Failure to Treat Life-Threatening Ventricular Tachyarrhythmias in Contemporary Implantable Cardioverter–Defibrillators: Implications for Strategic Programming
BACKGROUND: In clinical trials, manufacturer-specific, strategic programming of implantable cardioverter–defibrillators (ICDs), including faster detection rates, reduces unnecessary therapy but permits therapy for ventricular tachycardia/ventricular fibrillation (VF). Present consensus recommends a...
Autores principales: | , , , , |
---|---|
Formato: | Online Artículo Texto |
Lenguaje: | English |
Publicado: |
Lippincott Williams & Wilkins
2017
|
Materias: | |
Acceso en línea: | https://www.ncbi.nlm.nih.gov/pmc/articles/PMC5610568/ https://www.ncbi.nlm.nih.gov/pubmed/28916511 http://dx.doi.org/10.1161/CIRCEP.117.005305 |
Sumario: | BACKGROUND: In clinical trials, manufacturer-specific, strategic programming of implantable cardioverter–defibrillators (ICDs), including faster detection rates, reduces unnecessary therapy but permits therapy for ventricular tachycardia/ventricular fibrillation (VF). Present consensus recommends a generic rate threshold between 185 and 200 beats per minute, which exceeds the rate tested in clinical trials for some manufacturers. In a case series, we sought to determine the relationship between programmed parameters and failure of modern ICDs to treat VF. METHODS AND RESULTS: We reviewed cases in which normally functioning ICDs failed to deliver timely therapy for VF from April 2015 to January 2017 at 4 institutions. Of 10 ambulatory patients, 5 died from untreated VF, 4 had cardiac arrests requiring external shocks, and 1 was rescued by a delayed ICD shock. VF did not satisfy programmed detection criteria in 9 patients (90%). Seven of these patients had slowest detection rates that were consistent with generic recommendations but not tested in a peer-reviewed trial for their manufacturer’s ICDs. Manufacturer-specific factors interacted with fast detection rates to withhold therapy, including strict VF episode termination rules, enhancements to minimize T-wave oversensing, and features that restrict therapy to regular rhythms in ventricular tachycardia zones. Untreated VF despite recommended programming accounted for 56% of sudden deaths and 11% of all deaths during the study period. CONCLUSIONS: Complex and unanticipated interactions between manufacturer-specific features and generic programming can prevent therapy for VF. More data are needed to assess the risks and benefits of translating evidence-based detection parameters from one manufacturer to another. |
---|