T Wave Safety Margin during the Process of ICD Implantation As a Novel Predictor of T Wave Oversensing
Introduction: T wave oversensing (TWOS) is a major drawback of implantable cardioverter defibrillator (ICD) and data on predictors of TWOS in ICD is limited. We aimed to calculate a novel index of T wave safety margin (TWSM) and assess its potential for evaluating TWOS during the procedure of ICD im...
Autores principales: | , , , , , , , , , , , , , |
---|---|
Formato: | Online Artículo Texto |
Lenguaje: | English |
Publicado: |
Frontiers Media S.A.
2017
|
Materias: | |
Acceso en línea: | https://www.ncbi.nlm.nih.gov/pmc/articles/PMC5585188/ https://www.ncbi.nlm.nih.gov/pubmed/28919866 http://dx.doi.org/10.3389/fphys.2017.00659 |
_version_ | 1783261568552665088 |
---|---|
author | Sun, Ya-Xun Gao, Jing Jiang, Chen-Yang Xue, Yu-Mei Xu, Yi-Zhou Liu, Gang Guo, Ji-Hong Sheng, Xia Ye, Yang He, Hong Zhao, Yun-Tao Barajas-Martinez, Hector Fu, Guo-Sheng Hu, Dan |
author_facet | Sun, Ya-Xun Gao, Jing Jiang, Chen-Yang Xue, Yu-Mei Xu, Yi-Zhou Liu, Gang Guo, Ji-Hong Sheng, Xia Ye, Yang He, Hong Zhao, Yun-Tao Barajas-Martinez, Hector Fu, Guo-Sheng Hu, Dan |
author_sort | Sun, Ya-Xun |
collection | PubMed |
description | Introduction: T wave oversensing (TWOS) is a major drawback of implantable cardioverter defibrillator (ICD) and data on predictors of TWOS in ICD is limited. We aimed to calculate a novel index of T wave safety margin (TWSM) and assess its potential for evaluating TWOS during the procedure of ICD implantation. Methods and Results: Thirty-two consecutive patients with ICD implantation were enrolled. During each procedure of ICD implantation, different ICD generators were connected to implanted sensing lead through active-fixation leads and bridging cables. R and T wave amplitudes were measured on ICD printouts according to the gain. The ICDs were programed to the most sensitive settings to reveal possible TWOS. A novel index TWSM was calculated according to the corresponding sensing algorithm of ICD. There was discrepancy of R wave amplitudes measured by different ICDs (P < 0.01). In Fortify and Teligen ICDs, T wave amplitudes showed no difference (P > 0.05) and TWSMs were sufficiently high (post sensing: 13.0 ± 7.6 and 28.3 ± 16.5, respectively, post pacing: 5.0 ± 2.2 and 4.6 ± 0.9, respectively). In nine patients with 10 TWOS episodes detected during the procedure of ICD implantation, generators with the highest TWSM were chosen. Only one TWOS episode during pacing was recorded during the 25 ± 7 mo follow-up period. Conclusions: We first propose the index of TWSM during ICD implantation as a potentially efficient predictor for TWOS. Evaluation of TWSM might help to reduce TWOS episodes in patients with high risk of TWOS. Prospective studies are warranted to validate this index and its potential to reduce TWOS episodes. |
format | Online Article Text |
id | pubmed-5585188 |
institution | National Center for Biotechnology Information |
language | English |
publishDate | 2017 |
publisher | Frontiers Media S.A. |
record_format | MEDLINE/PubMed |
spelling | pubmed-55851882017-09-15 T Wave Safety Margin during the Process of ICD Implantation As a Novel Predictor of T Wave Oversensing Sun, Ya-Xun Gao, Jing Jiang, Chen-Yang Xue, Yu-Mei Xu, Yi-Zhou Liu, Gang Guo, Ji-Hong Sheng, Xia Ye, Yang He, Hong Zhao, Yun-Tao Barajas-Martinez, Hector Fu, Guo-Sheng Hu, Dan Front Physiol Physiology Introduction: T wave oversensing (TWOS) is a major drawback of implantable cardioverter defibrillator (ICD) and data on predictors of TWOS in ICD is limited. We aimed to calculate a novel index of T wave safety margin (TWSM) and assess its potential for evaluating TWOS during the procedure of ICD implantation. Methods and Results: Thirty-two consecutive patients with ICD implantation were enrolled. During each procedure of ICD implantation, different ICD generators were connected to implanted sensing lead through active-fixation leads and bridging cables. R and T wave amplitudes were measured on ICD printouts according to the gain. The ICDs were programed to the most sensitive settings to reveal possible TWOS. A novel index TWSM was calculated according to the corresponding sensing algorithm of ICD. There was discrepancy of R wave amplitudes measured by different ICDs (P < 0.01). In Fortify and Teligen ICDs, T wave amplitudes showed no difference (P > 0.05) and TWSMs were sufficiently high (post sensing: 13.0 ± 7.6 and 28.3 ± 16.5, respectively, post pacing: 5.0 ± 2.2 and 4.6 ± 0.9, respectively). In nine patients with 10 TWOS episodes detected during the procedure of ICD implantation, generators with the highest TWSM were chosen. Only one TWOS episode during pacing was recorded during the 25 ± 7 mo follow-up period. Conclusions: We first propose the index of TWSM during ICD implantation as a potentially efficient predictor for TWOS. Evaluation of TWSM might help to reduce TWOS episodes in patients with high risk of TWOS. Prospective studies are warranted to validate this index and its potential to reduce TWOS episodes. Frontiers Media S.A. 2017-09-01 /pmc/articles/PMC5585188/ /pubmed/28919866 http://dx.doi.org/10.3389/fphys.2017.00659 Text en Copyright © 2017 Sun, Gao, Jiang, Xue, Xu, Liu, Guo, Sheng, Ye, He, Zhao, Barajas-Martinez, Fu and Hu. http://creativecommons.org/licenses/by/4.0/ This is an open-access article distributed under the terms of the Creative Commons Attribution License (CC BY). The use, distribution or reproduction in other forums is permitted, provided the original author(s) or licensor are credited and that the original publication in this journal is cited, in accordance with accepted academic practice. No use, distribution or reproduction is permitted which does not comply with these terms. |
spellingShingle | Physiology Sun, Ya-Xun Gao, Jing Jiang, Chen-Yang Xue, Yu-Mei Xu, Yi-Zhou Liu, Gang Guo, Ji-Hong Sheng, Xia Ye, Yang He, Hong Zhao, Yun-Tao Barajas-Martinez, Hector Fu, Guo-Sheng Hu, Dan T Wave Safety Margin during the Process of ICD Implantation As a Novel Predictor of T Wave Oversensing |
title | T Wave Safety Margin during the Process of ICD Implantation As a Novel Predictor of T Wave Oversensing |
title_full | T Wave Safety Margin during the Process of ICD Implantation As a Novel Predictor of T Wave Oversensing |
title_fullStr | T Wave Safety Margin during the Process of ICD Implantation As a Novel Predictor of T Wave Oversensing |
title_full_unstemmed | T Wave Safety Margin during the Process of ICD Implantation As a Novel Predictor of T Wave Oversensing |
title_short | T Wave Safety Margin during the Process of ICD Implantation As a Novel Predictor of T Wave Oversensing |
title_sort | t wave safety margin during the process of icd implantation as a novel predictor of t wave oversensing |
topic | Physiology |
url | https://www.ncbi.nlm.nih.gov/pmc/articles/PMC5585188/ https://www.ncbi.nlm.nih.gov/pubmed/28919866 http://dx.doi.org/10.3389/fphys.2017.00659 |
work_keys_str_mv | AT sunyaxun twavesafetymarginduringtheprocessoficdimplantationasanovelpredictoroftwaveoversensing AT gaojing twavesafetymarginduringtheprocessoficdimplantationasanovelpredictoroftwaveoversensing AT jiangchenyang twavesafetymarginduringtheprocessoficdimplantationasanovelpredictoroftwaveoversensing AT xueyumei twavesafetymarginduringtheprocessoficdimplantationasanovelpredictoroftwaveoversensing AT xuyizhou twavesafetymarginduringtheprocessoficdimplantationasanovelpredictoroftwaveoversensing AT liugang twavesafetymarginduringtheprocessoficdimplantationasanovelpredictoroftwaveoversensing AT guojihong twavesafetymarginduringtheprocessoficdimplantationasanovelpredictoroftwaveoversensing AT shengxia twavesafetymarginduringtheprocessoficdimplantationasanovelpredictoroftwaveoversensing AT yeyang twavesafetymarginduringtheprocessoficdimplantationasanovelpredictoroftwaveoversensing AT hehong twavesafetymarginduringtheprocessoficdimplantationasanovelpredictoroftwaveoversensing AT zhaoyuntao twavesafetymarginduringtheprocessoficdimplantationasanovelpredictoroftwaveoversensing AT barajasmartinezhector twavesafetymarginduringtheprocessoficdimplantationasanovelpredictoroftwaveoversensing AT fuguosheng twavesafetymarginduringtheprocessoficdimplantationasanovelpredictoroftwaveoversensing AT hudan twavesafetymarginduringtheprocessoficdimplantationasanovelpredictoroftwaveoversensing |