Performance characteristics of loci method for measuring cardiac troponin I on the dimension EXL

OBJECTIVE: To define the performance characteristics of the LOCI(®) method for cardiac troponin I on the Dimension EXL system. DESIGNS AND METHODS: Three different levels of commercial control (mean concentrations 0.426, 1.42, and 18.64 µg/L) were used for the imprecision study, quantifying separate...

Descripción completa

Detalles Bibliográficos
Autores principales: Calderón, José L. Martín, Pérez, Julia M. Varona, Bustos, Fernando, Gómez, Juan C. Sánchez
Formato: Online Artículo Texto
Lenguaje:English
Publicado: Elsevier 2015
Materias:
Acceso en línea:https://www.ncbi.nlm.nih.gov/pmc/articles/PMC5597712/
https://www.ncbi.nlm.nih.gov/pubmed/28932798
http://dx.doi.org/10.1016/j.plabm.2015.03.006
_version_ 1783263759517614080
author Calderón, José L. Martín
Pérez, Julia M. Varona
Bustos, Fernando
Gómez, Juan C. Sánchez
author_facet Calderón, José L. Martín
Pérez, Julia M. Varona
Bustos, Fernando
Gómez, Juan C. Sánchez
author_sort Calderón, José L. Martín
collection PubMed
description OBJECTIVE: To define the performance characteristics of the LOCI(®) method for cardiac troponin I on the Dimension EXL system. DESIGNS AND METHODS: Three different levels of commercial control (mean concentrations 0.426, 1.42, and 18.64 µg/L) were used for the imprecision study, quantifying separately within-run and between-run over 20 days. The limit of blank (LoB) and limit of detection (LoD) were assessed with 20 replicates of a sample without troponin I. Linearity was assessed by regression analysis. In addition, we studied inaccuracy, carry-over and limit of quantitation and conducted a method comparison with the Stratus CS (n=69). The reference interval was determined in 146 healthy blood donors using non-parametric method. RESULTS: The within-run imprecision (coefficient of variation [CV], %) obtained at each level was 2.4, 1.4% and 2.2%, while the between-run imprecision (CV,%) was 3.3%, 2.9% and 2.5%. Total imprecision was 4.06%, 3.3% and 3.4% for each control level. The limit of quantitation which corresponds to the troponin I concentration at which CV=10% was 0.05 µg/L. Method comparison with the Stratus CS assay produced the equation: Dimension EXL=−0.002698+1.0233(⁎)(Stratus CS) with a confidence interval from −0.01562 to 0.00626 for the intercept and (0.979 to 1.0875) for the slope. The 99th percentile obtained for the reference population was 0.047 µg/L. CONCLUSIONS: The LOCI method for cardiac troponin I on the Dimension EXL meets all guidelines recommended criteria referring to limit of quantitation, imprecision and shows excellent transferability with the Stratus CS method.
format Online
Article
Text
id pubmed-5597712
institution National Center for Biotechnology Information
language English
publishDate 2015
publisher Elsevier
record_format MEDLINE/PubMed
spelling pubmed-55977122017-09-20 Performance characteristics of loci method for measuring cardiac troponin I on the dimension EXL Calderón, José L. Martín Pérez, Julia M. Varona Bustos, Fernando Gómez, Juan C. Sánchez Pract Lab Med Research Article OBJECTIVE: To define the performance characteristics of the LOCI(®) method for cardiac troponin I on the Dimension EXL system. DESIGNS AND METHODS: Three different levels of commercial control (mean concentrations 0.426, 1.42, and 18.64 µg/L) were used for the imprecision study, quantifying separately within-run and between-run over 20 days. The limit of blank (LoB) and limit of detection (LoD) were assessed with 20 replicates of a sample without troponin I. Linearity was assessed by regression analysis. In addition, we studied inaccuracy, carry-over and limit of quantitation and conducted a method comparison with the Stratus CS (n=69). The reference interval was determined in 146 healthy blood donors using non-parametric method. RESULTS: The within-run imprecision (coefficient of variation [CV], %) obtained at each level was 2.4, 1.4% and 2.2%, while the between-run imprecision (CV,%) was 3.3%, 2.9% and 2.5%. Total imprecision was 4.06%, 3.3% and 3.4% for each control level. The limit of quantitation which corresponds to the troponin I concentration at which CV=10% was 0.05 µg/L. Method comparison with the Stratus CS assay produced the equation: Dimension EXL=−0.002698+1.0233(⁎)(Stratus CS) with a confidence interval from −0.01562 to 0.00626 for the intercept and (0.979 to 1.0875) for the slope. The 99th percentile obtained for the reference population was 0.047 µg/L. CONCLUSIONS: The LOCI method for cardiac troponin I on the Dimension EXL meets all guidelines recommended criteria referring to limit of quantitation, imprecision and shows excellent transferability with the Stratus CS method. Elsevier 2015-04-08 /pmc/articles/PMC5597712/ /pubmed/28932798 http://dx.doi.org/10.1016/j.plabm.2015.03.006 Text en © 2015 Published by Elsevier B.V. http://creativecommons.org/licenses/by-nc-nd/4.0/ This is an open access article under the CC BY-NC-ND license (http://creativecommons.org/licenses/by-nc-nd/4.0/).
spellingShingle Research Article
Calderón, José L. Martín
Pérez, Julia M. Varona
Bustos, Fernando
Gómez, Juan C. Sánchez
Performance characteristics of loci method for measuring cardiac troponin I on the dimension EXL
title Performance characteristics of loci method for measuring cardiac troponin I on the dimension EXL
title_full Performance characteristics of loci method for measuring cardiac troponin I on the dimension EXL
title_fullStr Performance characteristics of loci method for measuring cardiac troponin I on the dimension EXL
title_full_unstemmed Performance characteristics of loci method for measuring cardiac troponin I on the dimension EXL
title_short Performance characteristics of loci method for measuring cardiac troponin I on the dimension EXL
title_sort performance characteristics of loci method for measuring cardiac troponin i on the dimension exl
topic Research Article
url https://www.ncbi.nlm.nih.gov/pmc/articles/PMC5597712/
https://www.ncbi.nlm.nih.gov/pubmed/28932798
http://dx.doi.org/10.1016/j.plabm.2015.03.006
work_keys_str_mv AT calderonjoselmartin performancecharacteristicsoflocimethodformeasuringcardiactroponinionthedimensionexl
AT perezjuliamvarona performancecharacteristicsoflocimethodformeasuringcardiactroponinionthedimensionexl
AT bustosfernando performancecharacteristicsoflocimethodformeasuringcardiactroponinionthedimensionexl
AT gomezjuancsanchez performancecharacteristicsoflocimethodformeasuringcardiactroponinionthedimensionexl