 120 and 140 kVp, excellent correlations (R = 0.97, P < 0.001) were found between true and measured HA concentrations. Mean error for all measurements at 120 kVp was -5.6 ± 5.7 mg/cm(3) (-3.6 ± 3.2%) and at 140 kVp -2.4 ± 3.7 mg/cm(3) (-0.8 ± 2.8%). Mean measurement errors were smaller than 6% for all acquisition protocols. Strong linear correlations (R(2) ≥ 0.970, P < 0.001) with DXA were found. CONCLUSIONS: SDCT allows for accurate BMD quantification and potentially opens up the possibility for osteoporosis evaluation and opportunistic screening in patients undergoing SDCT for other clinical indications. However, patient studies are needed to extend and translate our findings. KEY POINTS: • Dual-layer spectral detector CT allows for accurate bone mineral density quantification. • BMD measurements on SDCT are strongly linearly correlated to DXA. • SDCT, acquired for several indications, may allow for evaluation of osteoporosis. • This potentially opens up the possibility for opportunistic osteoporosis screening. Springer Berlin Heidelberg 2017-04-03 2017 /pmc/articles/PMC5579207/ /pubmed/28374079 http://dx.doi.org/10.1007/s00330-017-4801-4 Text en © The Author(s) 2017 Open Access This article is distributed under the terms of the Creative Commons Attribution 4.0 International License (http://creativecommons.org/licenses/by/4.0/), which permits unrestricted use, distribution, and reproduction in any medium, provided you give appropriate credit to the original author(s) and the source, provide a link to the Creative Commons license, and indicate if changes were made.
spellingShingle Musculoskeletal
van Hamersvelt, Robbert W.
Schilham, Arnold M. R.
Engelke, Klaus
den Harder, Annemarie M.
de Keizer, Bart
Verhaar, Harald J.
Leiner, Tim
de Jong, Pim A.
Willemink, Martin J.
Accuracy of bone mineral density quantification using dual-layer spectral detector CT: a phantom study
title Accuracy of bone mineral density quantification using dual-layer spectral detector CT: a phantom study
title_full Accuracy of bone mineral density quantification using dual-layer spectral detector CT: a phantom study
title_fullStr Accuracy of bone mineral density quantification using dual-layer spectral detector CT: a phantom study
title_full_unstemmed Accuracy of bone mineral density quantification using dual-layer spectral detector CT: a phantom study
title_short Accuracy of bone mineral density quantification using dual-layer spectral detector CT: a phantom study
title_sort accuracy of bone mineral density quantification using dual-layer spectral detector ct: a phantom study
topic Musculoskeletal
url https://www.ncbi.nlm.nih.gov/pmc/articles/PMC5579207/
https://www.ncbi.nlm.nih.gov/pubmed/28374079
http://dx.doi.org/10.1007/s00330-017-4801-4
work_keys_str_mv AT vanhamersveltrobbertw accuracyofbonemineraldensityquantificationusingduallayerspectraldetectorctaphantomstudy
AT schilhamarnoldmr accuracyofbonemineraldensityquantificationusingduallayerspectraldetectorctaphantomstudy
AT engelkeklaus accuracyofbonemineraldensityquantificationusingduallayerspectraldetectorctaphantomstudy
AT denharderannemariem accuracyofbonemineraldensityquantificationusingduallayerspectraldetectorctaphantomstudy
AT dekeizerbart accuracyofbonemineraldensityquantificationusingduallayerspectraldetectorctaphantomstudy
AT verhaarharaldj accuracyofbonemineraldensityquantificationusingduallayerspectraldetectorctaphantomstudy
AT leinertim accuracyofbonemineraldensityquantificationusingduallayerspectraldetectorctaphantomstudy
AT dejongpima accuracyofbonemineraldensityquantificationusingduallayerspectraldetectorctaphantomstudy
AT willeminkmartinj accuracyofbonemineraldensityquantificationusingduallayerspectraldetectorctaphantomstudy