Comparison of Methods Study between a Photonic Crystal Biosensor and Certified ELISA to Measure Biomarkers of Iron Deficiency in Chronic Kidney Disease Patients

The total analytical error of a photonic crystal (PC) biosensor in the determination of ferritin and soluble transferrin receptor (sTfR) as biomarkers of iron deficiency anemia in chronic kidney disease (CKD) patients was evaluated against certified ELISAs. Antigens were extracted from sera of CKD p...

Descripción completa

Detalles Bibliográficos
Autores principales: Peterson, Ross D., Wilund, Kenneth R., Cunningham, Brian T., Andrade, Juan E.
Formato: Online Artículo Texto
Lenguaje:English
Publicado: MDPI 2017
Materias:
Acceso en línea:https://www.ncbi.nlm.nih.gov/pmc/articles/PMC5677296/
https://www.ncbi.nlm.nih.gov/pubmed/28946680
http://dx.doi.org/10.3390/s17102203
_version_ 1783277212138471424
author Peterson, Ross D.
Wilund, Kenneth R.
Cunningham, Brian T.
Andrade, Juan E.
author_facet Peterson, Ross D.
Wilund, Kenneth R.
Cunningham, Brian T.
Andrade, Juan E.
author_sort Peterson, Ross D.
collection PubMed
description The total analytical error of a photonic crystal (PC) biosensor in the determination of ferritin and soluble transferrin receptor (sTfR) as biomarkers of iron deficiency anemia in chronic kidney disease (CKD) patients was evaluated against certified ELISAs. Antigens were extracted from sera of CKD patients using functionalized iron-oxide nanoparticles (fAb-IONs) followed by magnetic separation. Immuno-complexes were recognized by complementary detection Ab affixed to the PC biosensor surface, and their signals were followed using the BIND instrument. Quantification was conducted against actual protein standards. Total calculated error (TEcalc) was estimated based on systematic (SE) and random error (RE) and compared against total allowed error (TEa) based on established quality specifications. Both detection platforms showed adequate linearity, specificity, and sensitivity for biomarkers. Means, SD, and CV were similar between biomarkers for both detection platforms. Compared to ELISA, inherent imprecision was higher on the PC biosensor for ferritin, but not for sTfR. High SE or RE in the PC biosensor when measuring either biomarker resulted in TEcalc higher than the TEa. This did not influence the diagnostic ability of the PC biosensor to discriminate CKD patients with low iron stores. The performance of the PC biosensor is similar to certified ELISAs; however, optimization is required to reduce TEcalc.
format Online
Article
Text
id pubmed-5677296
institution National Center for Biotechnology Information
language English
publishDate 2017
publisher MDPI
record_format MEDLINE/PubMed
spelling pubmed-56772962017-11-17 Comparison of Methods Study between a Photonic Crystal Biosensor and Certified ELISA to Measure Biomarkers of Iron Deficiency in Chronic Kidney Disease Patients Peterson, Ross D. Wilund, Kenneth R. Cunningham, Brian T. Andrade, Juan E. Sensors (Basel) Article The total analytical error of a photonic crystal (PC) biosensor in the determination of ferritin and soluble transferrin receptor (sTfR) as biomarkers of iron deficiency anemia in chronic kidney disease (CKD) patients was evaluated against certified ELISAs. Antigens were extracted from sera of CKD patients using functionalized iron-oxide nanoparticles (fAb-IONs) followed by magnetic separation. Immuno-complexes were recognized by complementary detection Ab affixed to the PC biosensor surface, and their signals were followed using the BIND instrument. Quantification was conducted against actual protein standards. Total calculated error (TEcalc) was estimated based on systematic (SE) and random error (RE) and compared against total allowed error (TEa) based on established quality specifications. Both detection platforms showed adequate linearity, specificity, and sensitivity for biomarkers. Means, SD, and CV were similar between biomarkers for both detection platforms. Compared to ELISA, inherent imprecision was higher on the PC biosensor for ferritin, but not for sTfR. High SE or RE in the PC biosensor when measuring either biomarker resulted in TEcalc higher than the TEa. This did not influence the diagnostic ability of the PC biosensor to discriminate CKD patients with low iron stores. The performance of the PC biosensor is similar to certified ELISAs; however, optimization is required to reduce TEcalc. MDPI 2017-09-25 /pmc/articles/PMC5677296/ /pubmed/28946680 http://dx.doi.org/10.3390/s17102203 Text en © 2017 by the authors. Licensee MDPI, Basel, Switzerland. This article is an open access article distributed under the terms and conditions of the Creative Commons Attribution (CC BY) license (http://creativecommons.org/licenses/by/4.0/).
spellingShingle Article
Peterson, Ross D.
Wilund, Kenneth R.
Cunningham, Brian T.
Andrade, Juan E.
Comparison of Methods Study between a Photonic Crystal Biosensor and Certified ELISA to Measure Biomarkers of Iron Deficiency in Chronic Kidney Disease Patients
title Comparison of Methods Study between a Photonic Crystal Biosensor and Certified ELISA to Measure Biomarkers of Iron Deficiency in Chronic Kidney Disease Patients
title_full Comparison of Methods Study between a Photonic Crystal Biosensor and Certified ELISA to Measure Biomarkers of Iron Deficiency in Chronic Kidney Disease Patients
title_fullStr Comparison of Methods Study between a Photonic Crystal Biosensor and Certified ELISA to Measure Biomarkers of Iron Deficiency in Chronic Kidney Disease Patients
title_full_unstemmed Comparison of Methods Study between a Photonic Crystal Biosensor and Certified ELISA to Measure Biomarkers of Iron Deficiency in Chronic Kidney Disease Patients
title_short Comparison of Methods Study between a Photonic Crystal Biosensor and Certified ELISA to Measure Biomarkers of Iron Deficiency in Chronic Kidney Disease Patients
title_sort comparison of methods study between a photonic crystal biosensor and certified elisa to measure biomarkers of iron deficiency in chronic kidney disease patients
topic Article
url https://www.ncbi.nlm.nih.gov/pmc/articles/PMC5677296/
https://www.ncbi.nlm.nih.gov/pubmed/28946680
http://dx.doi.org/10.3390/s17102203
work_keys_str_mv AT petersonrossd comparisonofmethodsstudybetweenaphotoniccrystalbiosensorandcertifiedelisatomeasurebiomarkersofirondeficiencyinchronickidneydiseasepatients
AT wilundkennethr comparisonofmethodsstudybetweenaphotoniccrystalbiosensorandcertifiedelisatomeasurebiomarkersofirondeficiencyinchronickidneydiseasepatients
AT cunninghambriant comparisonofmethodsstudybetweenaphotoniccrystalbiosensorandcertifiedelisatomeasurebiomarkersofirondeficiencyinchronickidneydiseasepatients
AT andradejuane comparisonofmethodsstudybetweenaphotoniccrystalbiosensorandcertifiedelisatomeasurebiomarkersofirondeficiencyinchronickidneydiseasepatients