T Wave Safety Margin during the Process of ICD Implantation As a Novel Predictor of T Wave Oversensing

Introduction: T wave oversensing (TWOS) is a major drawback of implantable cardioverter defibrillator (ICD) and data on predictors of TWOS in ICD is limited. We aimed to calculate a novel index of T wave safety margin (TWSM) and assess its potential for evaluating TWOS during the procedure of ICD im...

Descripción completa

Detalles Bibliográficos
Autores principales: Sun, Ya-Xun, Gao, Jing, Jiang, Chen-Yang, Xue, Yu-Mei, Xu, Yi-Zhou, Liu, Gang, Guo, Ji-Hong, Sheng, Xia, Ye, Yang, He, Hong, Zhao, Yun-Tao, Barajas-Martinez, Hector, Fu, Guo-Sheng, Hu, Dan
Formato: Online Artículo Texto
Lenguaje:English
Publicado: Frontiers Media S.A. 2017
Materias:
Acceso en línea:https://www.ncbi.nlm.nih.gov/pmc/articles/PMC5585188/
https://www.ncbi.nlm.nih.gov/pubmed/28919866
http://dx.doi.org/10.3389/fphys.2017.00659
_version_ 1783261568552665088
author Sun, Ya-Xun
Gao, Jing
Jiang, Chen-Yang
Xue, Yu-Mei
Xu, Yi-Zhou
Liu, Gang
Guo, Ji-Hong
Sheng, Xia
Ye, Yang
He, Hong
Zhao, Yun-Tao
Barajas-Martinez, Hector
Fu, Guo-Sheng
Hu, Dan
author_facet Sun, Ya-Xun
Gao, Jing
Jiang, Chen-Yang
Xue, Yu-Mei
Xu, Yi-Zhou
Liu, Gang
Guo, Ji-Hong
Sheng, Xia
Ye, Yang
He, Hong
Zhao, Yun-Tao
Barajas-Martinez, Hector
Fu, Guo-Sheng
Hu, Dan
author_sort Sun, Ya-Xun
collection PubMed
description Introduction: T wave oversensing (TWOS) is a major drawback of implantable cardioverter defibrillator (ICD) and data on predictors of TWOS in ICD is limited. We aimed to calculate a novel index of T wave safety margin (TWSM) and assess its potential for evaluating TWOS during the procedure of ICD implantation. Methods and Results: Thirty-two consecutive patients with ICD implantation were enrolled. During each procedure of ICD implantation, different ICD generators were connected to implanted sensing lead through active-fixation leads and bridging cables. R and T wave amplitudes were measured on ICD printouts according to the gain. The ICDs were programed to the most sensitive settings to reveal possible TWOS. A novel index TWSM was calculated according to the corresponding sensing algorithm of ICD. There was discrepancy of R wave amplitudes measured by different ICDs (P < 0.01). In Fortify and Teligen ICDs, T wave amplitudes showed no difference (P > 0.05) and TWSMs were sufficiently high (post sensing: 13.0 ± 7.6 and 28.3 ± 16.5, respectively, post pacing: 5.0 ± 2.2 and 4.6 ± 0.9, respectively). In nine patients with 10 TWOS episodes detected during the procedure of ICD implantation, generators with the highest TWSM were chosen. Only one TWOS episode during pacing was recorded during the 25 ± 7 mo follow-up period. Conclusions: We first propose the index of TWSM during ICD implantation as a potentially efficient predictor for TWOS. Evaluation of TWSM might help to reduce TWOS episodes in patients with high risk of TWOS. Prospective studies are warranted to validate this index and its potential to reduce TWOS episodes.
format Online
Article
Text
id pubmed-5585188
institution National Center for Biotechnology Information
language English
publishDate 2017
publisher Frontiers Media S.A.
record_format MEDLINE/PubMed
spelling pubmed-55851882017-09-15 T Wave Safety Margin during the Process of ICD Implantation As a Novel Predictor of T Wave Oversensing Sun, Ya-Xun Gao, Jing Jiang, Chen-Yang Xue, Yu-Mei Xu, Yi-Zhou Liu, Gang Guo, Ji-Hong Sheng, Xia Ye, Yang He, Hong Zhao, Yun-Tao Barajas-Martinez, Hector Fu, Guo-Sheng Hu, Dan Front Physiol Physiology Introduction: T wave oversensing (TWOS) is a major drawback of implantable cardioverter defibrillator (ICD) and data on predictors of TWOS in ICD is limited. We aimed to calculate a novel index of T wave safety margin (TWSM) and assess its potential for evaluating TWOS during the procedure of ICD implantation. Methods and Results: Thirty-two consecutive patients with ICD implantation were enrolled. During each procedure of ICD implantation, different ICD generators were connected to implanted sensing lead through active-fixation leads and bridging cables. R and T wave amplitudes were measured on ICD printouts according to the gain. The ICDs were programed to the most sensitive settings to reveal possible TWOS. A novel index TWSM was calculated according to the corresponding sensing algorithm of ICD. There was discrepancy of R wave amplitudes measured by different ICDs (P < 0.01). In Fortify and Teligen ICDs, T wave amplitudes showed no difference (P > 0.05) and TWSMs were sufficiently high (post sensing: 13.0 ± 7.6 and 28.3 ± 16.5, respectively, post pacing: 5.0 ± 2.2 and 4.6 ± 0.9, respectively). In nine patients with 10 TWOS episodes detected during the procedure of ICD implantation, generators with the highest TWSM were chosen. Only one TWOS episode during pacing was recorded during the 25 ± 7 mo follow-up period. Conclusions: We first propose the index of TWSM during ICD implantation as a potentially efficient predictor for TWOS. Evaluation of TWSM might help to reduce TWOS episodes in patients with high risk of TWOS. Prospective studies are warranted to validate this index and its potential to reduce TWOS episodes. Frontiers Media S.A. 2017-09-01 /pmc/articles/PMC5585188/ /pubmed/28919866 http://dx.doi.org/10.3389/fphys.2017.00659 Text en Copyright © 2017 Sun, Gao, Jiang, Xue, Xu, Liu, Guo, Sheng, Ye, He, Zhao, Barajas-Martinez, Fu and Hu. http://creativecommons.org/licenses/by/4.0/ This is an open-access article distributed under the terms of the Creative Commons Attribution License (CC BY). The use, distribution or reproduction in other forums is permitted, provided the original author(s) or licensor are credited and that the original publication in this journal is cited, in accordance with accepted academic practice. No use, distribution or reproduction is permitted which does not comply with these terms.
spellingShingle Physiology
Sun, Ya-Xun
Gao, Jing
Jiang, Chen-Yang
Xue, Yu-Mei
Xu, Yi-Zhou
Liu, Gang
Guo, Ji-Hong
Sheng, Xia
Ye, Yang
He, Hong
Zhao, Yun-Tao
Barajas-Martinez, Hector
Fu, Guo-Sheng
Hu, Dan
T Wave Safety Margin during the Process of ICD Implantation As a Novel Predictor of T Wave Oversensing
title T Wave Safety Margin during the Process of ICD Implantation As a Novel Predictor of T Wave Oversensing
title_full T Wave Safety Margin during the Process of ICD Implantation As a Novel Predictor of T Wave Oversensing
title_fullStr T Wave Safety Margin during the Process of ICD Implantation As a Novel Predictor of T Wave Oversensing
title_full_unstemmed T Wave Safety Margin during the Process of ICD Implantation As a Novel Predictor of T Wave Oversensing
title_short T Wave Safety Margin during the Process of ICD Implantation As a Novel Predictor of T Wave Oversensing
title_sort t wave safety margin during the process of icd implantation as a novel predictor of t wave oversensing
topic Physiology
url https://www.ncbi.nlm.nih.gov/pmc/articles/PMC5585188/
https://www.ncbi.nlm.nih.gov/pubmed/28919866
http://dx.doi.org/10.3389/fphys.2017.00659
work_keys_str_mv AT sunyaxun twavesafetymarginduringtheprocessoficdimplantationasanovelpredictoroftwaveoversensing
AT gaojing twavesafetymarginduringtheprocessoficdimplantationasanovelpredictoroftwaveoversensing
AT jiangchenyang twavesafetymarginduringtheprocessoficdimplantationasanovelpredictoroftwaveoversensing
AT xueyumei twavesafetymarginduringtheprocessoficdimplantationasanovelpredictoroftwaveoversensing
AT xuyizhou twavesafetymarginduringtheprocessoficdimplantationasanovelpredictoroftwaveoversensing
AT liugang twavesafetymarginduringtheprocessoficdimplantationasanovelpredictoroftwaveoversensing
AT guojihong twavesafetymarginduringtheprocessoficdimplantationasanovelpredictoroftwaveoversensing
AT shengxia twavesafetymarginduringtheprocessoficdimplantationasanovelpredictoroftwaveoversensing
AT yeyang twavesafetymarginduringtheprocessoficdimplantationasanovelpredictoroftwaveoversensing
AT hehong twavesafetymarginduringtheprocessoficdimplantationasanovelpredictoroftwaveoversensing
AT zhaoyuntao twavesafetymarginduringtheprocessoficdimplantationasanovelpredictoroftwaveoversensing
AT barajasmartinezhector twavesafetymarginduringtheprocessoficdimplantationasanovelpredictoroftwaveoversensing
AT fuguosheng twavesafetymarginduringtheprocessoficdimplantationasanovelpredictoroftwaveoversensing
AT hudan twavesafetymarginduringtheprocessoficdimplantationasanovelpredictoroftwaveoversensing