Prospective Validation of Intra- and Interobserver Reproducibility of a New Point Shear Wave Elastographic Technique for Assessing Liver Stiffness in Patients with Chronic Liver Disease
OBJECTIVE: To assess intra- and inter-observer reproducibility of a new point shear wave elastography technique (pSWE, S-Shearwave, Samsung Medison) and compare its accuracy in assessing liver stiffness (LS) with an established pSWE technique (Virtual Touch Quantification, VTQ). MATERIALS AND METHOD...
Autores principales: | , , , , , , , , |
---|---|
Formato: | Online Artículo Texto |
Lenguaje: | English |
Publicado: |
The Korean Society of Radiology
2017
|
Materias: | |
Acceso en línea: | https://www.ncbi.nlm.nih.gov/pmc/articles/PMC5639158/ https://www.ncbi.nlm.nih.gov/pubmed/29089825 http://dx.doi.org/10.3348/kjr.2017.18.6.926 |
_version_ | 1783270838833774592 |
---|---|
author | Ahn, Su Joa Lee, Jeong Min Chang, Won Lee, Sang Min Kang, Hyo-Jin Yang, Hyunkyung Yoon, Jeong Hee Park, Sae Jin Han, Joon Koo |
author_facet | Ahn, Su Joa Lee, Jeong Min Chang, Won Lee, Sang Min Kang, Hyo-Jin Yang, Hyunkyung Yoon, Jeong Hee Park, Sae Jin Han, Joon Koo |
author_sort | Ahn, Su Joa |
collection | PubMed |
description | OBJECTIVE: To assess intra- and inter-observer reproducibility of a new point shear wave elastography technique (pSWE, S-Shearwave, Samsung Medison) and compare its accuracy in assessing liver stiffness (LS) with an established pSWE technique (Virtual Touch Quantification, VTQ). MATERIALS AND METHODS: Thirty-three patients were enrolled in this Institutional Review Board-approved prospective study. LS values were measured by VTQ on an Acuson S2000 system (Siemens Healthineer) and S-Shearwave on an RS-80A (Samsung Medison) in the same session, followed by two further S-Shearwave sessions for inter- and intra-observer variation at 8-hour intervals. The technical success rate (SR) and reliability of the measurements of both pSWE techniques were compared. The intra- and inter-observer reproducibility of S-Shearwave was determined by intraclass correlation coefficients (ICCs). LS values were measured by both methods of pSWE. The diagnostic performance in severe fibrosis (F ≥ 3) and cirrhosis (F = 4) was evaluated using the receiver operating characteristics curve analysis and the Obuchowski measure with the LS values of transient elastography as the referenced standard. RESULTS: The VTQ (100%, 33/33) and S-Shearwave (96.9%, 32/33) techniques did not display a significant difference in technical SR (p = 0.63) or reliability of LS measurements (96.9%, 32/33; 93.9%, 30/32, respectively, p = 0.61). The inter- and intra-observer agreement for LS measurements using the S-Shearwave technique was excellent (ICC = 0.98 and 0.99, respectively). The mean LS values of both pSWE techniques were not significantly different and exhibited a good correlation (r = 0.78). To detect F ≥ 3 and F = 4, VTQ and S-Shearwave showed comparable diagnostic accuracy as indicated by the following outcomes: areas under receiver operating characteristics curve (AUROC) = 0.87 (95% confidence intervals [CI] 0.70–0.96), 0.89 for VTQ (95% CI 0.74–0.97), respectively; and AUROC = 0.84 (95% CI 0.67–0.94), 0.94 (95% CI 0.80–0.99) for S-Shearwave (p > 0.48), respectively. The Obuchowski measures were similarly high for S-Shearwave and VTQ (0.94 vs. 0.95). CONCLUSION: S-Shearwave shows excellent inter- and intra-observer agreement and diagnostic effectiveness comparable to VTQ in detecting LS. |
format | Online Article Text |
id | pubmed-5639158 |
institution | National Center for Biotechnology Information |
language | English |
publishDate | 2017 |
publisher | The Korean Society of Radiology |
record_format | MEDLINE/PubMed |
spelling | pubmed-56391582017-11-01 Prospective Validation of Intra- and Interobserver Reproducibility of a New Point Shear Wave Elastographic Technique for Assessing Liver Stiffness in Patients with Chronic Liver Disease Ahn, Su Joa Lee, Jeong Min Chang, Won Lee, Sang Min Kang, Hyo-Jin Yang, Hyunkyung Yoon, Jeong Hee Park, Sae Jin Han, Joon Koo Korean J Radiol Gastrointestinal Imaging OBJECTIVE: To assess intra- and inter-observer reproducibility of a new point shear wave elastography technique (pSWE, S-Shearwave, Samsung Medison) and compare its accuracy in assessing liver stiffness (LS) with an established pSWE technique (Virtual Touch Quantification, VTQ). MATERIALS AND METHODS: Thirty-three patients were enrolled in this Institutional Review Board-approved prospective study. LS values were measured by VTQ on an Acuson S2000 system (Siemens Healthineer) and S-Shearwave on an RS-80A (Samsung Medison) in the same session, followed by two further S-Shearwave sessions for inter- and intra-observer variation at 8-hour intervals. The technical success rate (SR) and reliability of the measurements of both pSWE techniques were compared. The intra- and inter-observer reproducibility of S-Shearwave was determined by intraclass correlation coefficients (ICCs). LS values were measured by both methods of pSWE. The diagnostic performance in severe fibrosis (F ≥ 3) and cirrhosis (F = 4) was evaluated using the receiver operating characteristics curve analysis and the Obuchowski measure with the LS values of transient elastography as the referenced standard. RESULTS: The VTQ (100%, 33/33) and S-Shearwave (96.9%, 32/33) techniques did not display a significant difference in technical SR (p = 0.63) or reliability of LS measurements (96.9%, 32/33; 93.9%, 30/32, respectively, p = 0.61). The inter- and intra-observer agreement for LS measurements using the S-Shearwave technique was excellent (ICC = 0.98 and 0.99, respectively). The mean LS values of both pSWE techniques were not significantly different and exhibited a good correlation (r = 0.78). To detect F ≥ 3 and F = 4, VTQ and S-Shearwave showed comparable diagnostic accuracy as indicated by the following outcomes: areas under receiver operating characteristics curve (AUROC) = 0.87 (95% confidence intervals [CI] 0.70–0.96), 0.89 for VTQ (95% CI 0.74–0.97), respectively; and AUROC = 0.84 (95% CI 0.67–0.94), 0.94 (95% CI 0.80–0.99) for S-Shearwave (p > 0.48), respectively. The Obuchowski measures were similarly high for S-Shearwave and VTQ (0.94 vs. 0.95). CONCLUSION: S-Shearwave shows excellent inter- and intra-observer agreement and diagnostic effectiveness comparable to VTQ in detecting LS. The Korean Society of Radiology 2017 2017-09-21 /pmc/articles/PMC5639158/ /pubmed/29089825 http://dx.doi.org/10.3348/kjr.2017.18.6.926 Text en Copyright © 2017 The Korean Society of Radiology http://creativecommons.org/licenses/by-nc/4.0/ This is an Open Access article distributed under the terms of the Creative Commons Attribution Non-Commercial License (http://creativecommons.org/licenses/by-nc/4.0/) which permits unrestricted non-commercial use, distribution, and reproduction in any medium, provided the original work is properly cited. |
spellingShingle | Gastrointestinal Imaging Ahn, Su Joa Lee, Jeong Min Chang, Won Lee, Sang Min Kang, Hyo-Jin Yang, Hyunkyung Yoon, Jeong Hee Park, Sae Jin Han, Joon Koo Prospective Validation of Intra- and Interobserver Reproducibility of a New Point Shear Wave Elastographic Technique for Assessing Liver Stiffness in Patients with Chronic Liver Disease |
title | Prospective Validation of Intra- and Interobserver Reproducibility of a New Point Shear Wave Elastographic Technique for Assessing Liver Stiffness in Patients with Chronic Liver Disease |
title_full | Prospective Validation of Intra- and Interobserver Reproducibility of a New Point Shear Wave Elastographic Technique for Assessing Liver Stiffness in Patients with Chronic Liver Disease |
title_fullStr | Prospective Validation of Intra- and Interobserver Reproducibility of a New Point Shear Wave Elastographic Technique for Assessing Liver Stiffness in Patients with Chronic Liver Disease |
title_full_unstemmed | Prospective Validation of Intra- and Interobserver Reproducibility of a New Point Shear Wave Elastographic Technique for Assessing Liver Stiffness in Patients with Chronic Liver Disease |
title_short | Prospective Validation of Intra- and Interobserver Reproducibility of a New Point Shear Wave Elastographic Technique for Assessing Liver Stiffness in Patients with Chronic Liver Disease |
title_sort | prospective validation of intra- and interobserver reproducibility of a new point shear wave elastographic technique for assessing liver stiffness in patients with chronic liver disease |
topic | Gastrointestinal Imaging |
url | https://www.ncbi.nlm.nih.gov/pmc/articles/PMC5639158/ https://www.ncbi.nlm.nih.gov/pubmed/29089825 http://dx.doi.org/10.3348/kjr.2017.18.6.926 |
work_keys_str_mv | AT ahnsujoa prospectivevalidationofintraandinterobserverreproducibilityofanewpointshearwaveelastographictechniqueforassessingliverstiffnessinpatientswithchronicliverdisease AT leejeongmin prospectivevalidationofintraandinterobserverreproducibilityofanewpointshearwaveelastographictechniqueforassessingliverstiffnessinpatientswithchronicliverdisease AT changwon prospectivevalidationofintraandinterobserverreproducibilityofanewpointshearwaveelastographictechniqueforassessingliverstiffnessinpatientswithchronicliverdisease AT leesangmin prospectivevalidationofintraandinterobserverreproducibilityofanewpointshearwaveelastographictechniqueforassessingliverstiffnessinpatientswithchronicliverdisease AT kanghyojin prospectivevalidationofintraandinterobserverreproducibilityofanewpointshearwaveelastographictechniqueforassessingliverstiffnessinpatientswithchronicliverdisease AT yanghyunkyung prospectivevalidationofintraandinterobserverreproducibilityofanewpointshearwaveelastographictechniqueforassessingliverstiffnessinpatientswithchronicliverdisease AT yoonjeonghee prospectivevalidationofintraandinterobserverreproducibilityofanewpointshearwaveelastographictechniqueforassessingliverstiffnessinpatientswithchronicliverdisease AT parksaejin prospectivevalidationofintraandinterobserverreproducibilityofanewpointshearwaveelastographictechniqueforassessingliverstiffnessinpatientswithchronicliverdisease AT hanjoonkoo prospectivevalidationofintraandinterobserverreproducibilityofanewpointshearwaveelastographictechniqueforassessingliverstiffnessinpatientswithchronicliverdisease |