Performance evaluation of the RITG148(+) set of TomoTherapy quality assurance tools using RTQA(2) radiochromic film

Version 6.3 of the [Formula: see text] software package offers eight automated analysis routines for quality assurance of the TomoTherapy platform. A performance evaluation of each routine was performed in order to compare [Formula: see text] results with traditionally accepted analysis techniques a...

Descripción completa

Detalles Bibliográficos
Autor principal: Lobb, Eric C.
Formato: Online Artículo Texto
Lenguaje:English
Publicado: John Wiley and Sons Inc. 2016
Materias:
Acceso en línea:https://www.ncbi.nlm.nih.gov/pmc/articles/PMC5690037/
https://www.ncbi.nlm.nih.gov/pubmed/27455492
http://dx.doi.org/10.1120/jacmp.v17i4.6178
_version_ 1783279515241283584
author Lobb, Eric C.
author_facet Lobb, Eric C.
author_sort Lobb, Eric C.
collection PubMed
description Version 6.3 of the [Formula: see text] software package offers eight automated analysis routines for quality assurance of the TomoTherapy platform. A performance evaluation of each routine was performed in order to compare [Formula: see text] results with traditionally accepted analysis techniques and verify that simulated changes in machine parameters are correctly identified by the software. Reference films were exposed according to AAPM TG‐148 methodology for each routine and the [Formula: see text] results were compared with either alternative software analysis techniques or manual analysis techniques in order to assess baseline agreement. Changes in machine performance were simulated through translational and rotational adjustments to subsequently irradiated films, and these films were analyzed to verify that the applied changes were accurately detected by each of the [Formula: see text] routines. For the Hounsfield unit routine, an assessment of the “Frame Averaging” functionality and the effects of phantom roll on the routine results are presented. All [Formula: see text] routines reported acceptable baseline results consistent with alternative analysis techniques, with 9 of the 11 baseline test results showing agreement of 0.1 mm/0.1° or better. Simulated changes were correctly identified by the [Formula: see text] routines within approximately [Formula: see text] with the exception of the Field Center vs. Jaw Setting routine, which was found to have limited accuracy in cases where field centers were not aligned for all jaw settings due to inaccurate autorotation of the film during analysis. The performance of the [Formula: see text] software package was found to be acceptable for introduction into our clinical environment as an automated alternative to traditional analysis techniques for routine TomoTherapy quality assurance testing. PACS number(s): 87.55.Qr
format Online
Article
Text
id pubmed-5690037
institution National Center for Biotechnology Information
language English
publishDate 2016
publisher John Wiley and Sons Inc.
record_format MEDLINE/PubMed
spelling pubmed-56900372018-04-02 Performance evaluation of the RITG148(+) set of TomoTherapy quality assurance tools using RTQA(2) radiochromic film Lobb, Eric C. J Appl Clin Med Phys Radiation Oncology Physics Version 6.3 of the [Formula: see text] software package offers eight automated analysis routines for quality assurance of the TomoTherapy platform. A performance evaluation of each routine was performed in order to compare [Formula: see text] results with traditionally accepted analysis techniques and verify that simulated changes in machine parameters are correctly identified by the software. Reference films were exposed according to AAPM TG‐148 methodology for each routine and the [Formula: see text] results were compared with either alternative software analysis techniques or manual analysis techniques in order to assess baseline agreement. Changes in machine performance were simulated through translational and rotational adjustments to subsequently irradiated films, and these films were analyzed to verify that the applied changes were accurately detected by each of the [Formula: see text] routines. For the Hounsfield unit routine, an assessment of the “Frame Averaging” functionality and the effects of phantom roll on the routine results are presented. All [Formula: see text] routines reported acceptable baseline results consistent with alternative analysis techniques, with 9 of the 11 baseline test results showing agreement of 0.1 mm/0.1° or better. Simulated changes were correctly identified by the [Formula: see text] routines within approximately [Formula: see text] with the exception of the Field Center vs. Jaw Setting routine, which was found to have limited accuracy in cases where field centers were not aligned for all jaw settings due to inaccurate autorotation of the film during analysis. The performance of the [Formula: see text] software package was found to be acceptable for introduction into our clinical environment as an automated alternative to traditional analysis techniques for routine TomoTherapy quality assurance testing. PACS number(s): 87.55.Qr John Wiley and Sons Inc. 2016-07-08 /pmc/articles/PMC5690037/ /pubmed/27455492 http://dx.doi.org/10.1120/jacmp.v17i4.6178 Text en © 2016 The Authors. This is an open access article under the terms of the Creative Commons Attribution (http://creativecommons.org/licenses/by/3.0/) License, which permits use, distribution and reproduction in any medium, provided the original work is properly cited.
spellingShingle Radiation Oncology Physics
Lobb, Eric C.
Performance evaluation of the RITG148(+) set of TomoTherapy quality assurance tools using RTQA(2) radiochromic film
title Performance evaluation of the RITG148(+) set of TomoTherapy quality assurance tools using RTQA(2) radiochromic film
title_full Performance evaluation of the RITG148(+) set of TomoTherapy quality assurance tools using RTQA(2) radiochromic film
title_fullStr Performance evaluation of the RITG148(+) set of TomoTherapy quality assurance tools using RTQA(2) radiochromic film
title_full_unstemmed Performance evaluation of the RITG148(+) set of TomoTherapy quality assurance tools using RTQA(2) radiochromic film
title_short Performance evaluation of the RITG148(+) set of TomoTherapy quality assurance tools using RTQA(2) radiochromic film
title_sort performance evaluation of the ritg148(+) set of tomotherapy quality assurance tools using rtqa(2) radiochromic film
topic Radiation Oncology Physics
url https://www.ncbi.nlm.nih.gov/pmc/articles/PMC5690037/
https://www.ncbi.nlm.nih.gov/pubmed/27455492
http://dx.doi.org/10.1120/jacmp.v17i4.6178
work_keys_str_mv AT lobbericc performanceevaluationoftheritg148setoftomotherapyqualityassurancetoolsusingrtqa2radiochromicfilm