Comparisons of the Tono-Pen® and Goldmann Applanation Tonometer in the Measurement of Intraocular Pressure of Primary Open Angle Glaucoma Patients in a Hospital Population in Southwest Nigeria

OBJECTIVE: The aim of this study was to compare intraocular pressure (IOP) measured with the Tono-Pen® to that measured using the Goldmann applanation tonometer (GAT) in patients with primary open angle glaucoma attending a glaucoma clinic. SUBJECTS AND METHODS: A comparative clinic-based observatio...

Descripción completa

Detalles Bibliográficos
Autores principales: Onochie, Chizoba, Okoye, Obiekwe, Ogunro, Adunola, Aribaba, Tade, Hassan, Kunle, Onakoya, Adeola
Formato: Online Artículo Texto
Lenguaje:English
Publicado: S. Karger AG 2016
Materias:
Acceso en línea:https://www.ncbi.nlm.nih.gov/pmc/articles/PMC5588504/
https://www.ncbi.nlm.nih.gov/pubmed/27504638
http://dx.doi.org/10.1159/000448953
_version_ 1783262185821044736
author Onochie, Chizoba
Okoye, Obiekwe
Ogunro, Adunola
Aribaba, Tade
Hassan, Kunle
Onakoya, Adeola
author_facet Onochie, Chizoba
Okoye, Obiekwe
Ogunro, Adunola
Aribaba, Tade
Hassan, Kunle
Onakoya, Adeola
author_sort Onochie, Chizoba
collection PubMed
description OBJECTIVE: The aim of this study was to compare intraocular pressure (IOP) measured with the Tono-Pen® to that measured using the Goldmann applanation tonometer (GAT) in patients with primary open angle glaucoma attending a glaucoma clinic. SUBJECTS AND METHODS: A comparative clinic-based observational study was conducted involving 75 patients (39 men and 36 women) attending a glaucoma clinic in Southwest Nigeria. A pretested structured questionnaire was used to collect sociodemographic data. The Tono-Pen and the GAT were used to measure the IOP in each patient. Central corneal thickness (CCT) was measured with an ultrasonic pachymeter, and the corrected GAT value was calculated for each patient. The Tono-Pen readings were compared with the uncorrected and corrected GAT readings. Descriptive and comparative analyses were performed. Values for p < 0.05 were considered statistically significant. RESULTS: The mean age of the participants was 60.39 ± 16.71 years. The mean IOP using the Tono-Pen was 21.1 ± 6.8 mm Hg versus the mean uncorrected GAT value of 17.1 ± 6.9 mm Hg and the corrected GAT value of 18.9 ± 7.5 mm Hg. The mean CCT was 510.5 ± 29.6 µm. The mean differences between the Tono-Pen reading and uncorrected and corrected GAT readings were 3.9 ± 2.6 and 2.1 ± 3.5 mm Hg, respectively. Gender (Tono-Pen vs. GAT; p = 0.981 vs. 0.437) and corneal thickness (p = 0.057) did not significantly affect the IOP value. Of the 75 patients, 68 (90.7%) preferred the Tono-Pen to the GAT. CONCLUSION: In this study, the Tono-Pen gave a higher value for IOP than the uncorrected and corrected GAT values. Gender and corneal thickness did not significantly affect the measurements. Most patients found the Tono-Pen more acceptable than the GAT.
format Online
Article
Text
id pubmed-5588504
institution National Center for Biotechnology Information
language English
publishDate 2016
publisher S. Karger AG
record_format MEDLINE/PubMed
spelling pubmed-55885042017-11-01 Comparisons of the Tono-Pen® and Goldmann Applanation Tonometer in the Measurement of Intraocular Pressure of Primary Open Angle Glaucoma Patients in a Hospital Population in Southwest Nigeria Onochie, Chizoba Okoye, Obiekwe Ogunro, Adunola Aribaba, Tade Hassan, Kunle Onakoya, Adeola Med Princ Pract Original Paper OBJECTIVE: The aim of this study was to compare intraocular pressure (IOP) measured with the Tono-Pen® to that measured using the Goldmann applanation tonometer (GAT) in patients with primary open angle glaucoma attending a glaucoma clinic. SUBJECTS AND METHODS: A comparative clinic-based observational study was conducted involving 75 patients (39 men and 36 women) attending a glaucoma clinic in Southwest Nigeria. A pretested structured questionnaire was used to collect sociodemographic data. The Tono-Pen and the GAT were used to measure the IOP in each patient. Central corneal thickness (CCT) was measured with an ultrasonic pachymeter, and the corrected GAT value was calculated for each patient. The Tono-Pen readings were compared with the uncorrected and corrected GAT readings. Descriptive and comparative analyses were performed. Values for p < 0.05 were considered statistically significant. RESULTS: The mean age of the participants was 60.39 ± 16.71 years. The mean IOP using the Tono-Pen was 21.1 ± 6.8 mm Hg versus the mean uncorrected GAT value of 17.1 ± 6.9 mm Hg and the corrected GAT value of 18.9 ± 7.5 mm Hg. The mean CCT was 510.5 ± 29.6 µm. The mean differences between the Tono-Pen reading and uncorrected and corrected GAT readings were 3.9 ± 2.6 and 2.1 ± 3.5 mm Hg, respectively. Gender (Tono-Pen vs. GAT; p = 0.981 vs. 0.437) and corneal thickness (p = 0.057) did not significantly affect the IOP value. Of the 75 patients, 68 (90.7%) preferred the Tono-Pen to the GAT. CONCLUSION: In this study, the Tono-Pen gave a higher value for IOP than the uncorrected and corrected GAT values. Gender and corneal thickness did not significantly affect the measurements. Most patients found the Tono-Pen more acceptable than the GAT. S. Karger AG 2016-11 2016-08-09 /pmc/articles/PMC5588504/ /pubmed/27504638 http://dx.doi.org/10.1159/000448953 Text en Copyright © 2016 by S. Karger AG, Basel http://creativecommons.org/licenses/by-nc/3.0/ This is an Open Access article licensed under the terms of the Creative Commons Attribution-NonCommercial 3.0 Unported license (CC BY-NC) (www.karger.com/OA-license), applicable to the online version of the article only. Distribution permitted for non-commercial purposes only.
spellingShingle Original Paper
Onochie, Chizoba
Okoye, Obiekwe
Ogunro, Adunola
Aribaba, Tade
Hassan, Kunle
Onakoya, Adeola
Comparisons of the Tono-Pen® and Goldmann Applanation Tonometer in the Measurement of Intraocular Pressure of Primary Open Angle Glaucoma Patients in a Hospital Population in Southwest Nigeria
title Comparisons of the Tono-Pen® and Goldmann Applanation Tonometer in the Measurement of Intraocular Pressure of Primary Open Angle Glaucoma Patients in a Hospital Population in Southwest Nigeria
title_full Comparisons of the Tono-Pen® and Goldmann Applanation Tonometer in the Measurement of Intraocular Pressure of Primary Open Angle Glaucoma Patients in a Hospital Population in Southwest Nigeria
title_fullStr Comparisons of the Tono-Pen® and Goldmann Applanation Tonometer in the Measurement of Intraocular Pressure of Primary Open Angle Glaucoma Patients in a Hospital Population in Southwest Nigeria
title_full_unstemmed Comparisons of the Tono-Pen® and Goldmann Applanation Tonometer in the Measurement of Intraocular Pressure of Primary Open Angle Glaucoma Patients in a Hospital Population in Southwest Nigeria
title_short Comparisons of the Tono-Pen® and Goldmann Applanation Tonometer in the Measurement of Intraocular Pressure of Primary Open Angle Glaucoma Patients in a Hospital Population in Southwest Nigeria
title_sort comparisons of the tono-pen® and goldmann applanation tonometer in the measurement of intraocular pressure of primary open angle glaucoma patients in a hospital population in southwest nigeria
topic Original Paper
url https://www.ncbi.nlm.nih.gov/pmc/articles/PMC5588504/
https://www.ncbi.nlm.nih.gov/pubmed/27504638
http://dx.doi.org/10.1159/000448953
work_keys_str_mv AT onochiechizoba comparisonsofthetonopenandgoldmannapplanationtonometerinthemeasurementofintraocularpressureofprimaryopenangleglaucomapatientsinahospitalpopulationinsouthwestnigeria
AT okoyeobiekwe comparisonsofthetonopenandgoldmannapplanationtonometerinthemeasurementofintraocularpressureofprimaryopenangleglaucomapatientsinahospitalpopulationinsouthwestnigeria
AT ogunroadunola comparisonsofthetonopenandgoldmannapplanationtonometerinthemeasurementofintraocularpressureofprimaryopenangleglaucomapatientsinahospitalpopulationinsouthwestnigeria
AT aribabatade comparisonsofthetonopenandgoldmannapplanationtonometerinthemeasurementofintraocularpressureofprimaryopenangleglaucomapatientsinahospitalpopulationinsouthwestnigeria
AT hassankunle comparisonsofthetonopenandgoldmannapplanationtonometerinthemeasurementofintraocularpressureofprimaryopenangleglaucomapatientsinahospitalpopulationinsouthwestnigeria
AT onakoyaadeola comparisonsofthetonopenandgoldmannapplanationtonometerinthemeasurementofintraocularpressureofprimaryopenangleglaucomapatientsinahospitalpopulationinsouthwestnigeria