Fitbit Charge HR Wireless Heart Rate Monitor: Validation Study Conducted Under Free-Living Conditions

BACKGROUND: Many modern smart watches and activity trackers feature an optical sensor that estimates the wearer’s heart rate. Recent studies have evaluated the performance of these consumer devices in the laboratory. OBJECTIVE: The objective of our study was to examine the accuracy and sensitivity o...

Descripción completa

Detalles Bibliográficos
Autores principales: Gorny, Alexander Wilhelm, Liew, Seaw Jia, Tan, Chuen Seng, Müller-Riemenschneider, Falk
Formato: Online Artículo Texto
Lenguaje:English
Publicado: JMIR Publications 2017
Materias:
Acceso en línea:https://www.ncbi.nlm.nih.gov/pmc/articles/PMC5670315/
https://www.ncbi.nlm.nih.gov/pubmed/29055881
http://dx.doi.org/10.2196/mhealth.8233
_version_ 1783275999206572032
author Gorny, Alexander Wilhelm
Liew, Seaw Jia
Tan, Chuen Seng
Müller-Riemenschneider, Falk
author_facet Gorny, Alexander Wilhelm
Liew, Seaw Jia
Tan, Chuen Seng
Müller-Riemenschneider, Falk
author_sort Gorny, Alexander Wilhelm
collection PubMed
description BACKGROUND: Many modern smart watches and activity trackers feature an optical sensor that estimates the wearer’s heart rate. Recent studies have evaluated the performance of these consumer devices in the laboratory. OBJECTIVE: The objective of our study was to examine the accuracy and sensitivity of a common wrist-worn tracker device in measuring heart rates and detecting 1-min bouts of moderate to vigorous physical activity (MVPA) under free-living conditions. METHODS: Ten healthy volunteers were recruited from a large university in Singapore to participate in a limited field test, followed by a month of continuous data collection. During the field test, each participant would wear one Fitbit Charge HR activity tracker and one Polar H6 heart rate monitor. Fitbit measures were accessed at 1-min intervals, while Polar readings were available for 10-s intervals. We derived intraclass correlation coefficients (ICCs) for individual participants comparing heart rate estimates. We applied Centers for Disease Control and Prevention heart rate zone cut-offs to ascertain the sensitivity and specificity of Fitbit in identifying 1-min epochs falling into MVPA heart rate zone. RESULTS: We collected paired heart rate data for 2509 1-min epochs in 10 individuals under free-living conditions of 3 to 6 hours. The overall ICC comparing 1-min Fitbit measures with average 10-s Polar H6 measures for the same epoch was .83 (95% CI .63-.91). On average, the Fitbit tracker underestimated heart rate measures by −5.96 bpm (standard error, SE=0.18). At the low intensity heart rate zone, the underestimate was smaller at −4.22 bpm (SE=0.15). This underestimate grew to −16.2 bpm (SE=0.74) in the MVPA heart rate zone. Fitbit devices detected 52.9% (192/363) of MVPA heart rate zone epochs correctly. Positive and negative predictive values were 86.1% (192/223) and 92.52% (2115/2286), respectively. During subsequent 1 month of continuous data collection (270 person-days), only 3.9% of 1-min epochs could be categorized as MVPA according to heart rate zones. This measure was affected by decreasing wear time and adherence over the period of follow-up. CONCLUSIONS: Under free-living conditions, Fitbit trackers are affected by significant systematic errors. Improvements in tracker accuracy and sensitivity when measuring MVPA are required before they can be considered for use in the context of exercise prescription to promote better health.
format Online
Article
Text
id pubmed-5670315
institution National Center for Biotechnology Information
language English
publishDate 2017
publisher JMIR Publications
record_format MEDLINE/PubMed
spelling pubmed-56703152017-11-14 Fitbit Charge HR Wireless Heart Rate Monitor: Validation Study Conducted Under Free-Living Conditions Gorny, Alexander Wilhelm Liew, Seaw Jia Tan, Chuen Seng Müller-Riemenschneider, Falk JMIR Mhealth Uhealth Original Paper BACKGROUND: Many modern smart watches and activity trackers feature an optical sensor that estimates the wearer’s heart rate. Recent studies have evaluated the performance of these consumer devices in the laboratory. OBJECTIVE: The objective of our study was to examine the accuracy and sensitivity of a common wrist-worn tracker device in measuring heart rates and detecting 1-min bouts of moderate to vigorous physical activity (MVPA) under free-living conditions. METHODS: Ten healthy volunteers were recruited from a large university in Singapore to participate in a limited field test, followed by a month of continuous data collection. During the field test, each participant would wear one Fitbit Charge HR activity tracker and one Polar H6 heart rate monitor. Fitbit measures were accessed at 1-min intervals, while Polar readings were available for 10-s intervals. We derived intraclass correlation coefficients (ICCs) for individual participants comparing heart rate estimates. We applied Centers for Disease Control and Prevention heart rate zone cut-offs to ascertain the sensitivity and specificity of Fitbit in identifying 1-min epochs falling into MVPA heart rate zone. RESULTS: We collected paired heart rate data for 2509 1-min epochs in 10 individuals under free-living conditions of 3 to 6 hours. The overall ICC comparing 1-min Fitbit measures with average 10-s Polar H6 measures for the same epoch was .83 (95% CI .63-.91). On average, the Fitbit tracker underestimated heart rate measures by −5.96 bpm (standard error, SE=0.18). At the low intensity heart rate zone, the underestimate was smaller at −4.22 bpm (SE=0.15). This underestimate grew to −16.2 bpm (SE=0.74) in the MVPA heart rate zone. Fitbit devices detected 52.9% (192/363) of MVPA heart rate zone epochs correctly. Positive and negative predictive values were 86.1% (192/223) and 92.52% (2115/2286), respectively. During subsequent 1 month of continuous data collection (270 person-days), only 3.9% of 1-min epochs could be categorized as MVPA according to heart rate zones. This measure was affected by decreasing wear time and adherence over the period of follow-up. CONCLUSIONS: Under free-living conditions, Fitbit trackers are affected by significant systematic errors. Improvements in tracker accuracy and sensitivity when measuring MVPA are required before they can be considered for use in the context of exercise prescription to promote better health. JMIR Publications 2017-10-20 /pmc/articles/PMC5670315/ /pubmed/29055881 http://dx.doi.org/10.2196/mhealth.8233 Text en ©Alexander Wilhelm Gorny, Seaw Jia Liew, Chuen Seng Tan, Falk Müller-Riemenschneider. Originally published in JMIR Mhealth and Uhealth (http://mhealth.jmir.org), 20.10.2017. https://creativecommons.org/licenses/by/4.0/ This is an open-access article distributed under the terms of the Creative Commons Attribution License (https://creativecommons.org/licenses/by/4.0/), which permits unrestricted use, distribution, and reproduction in any medium, provided the original work, first published in JMIR mhealth and uhealth, is properly cited. The complete bibliographic information, a link to the original publication on http://mhealth.jmir.org/, as well as this copyright and license information must be included.
spellingShingle Original Paper
Gorny, Alexander Wilhelm
Liew, Seaw Jia
Tan, Chuen Seng
Müller-Riemenschneider, Falk
Fitbit Charge HR Wireless Heart Rate Monitor: Validation Study Conducted Under Free-Living Conditions
title Fitbit Charge HR Wireless Heart Rate Monitor: Validation Study Conducted Under Free-Living Conditions
title_full Fitbit Charge HR Wireless Heart Rate Monitor: Validation Study Conducted Under Free-Living Conditions
title_fullStr Fitbit Charge HR Wireless Heart Rate Monitor: Validation Study Conducted Under Free-Living Conditions
title_full_unstemmed Fitbit Charge HR Wireless Heart Rate Monitor: Validation Study Conducted Under Free-Living Conditions
title_short Fitbit Charge HR Wireless Heart Rate Monitor: Validation Study Conducted Under Free-Living Conditions
title_sort fitbit charge hr wireless heart rate monitor: validation study conducted under free-living conditions
topic Original Paper
url https://www.ncbi.nlm.nih.gov/pmc/articles/PMC5670315/
https://www.ncbi.nlm.nih.gov/pubmed/29055881
http://dx.doi.org/10.2196/mhealth.8233
work_keys_str_mv AT gornyalexanderwilhelm fitbitchargehrwirelessheartratemonitorvalidationstudyconductedunderfreelivingconditions
AT liewseawjia fitbitchargehrwirelessheartratemonitorvalidationstudyconductedunderfreelivingconditions
AT tanchuenseng fitbitchargehrwirelessheartratemonitorvalidationstudyconductedunderfreelivingconditions
AT mullerriemenschneiderfalk fitbitchargehrwirelessheartratemonitorvalidationstudyconductedunderfreelivingconditions