Cargando…
Effect of beam profile measurement on arc therapy plan quality assurance: a case study
We present an example when profile measurement and modeling of an Elekta Agility multileaf collimator (MLC) had a large effect specifically on arc therapy plan quality assurance (QA) results using ArcCheck. ArcCheck absolute dose measurements of these plans were systematically lower than planned by...
Autores principales: | , , , , |
---|---|
Formato: | Online Artículo Texto |
Lenguaje: | English |
Publicado: |
John Wiley and Sons Inc.
2017
|
Materias: | |
Acceso en línea: | https://www.ncbi.nlm.nih.gov/pmc/articles/PMC5689840/ https://www.ncbi.nlm.nih.gov/pubmed/28370918 http://dx.doi.org/10.1002/acm2.12071 |
_version_ | 1783279468747423744 |
---|---|
author | Kim, Leonard H. Malajovich, Irina Reyhan, Meral L. Xue, Jinyu Park, Joo Han |
author_facet | Kim, Leonard H. Malajovich, Irina Reyhan, Meral L. Xue, Jinyu Park, Joo Han |
author_sort | Kim, Leonard H. |
collection | PubMed |
description | We present an example when profile measurement and modeling of an Elekta Agility multileaf collimator (MLC) had a large effect specifically on arc therapy plan quality assurance (QA) results using ArcCheck. ArcCheck absolute dose measurements of these plans were systematically lower than planned by 3–10%. Failing QA results were seen even with unmodulated static and conformal arcs. Furthermore, the effect was found to be dependent on collimator angle, with worse results associated with near‐zero collimator angles. In contrast, step‐and‐shoot QA results were not affected. Changing the beam model to match steeper profile measurements obtained using a different measurement device resolved the problem. This case study demonstrates that conventional gamma index analysis can be sensitive to small profile modeling changes. |
format | Online Article Text |
id | pubmed-5689840 |
institution | National Center for Biotechnology Information |
language | English |
publishDate | 2017 |
publisher | John Wiley and Sons Inc. |
record_format | MEDLINE/PubMed |
spelling | pubmed-56898402018-04-02 Effect of beam profile measurement on arc therapy plan quality assurance: a case study Kim, Leonard H. Malajovich, Irina Reyhan, Meral L. Xue, Jinyu Park, Joo Han J Appl Clin Med Phys Radiation Oncology Physics We present an example when profile measurement and modeling of an Elekta Agility multileaf collimator (MLC) had a large effect specifically on arc therapy plan quality assurance (QA) results using ArcCheck. ArcCheck absolute dose measurements of these plans were systematically lower than planned by 3–10%. Failing QA results were seen even with unmodulated static and conformal arcs. Furthermore, the effect was found to be dependent on collimator angle, with worse results associated with near‐zero collimator angles. In contrast, step‐and‐shoot QA results were not affected. Changing the beam model to match steeper profile measurements obtained using a different measurement device resolved the problem. This case study demonstrates that conventional gamma index analysis can be sensitive to small profile modeling changes. John Wiley and Sons Inc. 2017-03-29 /pmc/articles/PMC5689840/ /pubmed/28370918 http://dx.doi.org/10.1002/acm2.12071 Text en © 2017 The Authors. Journal of Applied Clinical Medical Physics published by Wiley Periodicals, Inc. on behalf of American Association of Physicists in Medicine. This is an open access article under the terms of the Creative Commons Attribution (http://creativecommons.org/licenses/by/4.0/) License, which permits use, distribution and reproduction in any medium, provided the original work is properly cited. |
spellingShingle | Radiation Oncology Physics Kim, Leonard H. Malajovich, Irina Reyhan, Meral L. Xue, Jinyu Park, Joo Han Effect of beam profile measurement on arc therapy plan quality assurance: a case study |
title | Effect of beam profile measurement on arc therapy plan quality assurance: a case study |
title_full | Effect of beam profile measurement on arc therapy plan quality assurance: a case study |
title_fullStr | Effect of beam profile measurement on arc therapy plan quality assurance: a case study |
title_full_unstemmed | Effect of beam profile measurement on arc therapy plan quality assurance: a case study |
title_short | Effect of beam profile measurement on arc therapy plan quality assurance: a case study |
title_sort | effect of beam profile measurement on arc therapy plan quality assurance: a case study |
topic | Radiation Oncology Physics |
url | https://www.ncbi.nlm.nih.gov/pmc/articles/PMC5689840/ https://www.ncbi.nlm.nih.gov/pubmed/28370918 http://dx.doi.org/10.1002/acm2.12071 |
work_keys_str_mv | AT kimleonardh effectofbeamprofilemeasurementonarctherapyplanqualityassuranceacasestudy AT malajovichirina effectofbeamprofilemeasurementonarctherapyplanqualityassuranceacasestudy AT reyhanmerall effectofbeamprofilemeasurementonarctherapyplanqualityassuranceacasestudy AT xuejinyu effectofbeamprofilemeasurementonarctherapyplanqualityassuranceacasestudy AT parkjoohan effectofbeamprofilemeasurementonarctherapyplanqualityassuranceacasestudy |