Comparison of fit accuracy and torque maintenance of zirconia and titanium abutments for internal tri-channel and external-hex implant connections

PURPOSE: This in vitro study aimed to evaluate the effect of implant connection design (external vs. internal) on the fit discrepancy and torque loss of zirconia and titanium abutments. MATERIALS AND METHODS: Two regular platform dental implants, one with external connection (Brånemark, Nobel Biocar...

Descripción completa

Detalles Bibliográficos
Autores principales: Siadat, Hakimeh, Beyabanaki, Elaheh, Mousavi, Niloufar, Alikhasi, Marzieh
Formato: Online Artículo Texto
Lenguaje:English
Publicado: The Korean Academy of Prosthodontics 2017
Materias:
Acceso en línea:https://www.ncbi.nlm.nih.gov/pmc/articles/PMC5582093/
https://www.ncbi.nlm.nih.gov/pubmed/28874994
http://dx.doi.org/10.4047/jap.2017.9.4.271
_version_ 1783261127481753600
author Siadat, Hakimeh
Beyabanaki, Elaheh
Mousavi, Niloufar
Alikhasi, Marzieh
author_facet Siadat, Hakimeh
Beyabanaki, Elaheh
Mousavi, Niloufar
Alikhasi, Marzieh
author_sort Siadat, Hakimeh
collection PubMed
description PURPOSE: This in vitro study aimed to evaluate the effect of implant connection design (external vs. internal) on the fit discrepancy and torque loss of zirconia and titanium abutments. MATERIALS AND METHODS: Two regular platform dental implants, one with external connection (Brånemark, Nobel Biocare AB) and the other with internal connection (Noble Replace, Nobel Biocare AB), were selected. Seven titanium and seven customized zirconia abutments were used for each connection design. Measurements of geometry, marginal discrepancy, and rotational freedom were done using video measuring machine. To measure the torque loss, each abutment was torqued to 35 Ncm and then opened by means of a digital torque wrench. Data were analyzed with two-way ANOVA and t-test at α=0.05 of significance. RESULTS: There were significant differences in the geometrical measurements and rotational freedom between abutments of two connection groups (P<.001). Also, the results showed significant differences between titanium abutments of internal and external connection implants in terms of rotational freedom (P<.001). Not only customized internal abutments but also customized external abutments did not have the exact geometry of prefabricated abutments (P<.001). However, neither connection type (P=.15) nor abutment material (P=.38) affected torque loss. CONCLUSION: Abutments with internal connection showed less rotational freedom. However, better marginal fit was observed in externally connected abutments. Also, customized abutments with either connection could not duplicate the exact geometry of their corresponding prefabricated abutment. However, neither abutment connection nor material affected torque loss values.
format Online
Article
Text
id pubmed-5582093
institution National Center for Biotechnology Information
language English
publishDate 2017
publisher The Korean Academy of Prosthodontics
record_format MEDLINE/PubMed
spelling pubmed-55820932017-09-05 Comparison of fit accuracy and torque maintenance of zirconia and titanium abutments for internal tri-channel and external-hex implant connections Siadat, Hakimeh Beyabanaki, Elaheh Mousavi, Niloufar Alikhasi, Marzieh J Adv Prosthodont Original Article PURPOSE: This in vitro study aimed to evaluate the effect of implant connection design (external vs. internal) on the fit discrepancy and torque loss of zirconia and titanium abutments. MATERIALS AND METHODS: Two regular platform dental implants, one with external connection (Brånemark, Nobel Biocare AB) and the other with internal connection (Noble Replace, Nobel Biocare AB), were selected. Seven titanium and seven customized zirconia abutments were used for each connection design. Measurements of geometry, marginal discrepancy, and rotational freedom were done using video measuring machine. To measure the torque loss, each abutment was torqued to 35 Ncm and then opened by means of a digital torque wrench. Data were analyzed with two-way ANOVA and t-test at α=0.05 of significance. RESULTS: There were significant differences in the geometrical measurements and rotational freedom between abutments of two connection groups (P<.001). Also, the results showed significant differences between titanium abutments of internal and external connection implants in terms of rotational freedom (P<.001). Not only customized internal abutments but also customized external abutments did not have the exact geometry of prefabricated abutments (P<.001). However, neither connection type (P=.15) nor abutment material (P=.38) affected torque loss. CONCLUSION: Abutments with internal connection showed less rotational freedom. However, better marginal fit was observed in externally connected abutments. Also, customized abutments with either connection could not duplicate the exact geometry of their corresponding prefabricated abutment. However, neither abutment connection nor material affected torque loss values. The Korean Academy of Prosthodontics 2017-08 2017-08-16 /pmc/articles/PMC5582093/ /pubmed/28874994 http://dx.doi.org/10.4047/jap.2017.9.4.271 Text en © 2017 The Korean Academy of Prosthodontics http://creativecommons.org/licenses/by-nc/3.0/ This is an Open Access article distributed under the terms of the Creative Commons Attribution Non-Commercial License (http://creativecommons.org/licenses/by-nc/3.0) which permits unrestricted non-commercial use, distribution, and reproduction in any medium, provided the original work is properly cited.
spellingShingle Original Article
Siadat, Hakimeh
Beyabanaki, Elaheh
Mousavi, Niloufar
Alikhasi, Marzieh
Comparison of fit accuracy and torque maintenance of zirconia and titanium abutments for internal tri-channel and external-hex implant connections
title Comparison of fit accuracy and torque maintenance of zirconia and titanium abutments for internal tri-channel and external-hex implant connections
title_full Comparison of fit accuracy and torque maintenance of zirconia and titanium abutments for internal tri-channel and external-hex implant connections
title_fullStr Comparison of fit accuracy and torque maintenance of zirconia and titanium abutments for internal tri-channel and external-hex implant connections
title_full_unstemmed Comparison of fit accuracy and torque maintenance of zirconia and titanium abutments for internal tri-channel and external-hex implant connections
title_short Comparison of fit accuracy and torque maintenance of zirconia and titanium abutments for internal tri-channel and external-hex implant connections
title_sort comparison of fit accuracy and torque maintenance of zirconia and titanium abutments for internal tri-channel and external-hex implant connections
topic Original Article
url https://www.ncbi.nlm.nih.gov/pmc/articles/PMC5582093/
https://www.ncbi.nlm.nih.gov/pubmed/28874994
http://dx.doi.org/10.4047/jap.2017.9.4.271
work_keys_str_mv AT siadathakimeh comparisonoffitaccuracyandtorquemaintenanceofzirconiaandtitaniumabutmentsforinternaltrichannelandexternalheximplantconnections
AT beyabanakielaheh comparisonoffitaccuracyandtorquemaintenanceofzirconiaandtitaniumabutmentsforinternaltrichannelandexternalheximplantconnections
AT mousaviniloufar comparisonoffitaccuracyandtorquemaintenanceofzirconiaandtitaniumabutmentsforinternaltrichannelandexternalheximplantconnections
AT alikhasimarzieh comparisonoffitaccuracyandtorquemaintenanceofzirconiaandtitaniumabutmentsforinternaltrichannelandexternalheximplantconnections