Single‐fraction spine SBRT end‐to‐end testing on TomoTherapy, Vero, TrueBeam, and CyberKnife treatment platforms using a novel anthropomorphic phantom

Spine SBRT involves the delivery of very high doses of radiation to targets adjacent to the spinal cord and is most commonly delivered in a single fraction. Highly conformal planning and accurate delivery of such plans is imperative for successful treatment without catastrophic adverse effects. End–...

Descripción completa

Detalles Bibliográficos
Autores principales: Gallo, John J., Kaufman, Isaac, Powell, Rachel, Pandya, Shalini, Somnay, Archana, Bossenberger, Todd, Ramirez, Ezequiel, Reynolds, Robert, Solberg, Timothy, Burmeister, Jay
Formato: Online Artículo Texto
Lenguaje:English
Publicado: John Wiley and Sons Inc. 2015
Materias:
Acceso en línea:https://www.ncbi.nlm.nih.gov/pmc/articles/PMC5689980/
https://www.ncbi.nlm.nih.gov/pubmed/25679169
http://dx.doi.org/10.1120/jacmp.v16i1.5120
_version_ 1783279501714653184
author Gallo, John J.
Kaufman, Isaac
Powell, Rachel
Pandya, Shalini
Somnay, Archana
Bossenberger, Todd
Ramirez, Ezequiel
Reynolds, Robert
Solberg, Timothy
Burmeister, Jay
author_facet Gallo, John J.
Kaufman, Isaac
Powell, Rachel
Pandya, Shalini
Somnay, Archana
Bossenberger, Todd
Ramirez, Ezequiel
Reynolds, Robert
Solberg, Timothy
Burmeister, Jay
author_sort Gallo, John J.
collection PubMed
description Spine SBRT involves the delivery of very high doses of radiation to targets adjacent to the spinal cord and is most commonly delivered in a single fraction. Highly conformal planning and accurate delivery of such plans is imperative for successful treatment without catastrophic adverse effects. End–to‐end testing is an important practice for evaluating the entire treatment process from simulation through treatment delivery. We performed end‐to‐end testing for a set of representative spine targets planned and delivered using four different treatment planning systems (TPSs) and delivery systems to evaluate the various capabilities of each. An anthropomorphic E2E SBRT phantom was simulated and treated on each system to evaluate agreement between measured and calculated doses. The phantom accepts ion chambers in the thoracic region and radiochromic film in the lumbar region. Four representative targets were developed within each region (thoracic and lumbar) to represent different presentations of spinal metastases and planned according to RTOG 0631 constraints. Plans were created using the TomoTherapy TPS for delivery using the Hi·Art system, the iPlan TPS for delivery using the Vero system, the Eclipse TPS for delivery using the TrueBeam system in both flattened and flattening filter free (FFF), and the MultiPlan TPS for delivery using the CyberKnife system. Delivered doses were measured using a 0.007 cm(3) ion chamber in the thoracic region and EBT3 GAFCHROMIC film in the lumbar region. Films were scanned and analyzed using an Epson Expression 10000XL flatbed scanner in conjunction with FilmQAPro2013. All treatment platforms met all dose constraints required by RTOG 0631. Ion chamber measurements in the thoracic targets delivered an overall average difference of 1.5%. Specifically, measurements agreed with the TPS to within 2.2%, 3.2%, 1.4%, 3.1%, and 3.0% for all three measureable cases on TomoTherapy, Vero, TrueBeam (FFF), TrueBeam (flattened), and CyberKnife, respectively. Film measurements for the lumbar targets resulted in average global gamma index passing rates of 100% at 3%/3 mm, 96.9% at 2%/2 mm, and 61.8% at 1%/1 mm, with a 10% minimum threshold for all plans on all platforms. Local gamma analysis was also performed with similar results. While gamma passing rates were consistently accurate across all platforms through 2%/2 mm, treatment beam‐on delivery times varied greatly between each platform with TrueBeam FFF being shortest, averaging 4.4 min, TrueBeam using flattened beam at 9.5 min, TomoTherapy at 30.5 min, Vero at 19 min, and CyberKnife at 46.0 min. In spite of the complexity of the representative targets and their proximity to the spinal cord, all treatment platforms were able to create plans meeting all RTOG 0631 dose constraints and produced exceptional agreement between calculated and measured doses. However, there were differences in the plan characteristics and significant differences in the beam‐on delivery time between platforms. Thus, clinical judgment is required for each particular case to determine most appropriate treatment planning/delivery platform. PACS number: 87.53.Ly
format Online
Article
Text
id pubmed-5689980
institution National Center for Biotechnology Information
language English
publishDate 2015
publisher John Wiley and Sons Inc.
record_format MEDLINE/PubMed
spelling pubmed-56899802018-04-02 Single‐fraction spine SBRT end‐to‐end testing on TomoTherapy, Vero, TrueBeam, and CyberKnife treatment platforms using a novel anthropomorphic phantom Gallo, John J. Kaufman, Isaac Powell, Rachel Pandya, Shalini Somnay, Archana Bossenberger, Todd Ramirez, Ezequiel Reynolds, Robert Solberg, Timothy Burmeister, Jay J Appl Clin Med Phys Radiation Oncology Physics Spine SBRT involves the delivery of very high doses of radiation to targets adjacent to the spinal cord and is most commonly delivered in a single fraction. Highly conformal planning and accurate delivery of such plans is imperative for successful treatment without catastrophic adverse effects. End–to‐end testing is an important practice for evaluating the entire treatment process from simulation through treatment delivery. We performed end‐to‐end testing for a set of representative spine targets planned and delivered using four different treatment planning systems (TPSs) and delivery systems to evaluate the various capabilities of each. An anthropomorphic E2E SBRT phantom was simulated and treated on each system to evaluate agreement between measured and calculated doses. The phantom accepts ion chambers in the thoracic region and radiochromic film in the lumbar region. Four representative targets were developed within each region (thoracic and lumbar) to represent different presentations of spinal metastases and planned according to RTOG 0631 constraints. Plans were created using the TomoTherapy TPS for delivery using the Hi·Art system, the iPlan TPS for delivery using the Vero system, the Eclipse TPS for delivery using the TrueBeam system in both flattened and flattening filter free (FFF), and the MultiPlan TPS for delivery using the CyberKnife system. Delivered doses were measured using a 0.007 cm(3) ion chamber in the thoracic region and EBT3 GAFCHROMIC film in the lumbar region. Films were scanned and analyzed using an Epson Expression 10000XL flatbed scanner in conjunction with FilmQAPro2013. All treatment platforms met all dose constraints required by RTOG 0631. Ion chamber measurements in the thoracic targets delivered an overall average difference of 1.5%. Specifically, measurements agreed with the TPS to within 2.2%, 3.2%, 1.4%, 3.1%, and 3.0% for all three measureable cases on TomoTherapy, Vero, TrueBeam (FFF), TrueBeam (flattened), and CyberKnife, respectively. Film measurements for the lumbar targets resulted in average global gamma index passing rates of 100% at 3%/3 mm, 96.9% at 2%/2 mm, and 61.8% at 1%/1 mm, with a 10% minimum threshold for all plans on all platforms. Local gamma analysis was also performed with similar results. While gamma passing rates were consistently accurate across all platforms through 2%/2 mm, treatment beam‐on delivery times varied greatly between each platform with TrueBeam FFF being shortest, averaging 4.4 min, TrueBeam using flattened beam at 9.5 min, TomoTherapy at 30.5 min, Vero at 19 min, and CyberKnife at 46.0 min. In spite of the complexity of the representative targets and their proximity to the spinal cord, all treatment platforms were able to create plans meeting all RTOG 0631 dose constraints and produced exceptional agreement between calculated and measured doses. However, there were differences in the plan characteristics and significant differences in the beam‐on delivery time between platforms. Thus, clinical judgment is required for each particular case to determine most appropriate treatment planning/delivery platform. PACS number: 87.53.Ly John Wiley and Sons Inc. 2015-01-08 /pmc/articles/PMC5689980/ /pubmed/25679169 http://dx.doi.org/10.1120/jacmp.v16i1.5120 Text en © 2015 The Authors. This is an open access article under the terms of the Creative Commons Attribution (http://creativecommons.org/licenses/by/3.0/) License, which permits use, distribution and reproduction in any medium, provided the original work is properly cited.
spellingShingle Radiation Oncology Physics
Gallo, John J.
Kaufman, Isaac
Powell, Rachel
Pandya, Shalini
Somnay, Archana
Bossenberger, Todd
Ramirez, Ezequiel
Reynolds, Robert
Solberg, Timothy
Burmeister, Jay
Single‐fraction spine SBRT end‐to‐end testing on TomoTherapy, Vero, TrueBeam, and CyberKnife treatment platforms using a novel anthropomorphic phantom
title Single‐fraction spine SBRT end‐to‐end testing on TomoTherapy, Vero, TrueBeam, and CyberKnife treatment platforms using a novel anthropomorphic phantom
title_full Single‐fraction spine SBRT end‐to‐end testing on TomoTherapy, Vero, TrueBeam, and CyberKnife treatment platforms using a novel anthropomorphic phantom
title_fullStr Single‐fraction spine SBRT end‐to‐end testing on TomoTherapy, Vero, TrueBeam, and CyberKnife treatment platforms using a novel anthropomorphic phantom
title_full_unstemmed Single‐fraction spine SBRT end‐to‐end testing on TomoTherapy, Vero, TrueBeam, and CyberKnife treatment platforms using a novel anthropomorphic phantom
title_short Single‐fraction spine SBRT end‐to‐end testing on TomoTherapy, Vero, TrueBeam, and CyberKnife treatment platforms using a novel anthropomorphic phantom
title_sort single‐fraction spine sbrt end‐to‐end testing on tomotherapy, vero, truebeam, and cyberknife treatment platforms using a novel anthropomorphic phantom
topic Radiation Oncology Physics
url https://www.ncbi.nlm.nih.gov/pmc/articles/PMC5689980/
https://www.ncbi.nlm.nih.gov/pubmed/25679169
http://dx.doi.org/10.1120/jacmp.v16i1.5120
work_keys_str_mv AT gallojohnj singlefractionspinesbrtendtoendtestingontomotherapyverotruebeamandcyberknifetreatmentplatformsusinganovelanthropomorphicphantom
AT kaufmanisaac singlefractionspinesbrtendtoendtestingontomotherapyverotruebeamandcyberknifetreatmentplatformsusinganovelanthropomorphicphantom
AT powellrachel singlefractionspinesbrtendtoendtestingontomotherapyverotruebeamandcyberknifetreatmentplatformsusinganovelanthropomorphicphantom
AT pandyashalini singlefractionspinesbrtendtoendtestingontomotherapyverotruebeamandcyberknifetreatmentplatformsusinganovelanthropomorphicphantom
AT somnayarchana singlefractionspinesbrtendtoendtestingontomotherapyverotruebeamandcyberknifetreatmentplatformsusinganovelanthropomorphicphantom
AT bossenbergertodd singlefractionspinesbrtendtoendtestingontomotherapyverotruebeamandcyberknifetreatmentplatformsusinganovelanthropomorphicphantom
AT ramirezezequiel singlefractionspinesbrtendtoendtestingontomotherapyverotruebeamandcyberknifetreatmentplatformsusinganovelanthropomorphicphantom
AT reynoldsrobert singlefractionspinesbrtendtoendtestingontomotherapyverotruebeamandcyberknifetreatmentplatformsusinganovelanthropomorphicphantom
AT solbergtimothy singlefractionspinesbrtendtoendtestingontomotherapyverotruebeamandcyberknifetreatmentplatformsusinganovelanthropomorphicphantom
AT burmeisterjay singlefractionspinesbrtendtoendtestingontomotherapyverotruebeamandcyberknifetreatmentplatformsusinganovelanthropomorphicphantom