Cargando…
Quality control of CT systems by automated monitoring of key performance indicators: a two‐year study
The purpose of this study was to develop a method of performing routine periodical quality controls (QC) of CT systems by automatically analyzing key performance indicators (KPIs), obtainable from images of manufacturers' quality assurance (QA) phantoms. A KPI pertains to a measurable or determ...
Autores principales: | Nowik, Patrik, Bujila, Robert, Poludniowski, Gavin, Fransson, Annette |
---|---|
Formato: | Online Artículo Texto |
Lenguaje: | English |
Publicado: |
John Wiley and Sons Inc.
2015
|
Materias: | |
Acceso en línea: | https://www.ncbi.nlm.nih.gov/pmc/articles/PMC5690007/ https://www.ncbi.nlm.nih.gov/pubmed/26219012 http://dx.doi.org/10.1120/jacmp.v16i4.5469 |
Ejemplares similares
-
Applying three different methods of measuring CTDI
(free air) to the extended CTDI formalism for wide‐beam scanners (IEC 60601–2–44): A comparative study
por: Bujila, Robert, et al.
Publicado: (2018) -
Quantitation of clinical feedback on image quality differences between two CT scanner models
por: Bache, Steven T., et al.
Publicado: (2017) -
A comparative study based on image quality and clinical task performance for CT reconstruction algorithms in radiotherapy
por: Li, Hua, et al.
Publicado: (2016) -
An improved method for automated calculation of the water‐equivalent diameter for estimating size‐specific dose in CT
por: Anam, Choirul, et al.
Publicado: (2021) -
Automated development of the contrast–detail curve based on statistical low‐contrast detectability in CT images
por: Anam, Choirul, et al.
Publicado: (2022)