Cargando…

A virtual source model for Monte Carlo simulation of helical tomotherapy

The purpose of this study was to present a Monte Carlo (MC) simulation method based on a virtual source, jaw, and MLC model to calculate dose in patient for helical tomotherapy without the need of calculating phase‐space files (PSFs). Current studies on the tomotherapy MC simulation adopt a full MC...

Descripción completa

Detalles Bibliográficos
Autores principales: Yuan, Jiankui, Rong, Yi, Chen, Quan
Formato: Online Artículo Texto
Lenguaje:English
Publicado: John Wiley and Sons Inc. 2015
Materias:
Acceso en línea:https://www.ncbi.nlm.nih.gov/pmc/articles/PMC5689983/
https://www.ncbi.nlm.nih.gov/pubmed/25679157
http://dx.doi.org/10.1120/jacmp.v16i1.4992
_version_ 1783279502507376640
author Yuan, Jiankui
Rong, Yi
Chen, Quan
author_facet Yuan, Jiankui
Rong, Yi
Chen, Quan
author_sort Yuan, Jiankui
collection PubMed
description The purpose of this study was to present a Monte Carlo (MC) simulation method based on a virtual source, jaw, and MLC model to calculate dose in patient for helical tomotherapy without the need of calculating phase‐space files (PSFs). Current studies on the tomotherapy MC simulation adopt a full MC model, which includes extensive modeling of radiation source, primary and secondary jaws, and multileaf collimator (MLC). In the full MC model, PSFs need to be created at different scoring planes to facilitate the patient dose calculations. In the present work, the virtual source model (VSM) we established was based on the gold standard beam data of a tomotherapy unit, which can be exported from the treatment planning station (TPS). The TPS‐generated sinograms were extracted from the archived patient XML (eXtensible Markup Language) files. The fluence map for the MC sampling was created by incorporating the percentage leaf open time (LOT) with leaf filter, jaw penumbra, and leaf latency contained from sinogram files. The VSM was validated for various geometry setups and clinical situations involving heterogeneous media and delivery quality assurance (DQA) cases. An agreement of [Formula: see text] was obtained between the measured and simulated results for percent depth doses (PDDs) and open beam profiles for all three jaw settings in the VSM commissioning. The accuracy of the VSM leaf filter model was verified in comparing the measured and simulated results for a Picket Fence pattern. An agreement of [Formula: see text] was achieved between the presented VSM and a published full MC model for heterogeneous phantoms. For complex clinical head and neck (HN) cases, the VSM‐based MC simulation of DQA plans agreed with the film measurement with 98% of planar dose pixels passing on the 2%/2 mm gamma criteria. For patient treatment plans, results showed comparable dose‐volume histograms (DVHs) for planning target volumes (PTVs) and organs at risk (OARs). Deviations observed in this study were consistent with literature. The VSM‐based MC simulation approach can be feasibly built from the gold standard beam model of a tomotherapy unit. The accuracy of the VSM was validated against measurements in homogeneous media, as well as published full MC model in heterogeneous media. PACS numbers: 87.53.‐j, 87.55.K‐
format Online
Article
Text
id pubmed-5689983
institution National Center for Biotechnology Information
language English
publishDate 2015
publisher John Wiley and Sons Inc.
record_format MEDLINE/PubMed
spelling pubmed-56899832018-04-02 A virtual source model for Monte Carlo simulation of helical tomotherapy Yuan, Jiankui Rong, Yi Chen, Quan J Appl Clin Med Phys Radiation Oncology Physics The purpose of this study was to present a Monte Carlo (MC) simulation method based on a virtual source, jaw, and MLC model to calculate dose in patient for helical tomotherapy without the need of calculating phase‐space files (PSFs). Current studies on the tomotherapy MC simulation adopt a full MC model, which includes extensive modeling of radiation source, primary and secondary jaws, and multileaf collimator (MLC). In the full MC model, PSFs need to be created at different scoring planes to facilitate the patient dose calculations. In the present work, the virtual source model (VSM) we established was based on the gold standard beam data of a tomotherapy unit, which can be exported from the treatment planning station (TPS). The TPS‐generated sinograms were extracted from the archived patient XML (eXtensible Markup Language) files. The fluence map for the MC sampling was created by incorporating the percentage leaf open time (LOT) with leaf filter, jaw penumbra, and leaf latency contained from sinogram files. The VSM was validated for various geometry setups and clinical situations involving heterogeneous media and delivery quality assurance (DQA) cases. An agreement of [Formula: see text] was obtained between the measured and simulated results for percent depth doses (PDDs) and open beam profiles for all three jaw settings in the VSM commissioning. The accuracy of the VSM leaf filter model was verified in comparing the measured and simulated results for a Picket Fence pattern. An agreement of [Formula: see text] was achieved between the presented VSM and a published full MC model for heterogeneous phantoms. For complex clinical head and neck (HN) cases, the VSM‐based MC simulation of DQA plans agreed with the film measurement with 98% of planar dose pixels passing on the 2%/2 mm gamma criteria. For patient treatment plans, results showed comparable dose‐volume histograms (DVHs) for planning target volumes (PTVs) and organs at risk (OARs). Deviations observed in this study were consistent with literature. The VSM‐based MC simulation approach can be feasibly built from the gold standard beam model of a tomotherapy unit. The accuracy of the VSM was validated against measurements in homogeneous media, as well as published full MC model in heterogeneous media. PACS numbers: 87.53.‐j, 87.55.K‐ John Wiley and Sons Inc. 2015-01-08 /pmc/articles/PMC5689983/ /pubmed/25679157 http://dx.doi.org/10.1120/jacmp.v16i1.4992 Text en © 2015 The Authors. This is an open access article under the terms of the Creative Commons Attribution (http://creativecommons.org/licenses/by/3.0/) License, which permits use, distribution and reproduction in any medium, provided the original work is properly cited.
spellingShingle Radiation Oncology Physics
Yuan, Jiankui
Rong, Yi
Chen, Quan
A virtual source model for Monte Carlo simulation of helical tomotherapy
title A virtual source model for Monte Carlo simulation of helical tomotherapy
title_full A virtual source model for Monte Carlo simulation of helical tomotherapy
title_fullStr A virtual source model for Monte Carlo simulation of helical tomotherapy
title_full_unstemmed A virtual source model for Monte Carlo simulation of helical tomotherapy
title_short A virtual source model for Monte Carlo simulation of helical tomotherapy
title_sort virtual source model for monte carlo simulation of helical tomotherapy
topic Radiation Oncology Physics
url https://www.ncbi.nlm.nih.gov/pmc/articles/PMC5689983/
https://www.ncbi.nlm.nih.gov/pubmed/25679157
http://dx.doi.org/10.1120/jacmp.v16i1.4992
work_keys_str_mv AT yuanjiankui avirtualsourcemodelformontecarlosimulationofhelicaltomotherapy
AT rongyi avirtualsourcemodelformontecarlosimulationofhelicaltomotherapy
AT chenquan avirtualsourcemodelformontecarlosimulationofhelicaltomotherapy
AT yuanjiankui virtualsourcemodelformontecarlosimulationofhelicaltomotherapy
AT rongyi virtualsourcemodelformontecarlosimulationofhelicaltomotherapy
AT chenquan virtualsourcemodelformontecarlosimulationofhelicaltomotherapy