Cargando…

Measurement of image rotation angle in CT for radiotherapy treatment planning

The geometric accuracy of CT images is essential when they are used for planning radiation therapy. In this study, a method of quantitative testing of image rotation in CT is presented, based on the automated analysis of an image of a phantom half‐filled with water. A plug‐in was written for ImageJ...

Descripción completa

Detalles Bibliográficos
Autores principales: Skrzynski, Witold, Ślusarczyk‐Kacprzyk, Wioletta
Formato: Online Artículo Texto
Lenguaje:English
Publicado: John Wiley and Sons Inc. 2016
Materias:
Acceso en línea:https://www.ncbi.nlm.nih.gov/pmc/articles/PMC5690054/
https://www.ncbi.nlm.nih.gov/pubmed/27455498
http://dx.doi.org/10.1120/jacmp.v17i4.6203
_version_ 1783279519317098496
author Skrzynski, Witold
Ślusarczyk‐Kacprzyk, Wioletta
author_facet Skrzynski, Witold
Ślusarczyk‐Kacprzyk, Wioletta
author_sort Skrzynski, Witold
collection PubMed
description The geometric accuracy of CT images is essential when they are used for planning radiation therapy. In this study, a method of quantitative testing of image rotation in CT is presented, based on the automated analysis of an image of a phantom half‐filled with water. A plug‐in was written for ImageJ software for automated detection of a water surface in an image and measurement of its angle relative to a horizontal line. A water phantom can be used for the evaluation of image rotation for axial mode. In helical mode the movement of the table would cause movement of the water. The difference between image rotation for axial and helical scans can be evaluated by measuring and comparing the angles of the tabletop surface for both modes. Preliminary results of measurements for three CT scanners show that image rotation does not exceed 0.5°, and is less than 0.1° for the dedicated CT simulator. It was observed that, for one CT scanner image, the rotation angle depended on tube rotation time. PACS number(s): 87.57.Q‐, 87.55.Qr, 06.30.Bp
format Online
Article
Text
id pubmed-5690054
institution National Center for Biotechnology Information
language English
publishDate 2016
publisher John Wiley and Sons Inc.
record_format MEDLINE/PubMed
spelling pubmed-56900542018-04-02 Measurement of image rotation angle in CT for radiotherapy treatment planning Skrzynski, Witold Ślusarczyk‐Kacprzyk, Wioletta J Appl Clin Med Phys Medical Imaging The geometric accuracy of CT images is essential when they are used for planning radiation therapy. In this study, a method of quantitative testing of image rotation in CT is presented, based on the automated analysis of an image of a phantom half‐filled with water. A plug‐in was written for ImageJ software for automated detection of a water surface in an image and measurement of its angle relative to a horizontal line. A water phantom can be used for the evaluation of image rotation for axial mode. In helical mode the movement of the table would cause movement of the water. The difference between image rotation for axial and helical scans can be evaluated by measuring and comparing the angles of the tabletop surface for both modes. Preliminary results of measurements for three CT scanners show that image rotation does not exceed 0.5°, and is less than 0.1° for the dedicated CT simulator. It was observed that, for one CT scanner image, the rotation angle depended on tube rotation time. PACS number(s): 87.57.Q‐, 87.55.Qr, 06.30.Bp John Wiley and Sons Inc. 2016-07-08 /pmc/articles/PMC5690054/ /pubmed/27455498 http://dx.doi.org/10.1120/jacmp.v17i4.6203 Text en © 2016 The Authors. This is an open access article under the terms of the Creative Commons Attribution (http://creativecommons.org/licenses/by/3.0/) License, which permits use, distribution and reproduction in any medium, provided the original work is properly cited.
spellingShingle Medical Imaging
Skrzynski, Witold
Ślusarczyk‐Kacprzyk, Wioletta
Measurement of image rotation angle in CT for radiotherapy treatment planning
title Measurement of image rotation angle in CT for radiotherapy treatment planning
title_full Measurement of image rotation angle in CT for radiotherapy treatment planning
title_fullStr Measurement of image rotation angle in CT for radiotherapy treatment planning
title_full_unstemmed Measurement of image rotation angle in CT for radiotherapy treatment planning
title_short Measurement of image rotation angle in CT for radiotherapy treatment planning
title_sort measurement of image rotation angle in ct for radiotherapy treatment planning
topic Medical Imaging
url https://www.ncbi.nlm.nih.gov/pmc/articles/PMC5690054/
https://www.ncbi.nlm.nih.gov/pubmed/27455498
http://dx.doi.org/10.1120/jacmp.v17i4.6203
work_keys_str_mv AT skrzynskiwitold measurementofimagerotationangleinctforradiotherapytreatmentplanning
AT slusarczykkacprzykwioletta measurementofimagerotationangleinctforradiotherapytreatmentplanning