Effects of Imaging Parameters on the Quality of Contrast-Enhanced MR Angiography of Cerebral Aneurysms Treated Using Stent-Assisted Coiling: A Phantom Study
PURPOSE: To quantitatively investigate in vitro the effects of flip angle (FA), receiver bandwidth (BW), echo time (TE), and magnetic field strength (FS) on image noise and artifacts induced by stent-assisted coiling on contrast-enhanced MR angiography (CE-MRA) images, as a first step towards optimi...
Autores principales: | , , , |
---|---|
Formato: | Online Artículo Texto |
Lenguaje: | English |
Publicado: |
Japanese Society for Magnetic Resonance in Medicine
2016
|
Materias: | |
Acceso en línea: | https://www.ncbi.nlm.nih.gov/pmc/articles/PMC5600074/ https://www.ncbi.nlm.nih.gov/pubmed/27599584 http://dx.doi.org/10.2463/mrms.mp.2016-0042 |
_version_ | 1783264182338060288 |
---|---|
author | Ikushima, Yoichiro Hashido, Takashi Watanabe, Yoshiyuki Doi, Tsukasa |
author_facet | Ikushima, Yoichiro Hashido, Takashi Watanabe, Yoshiyuki Doi, Tsukasa |
author_sort | Ikushima, Yoichiro |
collection | PubMed |
description | PURPOSE: To quantitatively investigate in vitro the effects of flip angle (FA), receiver bandwidth (BW), echo time (TE), and magnetic field strength (FS) on image noise and artifacts induced by stent-assisted coiling on contrast-enhanced MR angiography (CE-MRA) images, as a first step towards optimization of imaging parameters. METHODS: A phantom simulating a cerebral aneurysm treated using stent-assisted coiling was filled with diluted gadolinium contrast medium, and MR angiography were obtained using varied parameters: FA (10°–60°), BW (164–780 Hz/pixel), and FS (1.5 and 3.0T). The TE varied automatically with BW because the TE was set to the smallest value. Three kinds of indices were semi-automatically calculated to quantify the severity of stent- and coil-induced artifacts: artificial lumen narrowing (ALN) representing a decrease in the in-stent luminal area, and relative in-stent signal (RIS(S)) and relative in-coil signal (RIS(C)) representing an increase in the in-stent and in-coil signal intensities, respectively. We also measured the ratio of in-stent signal to noise (IS/N) for each parameter. The variation in these indices with variations in FA, BW (TE), and FS was analyzed. RESULTS: An increase in FA led to an increase of up to 65% in the RIS(S), while the IS/N increased by up to three times. The 1.5T scanner indicated fewer artifacts (71% lower ALN, two times higher RIS(S), and 40% higher RIS(C)) than the 3.0T scanner. On the other hand, the 1.5T scanner worsened the IS/N compared with the 3.0T scanner, although the difference was relatively small. Variation in BW (and hence, TE) led to a trade-off between artifact severity and IS/N. CONCLUSION: A high FA and low FS should be used for improved artifact severity and IS/N on CE-MRA images of a stent-assisted coil. A wide BW (short TE) could improve artifact severity at the expense of the image noise. |
format | Online Article Text |
id | pubmed-5600074 |
institution | National Center for Biotechnology Information |
language | English |
publishDate | 2016 |
publisher | Japanese Society for Magnetic Resonance in Medicine |
record_format | MEDLINE/PubMed |
spelling | pubmed-56000742017-10-23 Effects of Imaging Parameters on the Quality of Contrast-Enhanced MR Angiography of Cerebral Aneurysms Treated Using Stent-Assisted Coiling: A Phantom Study Ikushima, Yoichiro Hashido, Takashi Watanabe, Yoshiyuki Doi, Tsukasa Magn Reson Med Sci Major Paper PURPOSE: To quantitatively investigate in vitro the effects of flip angle (FA), receiver bandwidth (BW), echo time (TE), and magnetic field strength (FS) on image noise and artifacts induced by stent-assisted coiling on contrast-enhanced MR angiography (CE-MRA) images, as a first step towards optimization of imaging parameters. METHODS: A phantom simulating a cerebral aneurysm treated using stent-assisted coiling was filled with diluted gadolinium contrast medium, and MR angiography were obtained using varied parameters: FA (10°–60°), BW (164–780 Hz/pixel), and FS (1.5 and 3.0T). The TE varied automatically with BW because the TE was set to the smallest value. Three kinds of indices were semi-automatically calculated to quantify the severity of stent- and coil-induced artifacts: artificial lumen narrowing (ALN) representing a decrease in the in-stent luminal area, and relative in-stent signal (RIS(S)) and relative in-coil signal (RIS(C)) representing an increase in the in-stent and in-coil signal intensities, respectively. We also measured the ratio of in-stent signal to noise (IS/N) for each parameter. The variation in these indices with variations in FA, BW (TE), and FS was analyzed. RESULTS: An increase in FA led to an increase of up to 65% in the RIS(S), while the IS/N increased by up to three times. The 1.5T scanner indicated fewer artifacts (71% lower ALN, two times higher RIS(S), and 40% higher RIS(C)) than the 3.0T scanner. On the other hand, the 1.5T scanner worsened the IS/N compared with the 3.0T scanner, although the difference was relatively small. Variation in BW (and hence, TE) led to a trade-off between artifact severity and IS/N. CONCLUSION: A high FA and low FS should be used for improved artifact severity and IS/N on CE-MRA images of a stent-assisted coil. A wide BW (short TE) could improve artifact severity at the expense of the image noise. Japanese Society for Magnetic Resonance in Medicine 2016-09-06 /pmc/articles/PMC5600074/ /pubmed/27599584 http://dx.doi.org/10.2463/mrms.mp.2016-0042 Text en © 2016 Japanese Society for Magnetic Resonance in Medicine http://creativecommons.org/licenses/by/3.0/ This work is licensed under a Creative Commons Attribution-NonCommercial-NoDerivatives International License. |
spellingShingle | Major Paper Ikushima, Yoichiro Hashido, Takashi Watanabe, Yoshiyuki Doi, Tsukasa Effects of Imaging Parameters on the Quality of Contrast-Enhanced MR Angiography of Cerebral Aneurysms Treated Using Stent-Assisted Coiling: A Phantom Study |
title | Effects of Imaging Parameters on the Quality of Contrast-Enhanced MR Angiography of Cerebral Aneurysms Treated Using Stent-Assisted Coiling: A Phantom Study |
title_full | Effects of Imaging Parameters on the Quality of Contrast-Enhanced MR Angiography of Cerebral Aneurysms Treated Using Stent-Assisted Coiling: A Phantom Study |
title_fullStr | Effects of Imaging Parameters on the Quality of Contrast-Enhanced MR Angiography of Cerebral Aneurysms Treated Using Stent-Assisted Coiling: A Phantom Study |
title_full_unstemmed | Effects of Imaging Parameters on the Quality of Contrast-Enhanced MR Angiography of Cerebral Aneurysms Treated Using Stent-Assisted Coiling: A Phantom Study |
title_short | Effects of Imaging Parameters on the Quality of Contrast-Enhanced MR Angiography of Cerebral Aneurysms Treated Using Stent-Assisted Coiling: A Phantom Study |
title_sort | effects of imaging parameters on the quality of contrast-enhanced mr angiography of cerebral aneurysms treated using stent-assisted coiling: a phantom study |
topic | Major Paper |
url | https://www.ncbi.nlm.nih.gov/pmc/articles/PMC5600074/ https://www.ncbi.nlm.nih.gov/pubmed/27599584 http://dx.doi.org/10.2463/mrms.mp.2016-0042 |
work_keys_str_mv | AT ikushimayoichiro effectsofimagingparametersonthequalityofcontrastenhancedmrangiographyofcerebralaneurysmstreatedusingstentassistedcoilingaphantomstudy AT hashidotakashi effectsofimagingparametersonthequalityofcontrastenhancedmrangiographyofcerebralaneurysmstreatedusingstentassistedcoilingaphantomstudy AT watanabeyoshiyuki effectsofimagingparametersonthequalityofcontrastenhancedmrangiographyofcerebralaneurysmstreatedusingstentassistedcoilingaphantomstudy AT doitsukasa effectsofimagingparametersonthequalityofcontrastenhancedmrangiographyofcerebralaneurysmstreatedusingstentassistedcoilingaphantomstudy |