Bore Disruption: An Unusual Mechanical Failure of Two Hip Hemiarthroplasties
We report a case of two hemiarthroplasties (HAs) that underwent an unexpected mechanical failure. Two patients affected by femoral fractures were treated with an HA. At 5 and 7 years after surgery, they showed a breakage of the HA at the head–stem junction. We analyzed macroscopically and microscopi...
Autores principales: | , , , , |
---|---|
Formato: | Online Artículo Texto |
Lenguaje: | English |
Publicado: |
Georg Thieme Verlag KG
2017
|
Acceso en línea: | https://www.ncbi.nlm.nih.gov/pmc/articles/PMC5672855/ https://www.ncbi.nlm.nih.gov/pubmed/29114631 http://dx.doi.org/10.1055/s-0037-1601407 |
_version_ | 1783276513440825344 |
---|---|
author | Dettoni, Federico Cottino, Umberto Spriano, Silvia Bonasia, Davide Edoardo Rossi, Roberto |
author_facet | Dettoni, Federico Cottino, Umberto Spriano, Silvia Bonasia, Davide Edoardo Rossi, Roberto |
author_sort | Dettoni, Federico |
collection | PubMed |
description | We report a case of two hemiarthroplasties (HAs) that underwent an unexpected mechanical failure. Two patients affected by femoral fractures were treated with an HA. At 5 and 7 years after surgery, they showed a breakage of the HA at the head–stem junction. We analyzed macroscopically and microscopically (by a scanning electron microscopy) the surfaces of the broken prostheses. In both cases, a fracture was detected starting at the joining point between the femoral head and the slot that receives the taper cone of the stem (the so-called bore), with the typical macroscopic and microscopic pattern of a fatigue fracture. The fatigue fracture resulted from repeated cycles of torsional stresses throughout the years. In literature, we did not find any other report of such HA failure. |
format | Online Article Text |
id | pubmed-5672855 |
institution | National Center for Biotechnology Information |
language | English |
publishDate | 2017 |
publisher | Georg Thieme Verlag KG |
record_format | MEDLINE/PubMed |
spelling | pubmed-56728552017-11-07 Bore Disruption: An Unusual Mechanical Failure of Two Hip Hemiarthroplasties Dettoni, Federico Cottino, Umberto Spriano, Silvia Bonasia, Davide Edoardo Rossi, Roberto Joints We report a case of two hemiarthroplasties (HAs) that underwent an unexpected mechanical failure. Two patients affected by femoral fractures were treated with an HA. At 5 and 7 years after surgery, they showed a breakage of the HA at the head–stem junction. We analyzed macroscopically and microscopically (by a scanning electron microscopy) the surfaces of the broken prostheses. In both cases, a fracture was detected starting at the joining point between the femoral head and the slot that receives the taper cone of the stem (the so-called bore), with the typical macroscopic and microscopic pattern of a fatigue fracture. The fatigue fracture resulted from repeated cycles of torsional stresses throughout the years. In literature, we did not find any other report of such HA failure. Georg Thieme Verlag KG 2017-06-05 /pmc/articles/PMC5672855/ /pubmed/29114631 http://dx.doi.org/10.1055/s-0037-1601407 Text en https://creativecommons.org/licenses/by-nc-nd/4.0/ This is an open-access article distributed under the terms of the Creative Commons Attribution-NonCommercial-NoDerivatives License, which permits unrestricted reproduction and distribution, for non-commercial purposes only; and use and reproduction, but not distribution, of adapted material for non-commercial purposes only, provided the original work is properly cited. |
spellingShingle | Dettoni, Federico Cottino, Umberto Spriano, Silvia Bonasia, Davide Edoardo Rossi, Roberto Bore Disruption: An Unusual Mechanical Failure of Two Hip Hemiarthroplasties |
title | Bore Disruption: An Unusual Mechanical Failure of Two Hip Hemiarthroplasties |
title_full | Bore Disruption: An Unusual Mechanical Failure of Two Hip Hemiarthroplasties |
title_fullStr | Bore Disruption: An Unusual Mechanical Failure of Two Hip Hemiarthroplasties |
title_full_unstemmed | Bore Disruption: An Unusual Mechanical Failure of Two Hip Hemiarthroplasties |
title_short | Bore Disruption: An Unusual Mechanical Failure of Two Hip Hemiarthroplasties |
title_sort | bore disruption: an unusual mechanical failure of two hip hemiarthroplasties |
url | https://www.ncbi.nlm.nih.gov/pmc/articles/PMC5672855/ https://www.ncbi.nlm.nih.gov/pubmed/29114631 http://dx.doi.org/10.1055/s-0037-1601407 |
work_keys_str_mv | AT dettonifederico boredisruptionanunusualmechanicalfailureoftwohiphemiarthroplasties AT cottinoumberto boredisruptionanunusualmechanicalfailureoftwohiphemiarthroplasties AT sprianosilvia boredisruptionanunusualmechanicalfailureoftwohiphemiarthroplasties AT bonasiadavideedoardo boredisruptionanunusualmechanicalfailureoftwohiphemiarthroplasties AT rossiroberto boredisruptionanunusualmechanicalfailureoftwohiphemiarthroplasties |