Choice of a Suitable Dosimeter for Photon Percentage Depth Dose Measurements in Flattening Filter-Free Beams
The International Atomic Energy Agency Technical Reports Series-398 code of practice for dosimetry recommends measuring photon percentage depth dose (PDD) curves with parallel-plate chambers. This code of practice was published before flattening filter-free (FFF) beams were widely used in clinical l...
Autores principales: | , |
---|---|
Formato: | Online Artículo Texto |
Lenguaje: | English |
Publicado: |
Medknow Publications & Media Pvt Ltd
2017
|
Materias: | |
Acceso en línea: | https://www.ncbi.nlm.nih.gov/pmc/articles/PMC5618460/ https://www.ncbi.nlm.nih.gov/pubmed/28974859 http://dx.doi.org/10.4103/jmp.JMP_11_17 |
_version_ | 1783267191051780096 |
---|---|
author | Vargas Castrillón, Silvia Cutanda Henríquez, Francisco |
author_facet | Vargas Castrillón, Silvia Cutanda Henríquez, Francisco |
author_sort | Vargas Castrillón, Silvia |
collection | PubMed |
description | The International Atomic Energy Agency Technical Reports Series-398 code of practice for dosimetry recommends measuring photon percentage depth dose (PDD) curves with parallel-plate chambers. This code of practice was published before flattening filter-free (FFF) beams were widely used in clinical linear accelerators. The choice of detector for PDD measurements needs to be reassessed for FFF beams given the physical differences between FFF beams and flattened ones. The present study compares PDD curves for FFF beams of nominal energies 6 MV, 6 FFF, 10 MV, and 10 FFF from a Varian TrueBeam linear accelerator (Varian Medical Systems, Palo Alto, USA) acquired with Scanditronix photon diodes, two scanning type chambers (both PTW 31010 Semiflex), two small volume chambers (Wellhofer CC04 and PTW 31016 PinPoint 3D), PTW 34001 Roos, Scanditronix Roos, and NACP 02 parallel-plate chambers. Results show that parallel-plate ion chambers can be used for photon PDD measurements, although for better accuracy, recombination effects should be taken into account. |
format | Online Article Text |
id | pubmed-5618460 |
institution | National Center for Biotechnology Information |
language | English |
publishDate | 2017 |
publisher | Medknow Publications & Media Pvt Ltd |
record_format | MEDLINE/PubMed |
spelling | pubmed-56184602017-10-03 Choice of a Suitable Dosimeter for Photon Percentage Depth Dose Measurements in Flattening Filter-Free Beams Vargas Castrillón, Silvia Cutanda Henríquez, Francisco J Med Phys Technical Note The International Atomic Energy Agency Technical Reports Series-398 code of practice for dosimetry recommends measuring photon percentage depth dose (PDD) curves with parallel-plate chambers. This code of practice was published before flattening filter-free (FFF) beams were widely used in clinical linear accelerators. The choice of detector for PDD measurements needs to be reassessed for FFF beams given the physical differences between FFF beams and flattened ones. The present study compares PDD curves for FFF beams of nominal energies 6 MV, 6 FFF, 10 MV, and 10 FFF from a Varian TrueBeam linear accelerator (Varian Medical Systems, Palo Alto, USA) acquired with Scanditronix photon diodes, two scanning type chambers (both PTW 31010 Semiflex), two small volume chambers (Wellhofer CC04 and PTW 31016 PinPoint 3D), PTW 34001 Roos, Scanditronix Roos, and NACP 02 parallel-plate chambers. Results show that parallel-plate ion chambers can be used for photon PDD measurements, although for better accuracy, recombination effects should be taken into account. Medknow Publications & Media Pvt Ltd 2017 /pmc/articles/PMC5618460/ /pubmed/28974859 http://dx.doi.org/10.4103/jmp.JMP_11_17 Text en Copyright: © 2017 Journal of Medical Physics http://creativecommons.org/licenses/by-nc-sa/3.0 This is an open access article distributed under the terms of the Creative Commons Attribution-NonCommercial-ShareAlike 3.0 License, which allows others to remix, tweak, and build upon the work non-commercially, as long as the author is credited and the new creations are licensed under the identical terms. |
spellingShingle | Technical Note Vargas Castrillón, Silvia Cutanda Henríquez, Francisco Choice of a Suitable Dosimeter for Photon Percentage Depth Dose Measurements in Flattening Filter-Free Beams |
title | Choice of a Suitable Dosimeter for Photon Percentage Depth Dose Measurements in Flattening Filter-Free Beams |
title_full | Choice of a Suitable Dosimeter for Photon Percentage Depth Dose Measurements in Flattening Filter-Free Beams |
title_fullStr | Choice of a Suitable Dosimeter for Photon Percentage Depth Dose Measurements in Flattening Filter-Free Beams |
title_full_unstemmed | Choice of a Suitable Dosimeter for Photon Percentage Depth Dose Measurements in Flattening Filter-Free Beams |
title_short | Choice of a Suitable Dosimeter for Photon Percentage Depth Dose Measurements in Flattening Filter-Free Beams |
title_sort | choice of a suitable dosimeter for photon percentage depth dose measurements in flattening filter-free beams |
topic | Technical Note |
url | https://www.ncbi.nlm.nih.gov/pmc/articles/PMC5618460/ https://www.ncbi.nlm.nih.gov/pubmed/28974859 http://dx.doi.org/10.4103/jmp.JMP_11_17 |
work_keys_str_mv | AT vargascastrillonsilvia choiceofasuitabledosimeterforphotonpercentagedepthdosemeasurementsinflatteningfilterfreebeams AT cutandahenriquezfrancisco choiceofasuitabledosimeterforphotonpercentagedepthdosemeasurementsinflatteningfilterfreebeams |