Suture Line Dehiscence after Endovascular Aneurysm Repair in a Patient with Marfan Syndrome

A 68-year-old man with Marfan syndrome developed de-novo leakage after endovascular aneurysm repair of a Dacron graft. Findings at subsequent reoperation suggested that the rebound force on the endograft generated by marked longitudinal deformation in the Dacron graft may have placed stress on the s...

Descripción completa

Detalles Bibliográficos
Autores principales: Kasahara, Hirofumi, Haijima, Norimasa, Hachiya, Takashi
Formato: Online Artículo Texto
Lenguaje:English
Publicado: Japanese College of Angiology / The Japanese Society for Vascular Surgery / Japanese Society of Phlebology 2017
Materias:
Acceso en línea:https://www.ncbi.nlm.nih.gov/pmc/articles/PMC5579775/
https://www.ncbi.nlm.nih.gov/pubmed/29034043
http://dx.doi.org/10.3400/avd.cr.17-00011
Descripción
Sumario:A 68-year-old man with Marfan syndrome developed de-novo leakage after endovascular aneurysm repair of a Dacron graft. Findings at subsequent reoperation suggested that the rebound force on the endograft generated by marked longitudinal deformation in the Dacron graft may have placed stress on the suture line, leading to partial dehiscence. The Dacron graft seemed to provide a stable proximal landing zone. However, the strong tendency of some endograft devices to return to their original shape may apply stress that affected weakened regions of the native aorta.