Cargando…
Knowing About Device Algorithms to Understand the Rhythm and Role of Managed Ventricular Pacing
A 66-year-old patient with a DR pacemaker for intermittent atrioventricular block presented with an electrocardiogram (ECG) showing some P waves non followed by QRS complexes, raising suspicion of device dysfunction. The device was equipped with a special algorithm (Managed Ventricular Pacing; Medtr...
Autores principales: | Kossaify, Antoine, Hamdar, Hiba |
---|---|
Formato: | Online Artículo Texto |
Lenguaje: | English |
Publicado: |
SAGE Publications
2017
|
Materias: | |
Acceso en línea: | https://www.ncbi.nlm.nih.gov/pmc/articles/PMC5588790/ https://www.ncbi.nlm.nih.gov/pubmed/28904525 http://dx.doi.org/10.1177/1179547617729651 |
Ejemplares similares
-
Assessment of Ventricular Pacing in the Setting of an Institutional Improvement Program: Insights into Physiological Pacing
por: Kossaify, Antoine, et al.
Publicado: (2012) -
The Apparent and the Effective PR Interval, Insights for Cardiac Pacing
por: Kossaify, Antoine
Publicado: (2011) -
Electrocardiographic Abnormalities of Takotsubo Cardiomyopathy in a Patient with Paced Ventricular Rhythm
por: Chauhan, Krati, et al.
Publicado: (2010) -
Left ventricular pacing induced polymorphic ventricular tachycardia via the adaptive left ventricle pacing algorithm
por: Khalil, Fouad, et al.
Publicado: (2020) -
Automatic adjustment of ventricular antitachycardia pacing and individualized device therapy
por: Tripathi, Rajiv, et al.
Publicado: (2021)