Measurement of skin surface dose distributions in radiation therapy using poly(vinyl alcohol) cryogel dosimeters

In external beam radiation therapy (EBRT), skin dose measurement is important to evaluate dose coverage of superficial target volumes. Treatment planning systems (TPSs) are often inaccurate in this region of the patient, so in vivo measurements are necessary for skin surface dose estimation. In this...

Descripción completa

Detalles Bibliográficos
Autores principales: Eyadeh, Molham M., Wierzbicki, Marcin, Diamond, Kevin R.
Formato: Online Artículo Texto
Lenguaje:English
Publicado: John Wiley and Sons Inc. 2017
Materias:
Acceso en línea:https://www.ncbi.nlm.nih.gov/pmc/articles/PMC5689842/
https://www.ncbi.nlm.nih.gov/pubmed/28436134
http://dx.doi.org/10.1002/acm2.12087
_version_ 1783279469236060160
author Eyadeh, Molham M.
Wierzbicki, Marcin
Diamond, Kevin R.
author_facet Eyadeh, Molham M.
Wierzbicki, Marcin
Diamond, Kevin R.
author_sort Eyadeh, Molham M.
collection PubMed
description In external beam radiation therapy (EBRT), skin dose measurement is important to evaluate dose coverage of superficial target volumes. Treatment planning systems (TPSs) are often inaccurate in this region of the patient, so in vivo measurements are necessary for skin surface dose estimation. In this work, superficial dose distributions were measured using radiochromic translucent poly(vinyl alcohol) cryogels. The cryogels simultaneously served as bolus material, providing the necessary buildup to achieve the desired superficial dose. The relationship between dose to the skin surface and dose measured with the bolus was established using a series of oblique irradiations with gantry angles ranging from 0° to 90°. EBT‐2 Gafchromic film was placed under the bolus, and the ratio of bolus‐film dose was determined ranging from 0.749 ± 0.005 to 0.930 ± 0.002 for 0° and 90° gantry angles, respectively. The average ratio over 0–67.5° (0.800 ± 0.064) was used as the single correction factor to convert dose in bolus to dose to the skin surface. The correction factor was applied to bolus measurements of skin dose from head and neck intensity‐modulated radiation therapy (IMRT) treatments delivered to a RANDO phantom. The resulting dose distributions were compared to film measurements using gamma analysis with a 3%/3 mm tolerance and a 10% threshold. The minimum gamma pass rate was 95.2% suggesting that the radiochromic bolus may provide an accurate estimation of skin surface dose using a simple correction factor. This study demonstrates the suitability of radiochromic cryogels for superficial dose measurements in megavoltage photon beams.
format Online
Article
Text
id pubmed-5689842
institution National Center for Biotechnology Information
language English
publishDate 2017
publisher John Wiley and Sons Inc.
record_format MEDLINE/PubMed
spelling pubmed-56898422018-04-02 Measurement of skin surface dose distributions in radiation therapy using poly(vinyl alcohol) cryogel dosimeters Eyadeh, Molham M. Wierzbicki, Marcin Diamond, Kevin R. J Appl Clin Med Phys Radiation Oncology Physics In external beam radiation therapy (EBRT), skin dose measurement is important to evaluate dose coverage of superficial target volumes. Treatment planning systems (TPSs) are often inaccurate in this region of the patient, so in vivo measurements are necessary for skin surface dose estimation. In this work, superficial dose distributions were measured using radiochromic translucent poly(vinyl alcohol) cryogels. The cryogels simultaneously served as bolus material, providing the necessary buildup to achieve the desired superficial dose. The relationship between dose to the skin surface and dose measured with the bolus was established using a series of oblique irradiations with gantry angles ranging from 0° to 90°. EBT‐2 Gafchromic film was placed under the bolus, and the ratio of bolus‐film dose was determined ranging from 0.749 ± 0.005 to 0.930 ± 0.002 for 0° and 90° gantry angles, respectively. The average ratio over 0–67.5° (0.800 ± 0.064) was used as the single correction factor to convert dose in bolus to dose to the skin surface. The correction factor was applied to bolus measurements of skin dose from head and neck intensity‐modulated radiation therapy (IMRT) treatments delivered to a RANDO phantom. The resulting dose distributions were compared to film measurements using gamma analysis with a 3%/3 mm tolerance and a 10% threshold. The minimum gamma pass rate was 95.2% suggesting that the radiochromic bolus may provide an accurate estimation of skin surface dose using a simple correction factor. This study demonstrates the suitability of radiochromic cryogels for superficial dose measurements in megavoltage photon beams. John Wiley and Sons Inc. 2017-04-24 /pmc/articles/PMC5689842/ /pubmed/28436134 http://dx.doi.org/10.1002/acm2.12087 Text en © 2017 The Authors. Journal of Applied Clinical Medical Physics published by Wiley Periodicals, Inc. on behalf of American Association of Physicists in Medicine. This is an open access article under the terms of the Creative Commons Attribution (http://creativecommons.org/licenses/by/4.0/) License, which permits use, distribution and reproduction in any medium, provided the original work is properly cited.
spellingShingle Radiation Oncology Physics
Eyadeh, Molham M.
Wierzbicki, Marcin
Diamond, Kevin R.
Measurement of skin surface dose distributions in radiation therapy using poly(vinyl alcohol) cryogel dosimeters
title Measurement of skin surface dose distributions in radiation therapy using poly(vinyl alcohol) cryogel dosimeters
title_full Measurement of skin surface dose distributions in radiation therapy using poly(vinyl alcohol) cryogel dosimeters
title_fullStr Measurement of skin surface dose distributions in radiation therapy using poly(vinyl alcohol) cryogel dosimeters
title_full_unstemmed Measurement of skin surface dose distributions in radiation therapy using poly(vinyl alcohol) cryogel dosimeters
title_short Measurement of skin surface dose distributions in radiation therapy using poly(vinyl alcohol) cryogel dosimeters
title_sort measurement of skin surface dose distributions in radiation therapy using poly(vinyl alcohol) cryogel dosimeters
topic Radiation Oncology Physics
url https://www.ncbi.nlm.nih.gov/pmc/articles/PMC5689842/
https://www.ncbi.nlm.nih.gov/pubmed/28436134
http://dx.doi.org/10.1002/acm2.12087
work_keys_str_mv AT eyadehmolhamm measurementofskinsurfacedosedistributionsinradiationtherapyusingpolyvinylalcoholcryogeldosimeters
AT wierzbickimarcin measurementofskinsurfacedosedistributionsinradiationtherapyusingpolyvinylalcoholcryogeldosimeters
AT diamondkevinr measurementofskinsurfacedosedistributionsinradiationtherapyusingpolyvinylalcoholcryogeldosimeters