Effect of a Novel Intracycle Motion Correction Algorithm on Dual-Energy Spectral Coronary CT Angiography: A Study with Pulsating Coronary Artery Phantom at High Heart Rates
OBJECTIVE: Using a pulsating coronary artery phantom at high heart rate settings, we investigated the efficacy of a motion correction algorithm (MCA) to improve the image quality in dual-energy spectral coronary CT angiography (CCTA). MATERIALS AND METHODS: Coronary flow phantoms were scanned at hea...
Autores principales: | , , , , , , |
---|---|
Formato: | Online Artículo Texto |
Lenguaje: | English |
Publicado: |
The Korean Society of Radiology
2017
|
Materias: | |
Acceso en línea: | https://www.ncbi.nlm.nih.gov/pmc/articles/PMC5639153/ https://www.ncbi.nlm.nih.gov/pubmed/29089820 http://dx.doi.org/10.3348/kjr.2017.18.6.881 |
_version_ | 1783270837613232128 |
---|---|
author | Xing, Yan Zhao, Yuan Guo, Ning Pan, Cun-Xue Azati, Gulina Wang, Yan-Wei Liu, Wen-Ya |
author_facet | Xing, Yan Zhao, Yuan Guo, Ning Pan, Cun-Xue Azati, Gulina Wang, Yan-Wei Liu, Wen-Ya |
author_sort | Xing, Yan |
collection | PubMed |
description | OBJECTIVE: Using a pulsating coronary artery phantom at high heart rate settings, we investigated the efficacy of a motion correction algorithm (MCA) to improve the image quality in dual-energy spectral coronary CT angiography (CCTA). MATERIALS AND METHODS: Coronary flow phantoms were scanned at heart rates of 60–100 beats/min at 10-beats/min increments, using dual-energy spectral CT mode. Virtual monochromatic images were reconstructed from 50 to 90 keV at 10-keV increments. Two blinded observers assessed image quality using a 4-point Likert Scale (1 = non-diagnostic, 4 = excellent) and the fraction of interpretable segments using MCA versus conventional algorithm (CA). Comparison of variables was performed with the Wilcoxon rank sum test and McNemar test. RESULTS: At heart rates of 70, 80, 90, and 100 beats/min, images with MCA were rated as higher image scores compared to those with CA on monochromatic levels of 50, 60, and 70 keV (each p < 0.05). Meanwhile, at a heart rate of 90 beats/min, image interpretability was improved by MCA at a monochromatic level of 60 keV (p < 0.05) and 70 keV (p < 0.05). At a heart rate of 100 beats/min, image interpretability was improved by MCA at monochromatic levels of 50 keV (from 69.4% to 86.1%, p < 0.05), 60 keV (from 55.6% to 83.3%, p < 0.05) and 70 keV (from 33.3% to 69.3%, p < 0.05). CONCLUSION: Low-keV monochromatic images combined with MCA improves image quality and image interpretability in CCTAs at high heart rates. |
format | Online Article Text |
id | pubmed-5639153 |
institution | National Center for Biotechnology Information |
language | English |
publishDate | 2017 |
publisher | The Korean Society of Radiology |
record_format | MEDLINE/PubMed |
spelling | pubmed-56391532017-11-01 Effect of a Novel Intracycle Motion Correction Algorithm on Dual-Energy Spectral Coronary CT Angiography: A Study with Pulsating Coronary Artery Phantom at High Heart Rates Xing, Yan Zhao, Yuan Guo, Ning Pan, Cun-Xue Azati, Gulina Wang, Yan-Wei Liu, Wen-Ya Korean J Radiol Cardiovascular Imaging OBJECTIVE: Using a pulsating coronary artery phantom at high heart rate settings, we investigated the efficacy of a motion correction algorithm (MCA) to improve the image quality in dual-energy spectral coronary CT angiography (CCTA). MATERIALS AND METHODS: Coronary flow phantoms were scanned at heart rates of 60–100 beats/min at 10-beats/min increments, using dual-energy spectral CT mode. Virtual monochromatic images were reconstructed from 50 to 90 keV at 10-keV increments. Two blinded observers assessed image quality using a 4-point Likert Scale (1 = non-diagnostic, 4 = excellent) and the fraction of interpretable segments using MCA versus conventional algorithm (CA). Comparison of variables was performed with the Wilcoxon rank sum test and McNemar test. RESULTS: At heart rates of 70, 80, 90, and 100 beats/min, images with MCA were rated as higher image scores compared to those with CA on monochromatic levels of 50, 60, and 70 keV (each p < 0.05). Meanwhile, at a heart rate of 90 beats/min, image interpretability was improved by MCA at a monochromatic level of 60 keV (p < 0.05) and 70 keV (p < 0.05). At a heart rate of 100 beats/min, image interpretability was improved by MCA at monochromatic levels of 50 keV (from 69.4% to 86.1%, p < 0.05), 60 keV (from 55.6% to 83.3%, p < 0.05) and 70 keV (from 33.3% to 69.3%, p < 0.05). CONCLUSION: Low-keV monochromatic images combined with MCA improves image quality and image interpretability in CCTAs at high heart rates. The Korean Society of Radiology 2017 2017-09-21 /pmc/articles/PMC5639153/ /pubmed/29089820 http://dx.doi.org/10.3348/kjr.2017.18.6.881 Text en Copyright © 2017 The Korean Society of Radiology http://creativecommons.org/licenses/by-nc/4.0/ This is an Open Access article distributed under the terms of the Creative Commons Attribution Non-Commercial License (http://creativecommons.org/licenses/by-nc/4.0/) which permits unrestricted non-commercial use, distribution, and reproduction in any medium, provided the original work is properly cited. |
spellingShingle | Cardiovascular Imaging Xing, Yan Zhao, Yuan Guo, Ning Pan, Cun-Xue Azati, Gulina Wang, Yan-Wei Liu, Wen-Ya Effect of a Novel Intracycle Motion Correction Algorithm on Dual-Energy Spectral Coronary CT Angiography: A Study with Pulsating Coronary Artery Phantom at High Heart Rates |
title | Effect of a Novel Intracycle Motion Correction Algorithm on Dual-Energy Spectral Coronary CT Angiography: A Study with Pulsating Coronary Artery Phantom at High Heart Rates |
title_full | Effect of a Novel Intracycle Motion Correction Algorithm on Dual-Energy Spectral Coronary CT Angiography: A Study with Pulsating Coronary Artery Phantom at High Heart Rates |
title_fullStr | Effect of a Novel Intracycle Motion Correction Algorithm on Dual-Energy Spectral Coronary CT Angiography: A Study with Pulsating Coronary Artery Phantom at High Heart Rates |
title_full_unstemmed | Effect of a Novel Intracycle Motion Correction Algorithm on Dual-Energy Spectral Coronary CT Angiography: A Study with Pulsating Coronary Artery Phantom at High Heart Rates |
title_short | Effect of a Novel Intracycle Motion Correction Algorithm on Dual-Energy Spectral Coronary CT Angiography: A Study with Pulsating Coronary Artery Phantom at High Heart Rates |
title_sort | effect of a novel intracycle motion correction algorithm on dual-energy spectral coronary ct angiography: a study with pulsating coronary artery phantom at high heart rates |
topic | Cardiovascular Imaging |
url | https://www.ncbi.nlm.nih.gov/pmc/articles/PMC5639153/ https://www.ncbi.nlm.nih.gov/pubmed/29089820 http://dx.doi.org/10.3348/kjr.2017.18.6.881 |
work_keys_str_mv | AT xingyan effectofanovelintracyclemotioncorrectionalgorithmondualenergyspectralcoronaryctangiographyastudywithpulsatingcoronaryarteryphantomathighheartrates AT zhaoyuan effectofanovelintracyclemotioncorrectionalgorithmondualenergyspectralcoronaryctangiographyastudywithpulsatingcoronaryarteryphantomathighheartrates AT guoning effectofanovelintracyclemotioncorrectionalgorithmondualenergyspectralcoronaryctangiographyastudywithpulsatingcoronaryarteryphantomathighheartrates AT pancunxue effectofanovelintracyclemotioncorrectionalgorithmondualenergyspectralcoronaryctangiographyastudywithpulsatingcoronaryarteryphantomathighheartrates AT azatigulina effectofanovelintracyclemotioncorrectionalgorithmondualenergyspectralcoronaryctangiographyastudywithpulsatingcoronaryarteryphantomathighheartrates AT wangyanwei effectofanovelintracyclemotioncorrectionalgorithmondualenergyspectralcoronaryctangiographyastudywithpulsatingcoronaryarteryphantomathighheartrates AT liuwenya effectofanovelintracyclemotioncorrectionalgorithmondualenergyspectralcoronaryctangiographyastudywithpulsatingcoronaryarteryphantomathighheartrates |