The new Greiner FC-Mix tubes equal the old Terumo ones and are useful as glucose stabilizer after prolonged storage of samples

INTRODUCTION: The aim of our study is to compare new Greiner tubes containing granulated citrate buffer with the Terumo ones and to verify if they are suitable for glucose stabilization after prolonged storage. MATERIALS AND METHODS: In Study 1, blood was collected in two Terumo and two Greiner tube...

Descripción completa

Detalles Bibliográficos
Autores principales: Bonetti, Graziella, Carta, Mariarosa
Formato: Online Artículo Texto
Lenguaje:English
Publicado: Croatian Society of Medical Biochemistry and Laboratory Medicine 2017
Materias:
Acceso en línea:https://www.ncbi.nlm.nih.gov/pmc/articles/PMC5628000/
https://www.ncbi.nlm.nih.gov/pubmed/29018304
http://dx.doi.org/10.11613/BM.2017.030901
Descripción
Sumario:INTRODUCTION: The aim of our study is to compare new Greiner tubes containing granulated citrate buffer with the Terumo ones and to verify if they are suitable for glucose stabilization after prolonged storage. MATERIALS AND METHODS: In Study 1, blood was collected in two Terumo and two Greiner tubes from 40 healthy volunteers. Samples were stored at room temperature (RT) for 1 and 2 hours, respectively. Comparison was made by Deming regression. In Study 2, glucose was measured in a reference tube (N = 50), according to the ADA-NACB guidelines and in aliquots of Greiner samples maintained un-centrifuged at RT for 1, 2, 4 (N = 50) and 24, 48, 72 hours (N = 35). RESULTS: There were insignificant mixed biases between the Terumo and Greiner tubes. Compared to reference (5.3 mmol/L), glucose concentration in the new tubes was 5.4 (P < 0.05), 5.4 (P < 0.05), 5.3 (P = 0.265), 5.2 (P = 0.156), 5.3 (P < 0.05) and 5.2 (P < 0.05) mmol/L after 1, 2, 4, 24, 48 and 72 hours at RT, respectively. There was no biological difference between any of the time points up to 48 h (bias < ± 1.95%). CONCLUSIONS: The study shows that the new tubes perform equally well as the Terumo ones and ensure glucose stabilization up to 48 h as well as permit to create a link between the previous studies demonstrating the clinical utility of granulated citrate buffer and the future ones.