Cargando…
Brachytherapy treatment planning commissioning: effect of the election of proper bibliography and finite size of TG‐43 input data on standard treatments
The aim of this work is to evaluate the performance of a commercial brachytherapy treatment planning system (TPS) with TG‐43 Vendors Input Data (VID), analyze possible discrepancies with respect to a proper reference source and its implications for standard treatments, and judge the effectiveness of...
Autores principales: | Valdés, Christian N., Píriz, Gustavo H., Lozano, Enrrique |
---|---|
Formato: | Online Artículo Texto |
Lenguaje: | English |
Publicado: |
John Wiley and Sons Inc.
2015
|
Materias: | |
Acceso en línea: | https://www.ncbi.nlm.nih.gov/pmc/articles/PMC5690009/ https://www.ncbi.nlm.nih.gov/pubmed/26218990 http://dx.doi.org/10.1120/jacmp.v16i4.4730 |
Ejemplares similares
-
Perturbation of TG‐43 parameters of the brachytherapy sources under insufficient scattering materials
por: Zehtabian, Mehdi, et al.
Publicado: (2013) -
Cylindrical coordinate‐based TG‐43U1 parameters for dose calculation around elongated brachytherapy sources
por: Awan, Shahid B., et al.
Publicado: (2008) -
Comparison of TG‐43 dosimetric parameters of brachytherapy sources obtained by three different versions of MCNP codes
por: Zaker, Neda, et al.
Publicado: (2016) -
TG‐43U1 parameterization of elongated RadioCoil [Formula: see text] brachytherapy sources
por: Dini, Sharifeh A., et al.
Publicado: (2007) -
Monte Carlo calculations and experimental measurements of the TG‐43U1‐recommended dosimetric parameters of (125)I (Model IR‐Seed2) brachytherapy source
por: Sheikholeslami, Sahar, et al.
Publicado: (2016)